Accuracy and Precision of a Veterinary Neuronavigation System for Radiation Oncology Positioning

Conformal radiation treatment plans such as IMRT and other radiosurgery techniques require very precise patient positioning, typically within a millimeter of error for best results. CT cone beam, real-time navigation, and infrared position sensors are potential options for success but rarely present...

Descripción completa

Detalles Bibliográficos
Autores principales: Vanhaezebrouck, Isabelle F., Ballegeer, Elizabeth A., Frey, Stephen, Sieffert, Rob
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5831826/
https://www.ncbi.nlm.nih.gov/pubmed/29666822
http://dx.doi.org/10.1155/2018/6431749
Descripción
Sumario:Conformal radiation treatment plans such as IMRT and other radiosurgery techniques require very precise patient positioning, typically within a millimeter of error for best results. CT cone beam, real-time navigation, and infrared position sensors are potential options for success but rarely present in veterinary radiation centers. A neuronavigation system (Brainsight Vet, Rogue Research) was tested 22 times on a skull for positioning accuracy and precision analysis. The first 6 manipulations allowed the authors to become familiar with the system but were still included in the analyses. Overall, the targeting mean error in 3D was 1.437 mm with SD 1.242 mm. This system could be used for positioning for radiation therapy or radiosurgery.