Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes

PURPOSE: To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster. METHODS: In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scan...

Descripción completa

Detalles Bibliográficos
Autores principales: Hua, Yanjun, Qiu, Wei, Xiao, Qiuyi, Wu, Qiang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5828443/
https://www.ncbi.nlm.nih.gov/pubmed/29486009
http://dx.doi.org/10.1371/journal.pone.0193023
_version_ 1783302647624761344
author Hua, Yanjun
Qiu, Wei
Xiao, Qiuyi
Wu, Qiang
author_facet Hua, Yanjun
Qiu, Wei
Xiao, Qiuyi
Wu, Qiang
author_sort Hua, Yanjun
collection PubMed
description PURPOSE: To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster. METHODS: In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scans were obtained by 2 observers using the Tomey OA-2000, and in the same session one observer used the IOLMaster (version 5.4.4.0006) for the measurements. About 1 week later, 3 scans were obtained by one observer using the Tomey OA-2000. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometer readings, pupil diameter (PD) and corneal diameter (CD) values measured by the Tomey OA-2000 and IOLMaster were analyzed. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within subject standard deviation (Sw) and 2.77Sw were calculated to assess the repeatability and reproducibility. The paired t test and Bland-Altman plots were used to analyze the differences and agreements of parameters measured by the two devices, respectively. RESULTS: Intraobserver repeatability, and interobserver and intersession reproducibility of the AL, CCT, ACD, LT, Kf, Ks, Km, PD and CD values measured by the Tomey OA-2000 biometer showed a CoV of less than 1% except that for PD, and an ICC of more than 0.97 except that for PD and CD. The AL, Kf, Ks, Km and CD values measured by the Tomey OA-2000 were 0.058 ± 0.094 mm, 0.088± 0.150 diopters (D), 0.163 ± 0.170 D, 0.127 ± 0.117 D and 0.171 ± 0.217 mm lower than those measured by the IOLMaster, respectively (all Ps < 0.05). However, the ACD values from the two devices were comparable (P = 0.169). The 95% linite of agreement (LoA) of the AL, ACD, CD and all keratometer readings were no more than 0.24 mm, 0.14 mm 0.60 mm and 0.5 D, respectively. CONCLUSION: Except for the PD and CD, the ocular parameters measured by the Tomey OA-2000 were highly repeatable and reproducible. Except for the CD value, there was good agreement of ocular parameters measured by the Tomey OA-2000 and the IOLMaster in healthy eyes.
format Online
Article
Text
id pubmed-5828443
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-58284432018-03-19 Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes Hua, Yanjun Qiu, Wei Xiao, Qiuyi Wu, Qiang PLoS One Research Article PURPOSE: To assess the precision (repeatability and reproducibility) of ocular parameters measured by the Tomey OA-2000 biometer, and to compare them with those measured by the IOLMaster. METHODS: In this prospective study, the right eyes of 108 healthy subjects were included. Three consecutive scans were obtained by 2 observers using the Tomey OA-2000, and in the same session one observer used the IOLMaster (version 5.4.4.0006) for the measurements. About 1 week later, 3 scans were obtained by one observer using the Tomey OA-2000. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), keratometer readings, pupil diameter (PD) and corneal diameter (CD) values measured by the Tomey OA-2000 and IOLMaster were analyzed. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within subject standard deviation (Sw) and 2.77Sw were calculated to assess the repeatability and reproducibility. The paired t test and Bland-Altman plots were used to analyze the differences and agreements of parameters measured by the two devices, respectively. RESULTS: Intraobserver repeatability, and interobserver and intersession reproducibility of the AL, CCT, ACD, LT, Kf, Ks, Km, PD and CD values measured by the Tomey OA-2000 biometer showed a CoV of less than 1% except that for PD, and an ICC of more than 0.97 except that for PD and CD. The AL, Kf, Ks, Km and CD values measured by the Tomey OA-2000 were 0.058 ± 0.094 mm, 0.088± 0.150 diopters (D), 0.163 ± 0.170 D, 0.127 ± 0.117 D and 0.171 ± 0.217 mm lower than those measured by the IOLMaster, respectively (all Ps < 0.05). However, the ACD values from the two devices were comparable (P = 0.169). The 95% linite of agreement (LoA) of the AL, ACD, CD and all keratometer readings were no more than 0.24 mm, 0.14 mm 0.60 mm and 0.5 D, respectively. CONCLUSION: Except for the PD and CD, the ocular parameters measured by the Tomey OA-2000 were highly repeatable and reproducible. Except for the CD value, there was good agreement of ocular parameters measured by the Tomey OA-2000 and the IOLMaster in healthy eyes. Public Library of Science 2018-02-27 /pmc/articles/PMC5828443/ /pubmed/29486009 http://dx.doi.org/10.1371/journal.pone.0193023 Text en © 2018 Hua et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Hua, Yanjun
Qiu, Wei
Xiao, Qiuyi
Wu, Qiang
Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes
title Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes
title_full Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes
title_fullStr Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes
title_full_unstemmed Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes
title_short Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes
title_sort precision (repeatability and reproducibility) of ocular parameters obtained by the tomey oa-2000 biometer compared to the iolmaster in healthy eyes
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5828443/
https://www.ncbi.nlm.nih.gov/pubmed/29486009
http://dx.doi.org/10.1371/journal.pone.0193023
work_keys_str_mv AT huayanjun precisionrepeatabilityandreproducibilityofocularparametersobtainedbythetomeyoa2000biometercomparedtotheiolmasterinhealthyeyes
AT qiuwei precisionrepeatabilityandreproducibilityofocularparametersobtainedbythetomeyoa2000biometercomparedtotheiolmasterinhealthyeyes
AT xiaoqiuyi precisionrepeatabilityandreproducibilityofocularparametersobtainedbythetomeyoa2000biometercomparedtotheiolmasterinhealthyeyes
AT wuqiang precisionrepeatabilityandreproducibilityofocularparametersobtainedbythetomeyoa2000biometercomparedtotheiolmasterinhealthyeyes