Cargando…

Optical surgical navigation system causes pulse oximeter malfunction

An optical surgical navigation system is used as a navigator to facilitate surgical approaches, and pulse oximeters provide valuable information for anesthetic management. However, saw-tooth waves on the monitor of a pulse oximeter and the inability of the pulse oximeter to accurately record the sat...

Descripción completa

Detalles Bibliográficos
Autores principales: Satoh, Masaaki, Hara, Tetsuhito, Tamai, Kenji, Shiba, Juntaro, Hotta, Kunihisa, Takeuchi, Mamoru, Watanabe, Eiju
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5818690/
https://www.ncbi.nlm.nih.gov/pubmed/29497637
http://dx.doi.org/10.1186/s40981-015-0007-4
_version_ 1783301077281538048
author Satoh, Masaaki
Hara, Tetsuhito
Tamai, Kenji
Shiba, Juntaro
Hotta, Kunihisa
Takeuchi, Mamoru
Watanabe, Eiju
author_facet Satoh, Masaaki
Hara, Tetsuhito
Tamai, Kenji
Shiba, Juntaro
Hotta, Kunihisa
Takeuchi, Mamoru
Watanabe, Eiju
author_sort Satoh, Masaaki
collection PubMed
description An optical surgical navigation system is used as a navigator to facilitate surgical approaches, and pulse oximeters provide valuable information for anesthetic management. However, saw-tooth waves on the monitor of a pulse oximeter and the inability of the pulse oximeter to accurately record the saturation of a percutaneous artery were observed when a surgeon started an optical navigation system. The current case is thought to be the first report of this navigation system interfering with pulse oximetry. The causes of pulse jamming and how to manage an optical navigation system are discussed.
format Online
Article
Text
id pubmed-5818690
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-58186902018-02-27 Optical surgical navigation system causes pulse oximeter malfunction Satoh, Masaaki Hara, Tetsuhito Tamai, Kenji Shiba, Juntaro Hotta, Kunihisa Takeuchi, Mamoru Watanabe, Eiju JA Clin Rep Case Report An optical surgical navigation system is used as a navigator to facilitate surgical approaches, and pulse oximeters provide valuable information for anesthetic management. However, saw-tooth waves on the monitor of a pulse oximeter and the inability of the pulse oximeter to accurately record the saturation of a percutaneous artery were observed when a surgeon started an optical navigation system. The current case is thought to be the first report of this navigation system interfering with pulse oximetry. The causes of pulse jamming and how to manage an optical navigation system are discussed. Springer Berlin Heidelberg 2015-08-27 2015 /pmc/articles/PMC5818690/ /pubmed/29497637 http://dx.doi.org/10.1186/s40981-015-0007-4 Text en © The Author(s) 2015 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited.
spellingShingle Case Report
Satoh, Masaaki
Hara, Tetsuhito
Tamai, Kenji
Shiba, Juntaro
Hotta, Kunihisa
Takeuchi, Mamoru
Watanabe, Eiju
Optical surgical navigation system causes pulse oximeter malfunction
title Optical surgical navigation system causes pulse oximeter malfunction
title_full Optical surgical navigation system causes pulse oximeter malfunction
title_fullStr Optical surgical navigation system causes pulse oximeter malfunction
title_full_unstemmed Optical surgical navigation system causes pulse oximeter malfunction
title_short Optical surgical navigation system causes pulse oximeter malfunction
title_sort optical surgical navigation system causes pulse oximeter malfunction
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5818690/
https://www.ncbi.nlm.nih.gov/pubmed/29497637
http://dx.doi.org/10.1186/s40981-015-0007-4
work_keys_str_mv AT satohmasaaki opticalsurgicalnavigationsystemcausespulseoximetermalfunction
AT haratetsuhito opticalsurgicalnavigationsystemcausespulseoximetermalfunction
AT tamaikenji opticalsurgicalnavigationsystemcausespulseoximetermalfunction
AT shibajuntaro opticalsurgicalnavigationsystemcausespulseoximetermalfunction
AT hottakunihisa opticalsurgicalnavigationsystemcausespulseoximetermalfunction
AT takeuchimamoru opticalsurgicalnavigationsystemcausespulseoximetermalfunction
AT watanabeeiju opticalsurgicalnavigationsystemcausespulseoximetermalfunction