Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards

BACKGROUND: Ultraviolet germicidal irradiation (UVGI) systems are gaining popularity, however objective comparisons of their characteristics are lacking. While environmental cultures and reduction of hospital-associated infections rates are excellent study endpoints, they are impractical for centers...

Descripción completa

Detalles Bibliográficos
Autores principales: Masse, Vincent, Hartley, Michael J., Edmond, Michael B., Diekema, Daniel J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5824448/
https://www.ncbi.nlm.nih.gov/pubmed/29484174
http://dx.doi.org/10.1186/s13756-018-0317-1
_version_ 1783302025586409472
author Masse, Vincent
Hartley, Michael J.
Edmond, Michael B.
Diekema, Daniel J.
author_facet Masse, Vincent
Hartley, Michael J.
Edmond, Michael B.
Diekema, Daniel J.
author_sort Masse, Vincent
collection PubMed
description BACKGROUND: Ultraviolet germicidal irradiation (UVGI) systems are gaining popularity, however objective comparisons of their characteristics are lacking. While environmental cultures and reduction of hospital-associated infections rates are excellent study endpoints, they are impractical for centers with limited resources who want to compare or optimize UVGI systems use. METHODS: We evaluated radiometry and commercial test cards, two simple and low cost tools, to compare 2 full size UVGI systems (Tru-D and Optimum-UV Enlight) and 2 small units (Lumalier EDU 435 and MRSA-UV Turbo-UV). RESULTS: Radiometry-derived output curves show that if both large devices emit enough energy to reach C. difficile lethal doses at 10 ft, the reduction in output in distance is almost perfectly logarithmic. In a patient room environment, Enlight and Tru-D performed similarly when compared using radiometry and commercial test cards. The two small devices reached C. difficile range around the bathroom with the device raised above the floor, but longer times are needed. CONCLUSIONS: Despite different workflows and price points, no clear superiority emerges between Tru-D and Enlight. Bathroom disinfection should be dealt with separately from the main room and small, cheaper units can be used. Radiometry and commercial test cards are promising ways to compare UVGI systems, but further validation is needed using correlation with environmental cultures. TRIAL REGISTRATION: Not applicable.
format Online
Article
Text
id pubmed-5824448
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-58244482018-02-26 Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards Masse, Vincent Hartley, Michael J. Edmond, Michael B. Diekema, Daniel J. Antimicrob Resist Infect Control Research BACKGROUND: Ultraviolet germicidal irradiation (UVGI) systems are gaining popularity, however objective comparisons of their characteristics are lacking. While environmental cultures and reduction of hospital-associated infections rates are excellent study endpoints, they are impractical for centers with limited resources who want to compare or optimize UVGI systems use. METHODS: We evaluated radiometry and commercial test cards, two simple and low cost tools, to compare 2 full size UVGI systems (Tru-D and Optimum-UV Enlight) and 2 small units (Lumalier EDU 435 and MRSA-UV Turbo-UV). RESULTS: Radiometry-derived output curves show that if both large devices emit enough energy to reach C. difficile lethal doses at 10 ft, the reduction in output in distance is almost perfectly logarithmic. In a patient room environment, Enlight and Tru-D performed similarly when compared using radiometry and commercial test cards. The two small devices reached C. difficile range around the bathroom with the device raised above the floor, but longer times are needed. CONCLUSIONS: Despite different workflows and price points, no clear superiority emerges between Tru-D and Enlight. Bathroom disinfection should be dealt with separately from the main room and small, cheaper units can be used. Radiometry and commercial test cards are promising ways to compare UVGI systems, but further validation is needed using correlation with environmental cultures. TRIAL REGISTRATION: Not applicable. BioMed Central 2018-02-22 /pmc/articles/PMC5824448/ /pubmed/29484174 http://dx.doi.org/10.1186/s13756-018-0317-1 Text en © The Author(s). 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Masse, Vincent
Hartley, Michael J.
Edmond, Michael B.
Diekema, Daniel J.
Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards
title Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards
title_full Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards
title_fullStr Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards
title_full_unstemmed Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards
title_short Comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards
title_sort comparing and optimizing ultraviolet germicidal irradiation systems use for patient room terminal disinfection: an exploratory study using radiometry and commercial test cards
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5824448/
https://www.ncbi.nlm.nih.gov/pubmed/29484174
http://dx.doi.org/10.1186/s13756-018-0317-1
work_keys_str_mv AT massevincent comparingandoptimizingultravioletgermicidalirradiationsystemsuseforpatientroomterminaldisinfectionanexploratorystudyusingradiometryandcommercialtestcards
AT hartleymichaelj comparingandoptimizingultravioletgermicidalirradiationsystemsuseforpatientroomterminaldisinfectionanexploratorystudyusingradiometryandcommercialtestcards
AT edmondmichaelb comparingandoptimizingultravioletgermicidalirradiationsystemsuseforpatientroomterminaldisinfectionanexploratorystudyusingradiometryandcommercialtestcards
AT diekemadanielj comparingandoptimizingultravioletgermicidalirradiationsystemsuseforpatientroomterminaldisinfectionanexploratorystudyusingradiometryandcommercialtestcards