Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures

The effect of the ceiling-mounted radiation shielding on the amount of the scatter radiation was assessed under conditions simulating obese patients for clinically relevant exposure parameters. Measurements were performed in different projections and with different positions of the ceiling-mounted s...

Descripción completa

Detalles Bibliográficos
Autores principales: Sukupova, Lucie, Hlavacek, Ondrej, Vedlich, Daniel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5831950/
https://www.ncbi.nlm.nih.gov/pubmed/29666706
http://dx.doi.org/10.1155/2018/4287973
_version_ 1783303234724560896
author Sukupova, Lucie
Hlavacek, Ondrej
Vedlich, Daniel
author_facet Sukupova, Lucie
Hlavacek, Ondrej
Vedlich, Daniel
author_sort Sukupova, Lucie
collection PubMed
description The effect of the ceiling-mounted radiation shielding on the amount of the scatter radiation was assessed under conditions simulating obese patients for clinically relevant exposure parameters. Measurements were performed in different projections and with different positions of the ceiling-mounted shielding: without shielding; shielding closest to the patient; and shielding closest to the physician performing the procedure. The protection provided by the shielding was assessed for cardiology when the femoral access is used and for radiology when the physician performs the procedure in the abdominal area. The results show that the use of the ceiling-mounted shielding can decrease the dose from the scatter radiation by 95% at the position of the performing physician. In cardiology, the impact is more pronounced when the left oblique projection is used. In radiology, a large decrease was observed for right oblique projections, compared to cardiology. The ceiling-mounted shielding should be placed as close to the physician as possible. The idea of creating the largest radiation shadow by placing the radiation shielding as close to the patient as possible does not provide as effective radiation protection of the operator as it might be thought.
format Online
Article
Text
id pubmed-5831950
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-58319502018-04-17 Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures Sukupova, Lucie Hlavacek, Ondrej Vedlich, Daniel Radiol Res Pract Research Article The effect of the ceiling-mounted radiation shielding on the amount of the scatter radiation was assessed under conditions simulating obese patients for clinically relevant exposure parameters. Measurements were performed in different projections and with different positions of the ceiling-mounted shielding: without shielding; shielding closest to the patient; and shielding closest to the physician performing the procedure. The protection provided by the shielding was assessed for cardiology when the femoral access is used and for radiology when the physician performs the procedure in the abdominal area. The results show that the use of the ceiling-mounted shielding can decrease the dose from the scatter radiation by 95% at the position of the performing physician. In cardiology, the impact is more pronounced when the left oblique projection is used. In radiology, a large decrease was observed for right oblique projections, compared to cardiology. The ceiling-mounted shielding should be placed as close to the physician as possible. The idea of creating the largest radiation shadow by placing the radiation shielding as close to the patient as possible does not provide as effective radiation protection of the operator as it might be thought. Hindawi 2018-01-30 /pmc/articles/PMC5831950/ /pubmed/29666706 http://dx.doi.org/10.1155/2018/4287973 Text en Copyright © 2018 Lucie Sukupova et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Sukupova, Lucie
Hlavacek, Ondrej
Vedlich, Daniel
Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures
title Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures
title_full Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures
title_fullStr Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures
title_full_unstemmed Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures
title_short Impact of the Ceiling-Mounted Radiation Shielding Position on the Physician's Dose from Scatter Radiation during Interventional Procedures
title_sort impact of the ceiling-mounted radiation shielding position on the physician's dose from scatter radiation during interventional procedures
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5831950/
https://www.ncbi.nlm.nih.gov/pubmed/29666706
http://dx.doi.org/10.1155/2018/4287973
work_keys_str_mv AT sukupovalucie impactoftheceilingmountedradiationshieldingpositiononthephysiciansdosefromscatterradiationduringinterventionalprocedures
AT hlavacekondrej impactoftheceilingmountedradiationshieldingpositiononthephysiciansdosefromscatterradiationduringinterventionalprocedures
AT vedlichdaniel impactoftheceilingmountedradiationshieldingpositiononthephysiciansdosefromscatterradiationduringinterventionalprocedures