Plugger temperature of cordless heat carriers according to the time elapsed

OBJECTIVE: The purpose of this study was to measure the temperature of the plugger tip of 3 cordless heat carriers set at 200°C. MATERIALS AND METHODS: Pluggers of the same taper (0.06, 0.08, 0.10) and similar tip sizes (sizes of 50 and 55) from 3 cordless heat carriers, namely SuperEndo-α(2) (B &am...

Descripción completa

Detalles Bibliográficos
Autores principales: Chang, Hoon-Sang, Park, Se-Hee, Cho, Kyung-Mo, Kim, Jin-Woo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Conservative Dentistry 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5816989/
https://www.ncbi.nlm.nih.gov/pubmed/29487842
http://dx.doi.org/10.5395/rde.2018.43.e12
_version_ 1783300792560648192
author Chang, Hoon-Sang
Park, Se-Hee
Cho, Kyung-Mo
Kim, Jin-Woo
author_facet Chang, Hoon-Sang
Park, Se-Hee
Cho, Kyung-Mo
Kim, Jin-Woo
author_sort Chang, Hoon-Sang
collection PubMed
description OBJECTIVE: The purpose of this study was to measure the temperature of the plugger tip of 3 cordless heat carriers set at 200°C. MATERIALS AND METHODS: Pluggers of the same taper (0.06, 0.08, 0.10) and similar tip sizes (sizes of 50 and 55) from 3 cordless heat carriers, namely SuperEndo-α(2) (B & L Biotech), Friendo (DXM), and Dia-Pen (Diadent), were used and an electric heat carrier, System B (SybronEndo), was used as the control. The plugger tips were covered with customized copper sleeves, heated for 10 seconds, and the temperature was recorded with a computerized measurement system attached to a K-type thermometer at room temperature (n = 10). The data were analyzed with 2-way analysis of variance at a 5% level of significance. RESULTS: The peak temperature of the plugger tips was significantly affected by the plugger taper and by the heat carrier brand (p < 0.05). The peak temperature of the plugger tips was between 177°C and 325°C. The temperature peaked at 207°C–231°C for the 0.06 taper pluggers, 195°C–313°C for the 0.08 taper pluggers, and 177°C–325°C for the 0.10 taper pluggers. Only 5 of the 12 plugger tips showed a temperature of 200°C ± 10°C. The time required to reach the highest temperature or 200°C ± 10°C was at least 4 seconds. CONCLUSION: When using cordless heat carriers, clinicians should pay attention to the temperature setting and to the activation time needed to reach the intended temperature of the pluggers.
format Online
Article
Text
id pubmed-5816989
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher The Korean Academy of Conservative Dentistry
record_format MEDLINE/PubMed
spelling pubmed-58169892018-02-27 Plugger temperature of cordless heat carriers according to the time elapsed Chang, Hoon-Sang Park, Se-Hee Cho, Kyung-Mo Kim, Jin-Woo Restor Dent Endod Research Article OBJECTIVE: The purpose of this study was to measure the temperature of the plugger tip of 3 cordless heat carriers set at 200°C. MATERIALS AND METHODS: Pluggers of the same taper (0.06, 0.08, 0.10) and similar tip sizes (sizes of 50 and 55) from 3 cordless heat carriers, namely SuperEndo-α(2) (B & L Biotech), Friendo (DXM), and Dia-Pen (Diadent), were used and an electric heat carrier, System B (SybronEndo), was used as the control. The plugger tips were covered with customized copper sleeves, heated for 10 seconds, and the temperature was recorded with a computerized measurement system attached to a K-type thermometer at room temperature (n = 10). The data were analyzed with 2-way analysis of variance at a 5% level of significance. RESULTS: The peak temperature of the plugger tips was significantly affected by the plugger taper and by the heat carrier brand (p < 0.05). The peak temperature of the plugger tips was between 177°C and 325°C. The temperature peaked at 207°C–231°C for the 0.06 taper pluggers, 195°C–313°C for the 0.08 taper pluggers, and 177°C–325°C for the 0.10 taper pluggers. Only 5 of the 12 plugger tips showed a temperature of 200°C ± 10°C. The time required to reach the highest temperature or 200°C ± 10°C was at least 4 seconds. CONCLUSION: When using cordless heat carriers, clinicians should pay attention to the temperature setting and to the activation time needed to reach the intended temperature of the pluggers. The Korean Academy of Conservative Dentistry 2018-02-07 /pmc/articles/PMC5816989/ /pubmed/29487842 http://dx.doi.org/10.5395/rde.2018.43.e12 Text en Copyright © 2018. The Korean Academy of Conservative Dentistry https://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Chang, Hoon-Sang
Park, Se-Hee
Cho, Kyung-Mo
Kim, Jin-Woo
Plugger temperature of cordless heat carriers according to the time elapsed
title Plugger temperature of cordless heat carriers according to the time elapsed
title_full Plugger temperature of cordless heat carriers according to the time elapsed
title_fullStr Plugger temperature of cordless heat carriers according to the time elapsed
title_full_unstemmed Plugger temperature of cordless heat carriers according to the time elapsed
title_short Plugger temperature of cordless heat carriers according to the time elapsed
title_sort plugger temperature of cordless heat carriers according to the time elapsed
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5816989/
https://www.ncbi.nlm.nih.gov/pubmed/29487842
http://dx.doi.org/10.5395/rde.2018.43.e12
work_keys_str_mv AT changhoonsang pluggertemperatureofcordlessheatcarriersaccordingtothetimeelapsed
AT parksehee pluggertemperatureofcordlessheatcarriersaccordingtothetimeelapsed
AT chokyungmo pluggertemperatureofcordlessheatcarriersaccordingtothetimeelapsed
AT kimjinwoo pluggertemperatureofcordlessheatcarriersaccordingtothetimeelapsed