Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population

INTRODUCTION: The Sysmex XN-series haematology analyser has newly adopted a fluorescent channel to measure immature platelet fraction (IPF). To promote the clinical utility of this promising parameter, establishing a reliable reference interval is mandatory. According to previous studies, IPF values...

Descripción completa

Detalles Bibliográficos
Autores principales: Imperiali, Claudia E., Arbiol-Roca, Ariadna, Sanchez-Navarro, Lourdes, Dastis-Arias, Macarena, Lopez-Delgado, Juan C., Cortes-Bosch, Anna, Sancho-Cerro, Ana, Dot-Bach, Dolors
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5812701/
https://www.ncbi.nlm.nih.gov/pubmed/29472803
http://dx.doi.org/10.11613/BM.2018.010708
_version_ 1783300073505947648
author Imperiali, Claudia E.
Arbiol-Roca, Ariadna
Sanchez-Navarro, Lourdes
Dastis-Arias, Macarena
Lopez-Delgado, Juan C.
Cortes-Bosch, Anna
Sancho-Cerro, Ana
Dot-Bach, Dolors
author_facet Imperiali, Claudia E.
Arbiol-Roca, Ariadna
Sanchez-Navarro, Lourdes
Dastis-Arias, Macarena
Lopez-Delgado, Juan C.
Cortes-Bosch, Anna
Sancho-Cerro, Ana
Dot-Bach, Dolors
author_sort Imperiali, Claudia E.
collection PubMed
description INTRODUCTION: The Sysmex XN-series haematology analyser has newly adopted a fluorescent channel to measure immature platelet fraction (IPF). To promote the clinical utility of this promising parameter, establishing a reliable reference interval is mandatory. According to previous studies, IPF values may be affected by the employed analyser and the ethnic background of the individual, but no differences seem to be found between individuals’ genders. Therefore, this study aimed to define the reference interval for IPF in a Spanish population following Clinical and Laboratory Standard Institute (CLSI) guidelines. MATERIALS AND METHODS: A total of 153 healthy Caucasian adults from Spain met the inclusion criteria. IPF measurement was performed by means of a Sysmex XN-2000 haematology analyser. A non-parametric percentile method was used to calculate the reference intervals in accordance with CLSI guidelines. RESULTS: The obtained reference interval for IPF on the Sysmex XN-2000 was 1.6–9.6% (90% confidence intervals (CIs) were 1.5–1.8 and 9.3–11.5, respectively). No significant gender difference in IPF reference intervals was observed (P = 0.101). CONCLUSIONS: This study provides, for the first time, a reference interval for IPF using a Sysmex XN-2000 in a Spanish population, ranging from 1.6 to 9.6%. These data are needed to evaluate platelet production in several conditions such as thrombocytopenia, inflammatory states and cardiovascular diseases, as well as for future research.
format Online
Article
Text
id pubmed-5812701
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-58127012018-02-22 Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population Imperiali, Claudia E. Arbiol-Roca, Ariadna Sanchez-Navarro, Lourdes Dastis-Arias, Macarena Lopez-Delgado, Juan C. Cortes-Bosch, Anna Sancho-Cerro, Ana Dot-Bach, Dolors Biochem Med (Zagreb) Original Papers INTRODUCTION: The Sysmex XN-series haematology analyser has newly adopted a fluorescent channel to measure immature platelet fraction (IPF). To promote the clinical utility of this promising parameter, establishing a reliable reference interval is mandatory. According to previous studies, IPF values may be affected by the employed analyser and the ethnic background of the individual, but no differences seem to be found between individuals’ genders. Therefore, this study aimed to define the reference interval for IPF in a Spanish population following Clinical and Laboratory Standard Institute (CLSI) guidelines. MATERIALS AND METHODS: A total of 153 healthy Caucasian adults from Spain met the inclusion criteria. IPF measurement was performed by means of a Sysmex XN-2000 haematology analyser. A non-parametric percentile method was used to calculate the reference intervals in accordance with CLSI guidelines. RESULTS: The obtained reference interval for IPF on the Sysmex XN-2000 was 1.6–9.6% (90% confidence intervals (CIs) were 1.5–1.8 and 9.3–11.5, respectively). No significant gender difference in IPF reference intervals was observed (P = 0.101). CONCLUSIONS: This study provides, for the first time, a reference interval for IPF using a Sysmex XN-2000 in a Spanish population, ranging from 1.6 to 9.6%. These data are needed to evaluate platelet production in several conditions such as thrombocytopenia, inflammatory states and cardiovascular diseases, as well as for future research. Croatian Society of Medical Biochemistry and Laboratory Medicine 2018-02-15 2018-02-15 /pmc/articles/PMC5812701/ /pubmed/29472803 http://dx.doi.org/10.11613/BM.2018.010708 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Papers
Imperiali, Claudia E.
Arbiol-Roca, Ariadna
Sanchez-Navarro, Lourdes
Dastis-Arias, Macarena
Lopez-Delgado, Juan C.
Cortes-Bosch, Anna
Sancho-Cerro, Ana
Dot-Bach, Dolors
Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population
title Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population
title_full Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population
title_fullStr Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population
title_full_unstemmed Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population
title_short Reference interval for immature platelet fraction on Sysmex XN haematology analyser in adult population
title_sort reference interval for immature platelet fraction on sysmex xn haematology analyser in adult population
topic Original Papers
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5812701/
https://www.ncbi.nlm.nih.gov/pubmed/29472803
http://dx.doi.org/10.11613/BM.2018.010708
work_keys_str_mv AT imperialiclaudiae referenceintervalforimmatureplateletfractiononsysmexxnhaematologyanalyserinadultpopulation
AT arbiolrocaariadna referenceintervalforimmatureplateletfractiononsysmexxnhaematologyanalyserinadultpopulation
AT sancheznavarrolourdes referenceintervalforimmatureplateletfractiononsysmexxnhaematologyanalyserinadultpopulation
AT dastisariasmacarena referenceintervalforimmatureplateletfractiononsysmexxnhaematologyanalyserinadultpopulation
AT lopezdelgadojuanc referenceintervalforimmatureplateletfractiononsysmexxnhaematologyanalyserinadultpopulation
AT cortesboschanna referenceintervalforimmatureplateletfractiononsysmexxnhaematologyanalyserinadultpopulation
AT sanchocerroana referenceintervalforimmatureplateletfractiononsysmexxnhaematologyanalyserinadultpopulation
AT dotbachdolors referenceintervalforimmatureplateletfractiononsysmexxnhaematologyanalyserinadultpopulation