Cargando…

Assessment of the Electronic Retinal Implant Alpha AMS in Restoring Vision to Blind Patients with End-Stage Retinitis Pigmentosa

PURPOSE: To report the initial efficacy results of the Retina Implant Alpha AMS (Retina Implant AG, Reutlingen, Germany) for partial restoration of vision in end-stage retinitis pigmentosa (RP). DESIGN: Prospective, single-arm, investigator-sponsored interventional clinical trial. Within-participant...

Descripción completa

Detalles Bibliográficos
Autores principales: Edwards, Thomas L., Cottriall, Charles L., Xue, Kanmin, Simunovic, Matthew P., Ramsden, James D., Zrenner, Eberhart, MacLaren, Robert E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5818267/
https://www.ncbi.nlm.nih.gov/pubmed/29110946
http://dx.doi.org/10.1016/j.ophtha.2017.09.019
Descripción
Sumario:PURPOSE: To report the initial efficacy results of the Retina Implant Alpha AMS (Retina Implant AG, Reutlingen, Germany) for partial restoration of vision in end-stage retinitis pigmentosa (RP). DESIGN: Prospective, single-arm, investigator-sponsored interventional clinical trial. Within-participant control comprising residual vision with the retinal implant switched ON versus OFF in the implanted eye. PARTICIPANTS: The Retina Implant Alpha AMS was implanted into the worse-seeing eye of 6 participants with end-stage RP and no useful perception of light vision. Eligibility criteria included previous normal vision for ≥12 years and no significant ocular or systemic comorbidity. METHODS: Vision assessments were scheduled at 1, 2, 3, 6, 9, and 12 months postimplantation. They comprised tabletop object recognition tasks, a self-assessment mobility questionnaire, and screen-based tests including Basic Light and Motion (BaLM), grating acuity, and greyscale contrast discrimination. A full-field stimulus test (FST) was also performed. MAIN OUTCOME MEASURES: Improvement in activities of daily living, recognition tasks, and assessments of light perception with the implant ON compared with OFF. RESULTS: All 6 participants underwent successful implantation. Light perception and temporal resolution with the implant ON were achieved in all participants. Light localization was achieved with the implant ON in all but 1 participant (P4) in whom the chip was not functioning optimally because of a combination of iatrogenic intraoperative implant damage and incorrect implantation. Implant ON correct grating detections (which were at chance level with implant OFF) were recorded in the other 5 participants, ranging from 0.1 to 3.33 cycles/degree on 1 occasion. The ability to locate high-contrast tabletop objects not seen with the implant OFF was partially restored with the implant ON in all but 1 participant (P4). There were 2 incidents of conjunctival erosion and 1 inferotemporal macula-on retinal detachment, which were successfully repaired, and 2 incidents of inadvertent damage to the implant during surgery (P3 and P4). CONCLUSIONS: The Alpha AMS subretinal implant improved visual performance in 5 of 6 participants and has exhibited ongoing function for up to 24 months. Although implantation surgery remains challenging, new developments such as OCT microscope guidance added refinements to the surgical technique.