Cargando…
Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy
BACKGROUND: Volumetric-modulated Dynamic WaveArc therapy (VMDWAT) is a non-coplanar continuous volumetric modulated radiation therapy (VMAT) delivery technique. Here, we monitored mechanical errors and their impact on dose distributions in VMDWAT using logfiles throughout the course of treatment. ME...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5813375/ https://www.ncbi.nlm.nih.gov/pubmed/29444693 http://dx.doi.org/10.1186/s13014-018-0972-7 |
_version_ | 1783300182032515072 |
---|---|
author | Hirashima, Hideaki Nakamura, Mitsuhiro Miyabe, Yuki Uto, Megumi Nakamura, Kiyonao Mizowaki, Takashi |
author_facet | Hirashima, Hideaki Nakamura, Mitsuhiro Miyabe, Yuki Uto, Megumi Nakamura, Kiyonao Mizowaki, Takashi |
author_sort | Hirashima, Hideaki |
collection | PubMed |
description | BACKGROUND: Volumetric-modulated Dynamic WaveArc therapy (VMDWAT) is a non-coplanar continuous volumetric modulated radiation therapy (VMAT) delivery technique. Here, we monitored mechanical errors and their impact on dose distributions in VMDWAT using logfiles throughout the course of treatment. METHODS: Fifteen patients were enrolled (2 skull base tumor patients and 13 prostate cancer patients). VMDWAT plans were created for the enrolled patients. The prescribed dose for the skull base tumor was set as 54 Gy at 1.8 Gy per fraction, and that for the prostate cancer was set as 72 to 78 Gy at 2 Gy per fraction. We acquired logfiles to monitor mechanical errors and their impact on dose distribution in each fraction. The root mean square error (RMSE) in the multi-leaf collimator (MLC), gantry angle, O-ring angle and monitor unit (MU) were calculated using logfiles throughout the course of VMDWAT for each patient. The dosimetric impact of mechanical errors throughout the course of VMDWAT was verified using a logfile-based dose reconstruction method. Dosimetric errors between the reconstructed plans and the original plans were assessed. RESULTS: A total of 517 datasets, including 55 datasets for the 2 skull base tumor patients and 462 datasets for the 13 prostate cancer patients, were acquired. The RMSE values were less than 0.1 mm, 0.2°, 0.1°, and 0.4 MU for MLC position, gantry angle, O-ring angle, and MU, respectively. For the skull base tumors, the absolute mean dosimetric errors and two standard deviations throughout the course of treatment were less than 1.4% and 1.1%, respectively. For prostate cancer, these absolute values were less than 0.3% and 0.5%, respectively. The largest dosimetric error of 2.5% was observed in a skull base tumor patient. The resultant dosimetric error in the accumulated daily delivered dose distribution, in the patient with the largest error, was up to 1.6% for all dose-volumetric parameters relative to the planned dose distribution. CONCLUSIONS: MLC position, gantry rotation, O-ring rotation and MU were highly accurate and stable throughout the course of treatment. The daily dosimetric errors due to mechanical errors were small. VMDWAT provided high delivery accuracy and stability throughout the course of treatment. TRIAL REGISTRATION: UMIN000023870. Registered: 1 October 2016. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s13014-018-0972-7) contains supplementary material, which is available to authorized users. |
format | Online Article Text |
id | pubmed-5813375 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-58133752018-02-16 Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy Hirashima, Hideaki Nakamura, Mitsuhiro Miyabe, Yuki Uto, Megumi Nakamura, Kiyonao Mizowaki, Takashi Radiat Oncol Research BACKGROUND: Volumetric-modulated Dynamic WaveArc therapy (VMDWAT) is a non-coplanar continuous volumetric modulated radiation therapy (VMAT) delivery technique. Here, we monitored mechanical errors and their impact on dose distributions in VMDWAT using logfiles throughout the course of treatment. METHODS: Fifteen patients were enrolled (2 skull base tumor