Reiterative ventricular fibrillation caused by R-on-T during temporary epicardial pacing: a case report
Epicardial pacemaker wire insertion is standard following cardiothoracic surgery. However, undersensing of pacing wires may cause the R-on-T phenomenon, which induces ventricular fibrillation. We report a case of a male patient with severe mitral regurgitation scheduled for mitral valve replacement...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5818722/ https://www.ncbi.nlm.nih.gov/pubmed/29497660 http://dx.doi.org/10.1186/s40981-016-0029-6 |
Sumario: | Epicardial pacemaker wire insertion is standard following cardiothoracic surgery. However, undersensing of pacing wires may cause the R-on-T phenomenon, which induces ventricular fibrillation. We report a case of a male patient with severe mitral regurgitation scheduled for mitral valve replacement who experienced two ventricular fibrillation episodes related to the R-on-T phenomenon caused by undersensing of the epicardial pacing wire. Both undersensing events happened despite an appropriately low sensing threshold. Notably, the stimulated T wave followed the QRS of the premature ventricular contraction (PVC). This case suggests that a PVC’s R wave may be undersensed despite a low sensing threshold. This critical complication may have occurred because pacemakers sense R waves using a slew rate, which is the quotient of voltage over time. As a result, pacemakers may undersense wide QRS waves such as PVCs. Avoiding this dangerous phenomenon completely is not possible using epicardial pacemakers; therefore we recommend carefully adapting epicardial pacing especially when PVC waves occur frequently. |
---|