patients and 13 prostate cancer patients). VMDWAT plans were created for the enrolled patients. The prescribed dose for the skull base tumor was set as 54 Gy at 1.8 Gy per fraction, and that for the prostate cancer was set as 72 to 78 Gy at 2 Gy per fraction. We acquired logfiles to monitor mechanical errors and their impact on dose distribution in each fraction. The root mean square error (RMSE) in the multi-leaf collimator (MLC), gantry angle, O-ring angle and monitor unit (MU) were calculated using logfiles throughout the course of VMDWAT for each patient. The dosimetric impact of mechanical errors throughout the course of VMDWAT was verified using a logfile-based dose reconstruction method. Dosimetric errors between the reconstructed plans and the original plans were assessed. RESULTS: A total of 517 datasets, including 55 datasets for the 2 skull base tumor patients and 462 datasets for the 13 prostate cancer patients, were acquired. The RMSE values were less than 0.1 mm, 0.2°, 0.1°, and 0.4 MU for MLC position, gantry angle, O-ring angle, and MU, respectively. For the skull base tumors, the absolute mean dosimetric errors and two standard deviations throughout the course of treatment were less than 1.4% and 1.1%, respectively. For prostate cancer, these absolute values were less than 0.3% and 0.5%, respectively. The largest dosimetric error of 2.5% was observed in a skull base tumor patient. The resultant dosimetric error in the accumulated daily delivered dose distribution, in the patient with the largest error, was up to 1.6% for all dose-volumetric parameters relative to the planned dose distribution. CONCLUSIONS: MLC position, gantry rotation, O-ring rotation and MU were highly accurate and stable throughout the course of treatment. The daily dosimetric errors due to mechanical errors were small. VMDWAT provided high delivery accuracy and stability throughout the course of treatment. TRIAL REGISTRATION: UMIN000023870. Registered: 1 October 2016. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s13014-018-0972-7) contains supplementary material, which is available to authorized users. BioMed Central 2018-02-14 /pmc/articles/PMC5813375/ /pubmed/29444693 http://dx.doi.org/10.1186/s13014-018-0972-7 Text en © The Author(s). 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Hirashima, Hideaki Nakamura, Mitsuhiro Miyabe, Yuki Uto, Megumi Nakamura, Kiyonao Mizowaki, Takashi Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy |
title | Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy |
title_full | Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy |
title_fullStr | Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy |
title_full_unstemmed | Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy |
title_short | Monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy |
title_sort | monitoring of mechanical errors and their dosimetric impact throughout the course of non-coplanar continuous volumetric-modulated arc therapy |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5813375/ https://www.ncbi.nlm.nih.gov/pubmed/29444693 http://dx.doi.org/10.1186/s13014-018-0972-7 |
work_keys_str_mv | AT hirashimahideaki monitoringofmechanicalerrorsandtheirdosimetricimpactthroughoutthecourseofnoncoplanarcontinuousvolumetricmodulatedarctherapy AT nakamuramitsuhiro monitoringofmechanicalerrorsandtheirdosimetricimpactthroughoutthecourseofnoncoplanarcontinuousvolumetricmodulatedarctherapy AT miyabeyuki monitoringofmechanicalerrorsandtheirdosimetricimpactthroughoutthecourseofnoncoplanarcontinuousvolumetricmodulatedarctherapy AT utomegumi monitoringofmechanicalerrorsandtheirdosimetricimpactthroughoutthecourseofnoncoplanarcontinuousvolumetricmodulatedarctherapy AT nakamurakiyonao monitoringofmechanicalerrorsandtheirdosimetricimpactthroughoutthecourseofnoncoplanarcontinuousvolumetricmodulatedarctherapy AT mizowakitakashi monitoringofmechanicalerrorsandtheirdosimetricimpactthroughoutthecourseofnoncoplanarcontinuousvolumetricmodulatedarctherapy |