Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects
BACKGROUND: Quantification and visualisation of left ventricular (LV) blood flow is afforded by three-dimensional, time resolved phase contrast cardiovascular magnetic resonance (CMR 4D flow). However, few data exist upon the repeatability and variability of these parameters in a healthy population....
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5833126/ https://www.ncbi.nlm.nih.gov/pubmed/29499706 http://dx.doi.org/10.1186/s12968-018-0432-4 |
_version_ | 1783303432920104960 |
---|---|
author | Stoll, Victoria M. Loudon, Margaret Eriksson, Jonatan Bissell, Malenka M. Dyverfeldt, Petter Ebbers, Tino Myerson, Saul G. Neubauer, Stefan Carlhäll, Carl- Johan Hess, Aaron T. |
author_facet | Stoll, Victoria M. Loudon, Margaret Eriksson, Jonatan Bissell, Malenka M. Dyverfeldt, Petter Ebbers, Tino Myerson, Saul G. Neubauer, Stefan Carlhäll, Carl- Johan Hess, Aaron T. |
author_sort | Stoll, Victoria M. |
collection | PubMed |
description | BACKGROUND: Quantification and visualisation of left ventricular (LV) blood flow is afforded by three-dimensional, time resolved phase contrast cardiovascular magnetic resonance (CMR 4D flow). However, few data exist upon the repeatability and variability of these parameters in a healthy population. We aimed to assess the repeatability and variability over time of LV 4D CMR flow measurements. METHODS: Forty five controls underwent CMR 4D flow data acquisition. Of these, 10 underwent a second scan within the same visit (scan-rescan), 25 returned for a second visit (interval scan; median interval 52 days, IQR 28–57 days). The LV-end diastolic volume (EDV) was divided into four flow components: 1) Direct flow: inflow that passes directly to ejection; 2) Retained inflow: inflow that enters and resides within the LV; 3) Delayed ejection flow: starts within the LV and is ejected and 4) Residual volume: blood that resides within the LV for > 2 cardiac cycles. Each flow components’ volume was related to the EDV (volume-ratio). The kinetic energy at end-diastole (ED) was measured and divided by the components’ volume. RESULTS: The dominant flow component in all 45 controls was the direct flow (volume ratio 38 ± 4%) followed by the residual volume (30 ± 4%), then delayed ejection flow (16 ± 3%) and retained inflow (16 ± 4%). The kinetic energy at ED for each component was direct flow (7.8 ± 3.0 microJ/ml), retained inflow (4.1 ± 2.0 microJ/ml), delayed ejection flow (6.3 ± 2.3 microJ/ml) and the residual volume (1.2 ± 0.5 microJ/ml). The coefficients of variation for the scan-rescan ranged from 2.5%–9.2% for the flow components’ volume ratio and between 13.5%–17.7% for the kinetic energy. The interval scan results showed higher coefficients of variation with values from 6.2–16.1% for the flow components’ volume ratio and 16.9–29.0% for the kinetic energy of the flow components. CONCLUSION: LV flow components’ volume and their associated kinetic energy values are repeatable and stable within a population over time. However, the variability of these measurements in individuals over time is greater than can be attributed to sources of error in the data acquisition and analysis, suggesting that additional physiological factors may influence LV flow measurements. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s12968-018-0432-4) contains supplementary material, which is available to authorized users. |
format | Online Article Text |
id | pubmed-5833126 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-58331262018-03-05 Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects Stoll, Victoria M. Loudon, Margaret Eriksson, Jonatan Bissell, Malenka M. Dyverfeldt, Petter Ebbers, Tino Myerson, Saul G. Neubauer, Stefan Carlhäll, Carl- Johan Hess, Aaron T. J Cardiovasc Magn Reson Research BACKGROUND: Quantification and visualisation of left ventricular (LV) blood flow is afforded by three-dimensional, time resolved phase contrast cardiovascular magnetic resonance (CMR 4D flow). However, few data exist upon the repeatability and variability of these parameters in a healthy population. We aimed to assess the repeatability and variability over time of LV 4D CMR flow measurements. METHODS: Forty five controls underwent CMR 4D flow data acquisition. Of these, 10 underwent a second scan within the same visit (scan-rescan), 25 returned for a second visit (interval scan; median interval 52 days, IQR 28–57 days). The LV-end diastolic volume (EDV) was divided into four flow components: 1) Direct flow: inflow that passes directly to ejection; 2) Retained inflow: inflow that enters and resides within the LV; 3) Delayed ejection flow: starts within the LV and is ejected and 4) Residual volume: blood that resides within the LV for > 2 cardiac cycles. Each flow components’ volume was related to the EDV (volume-ratio). The kinetic energy at end-diastole (ED) was measured and divided by the components’ volume. RESULTS: The dominant flow component in all 45 controls was the direct flow (volume ratio 38 ± 4%) followed by the residual volume (30 ± 4%), then delayed ejection flow (16 ± 3%) and retained inflow (16 ± 4%). The kinetic energy at ED for each component was direct flow (7.8 ± 3.0 microJ/ml), retained inflow (4.1 ± 2.0 microJ/ml), delayed ejection flow (6.3 ± 2.3 microJ/ml) and the residual volume (1.2 ± 0.5 microJ/ml). The coefficients of variation for the scan-rescan ranged from 2.5%–9.2% for the flow components’ volume ratio and between 13.5%–17.7% for the kinetic energy. The interval scan results showed higher coefficients of variation with values from 6.2–16.1% for the flow components’ volume ratio and 16.9–29.0% for the kinetic energy of the flow components. CONCLUSION: LV flow components’ volume and their associated kinetic energy values are repeatable and stable within a population over time. However, the variability of these measurements in individuals over time is greater than can be attributed to sources of error in the data acquisition and analysis, suggesting that additional physiological factors may influence LV flow measurements. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1186/s12968-018-0432-4) contains supplementary material, which is available to authorized users. BioMed Central 2018-03-02 /pmc/articles/PMC5833126/ /pubmed/29499706 http://dx.doi.org/10.1186/s12968-018-0432-4 Text en © The Author(s). 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Stoll, Victoria M. Loudon, Margaret Eriksson, Jonatan Bissell, Malenka M. Dyverfeldt, Petter Ebbers, Tino Myerson, Saul G. Neubauer, Stefan Carlhäll, Carl- Johan Hess, Aaron T. Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects |
title | Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects |
title_full | Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects |
title_fullStr | Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects |
title_full_unstemmed | Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects |
title_short | Test-retest variability of left ventricular 4D flow cardiovascular magnetic resonance measurements in healthy subjects |
title_sort | test-retest variability of left ventricular 4d flow cardiovascular magnetic resonance measurements in healthy subjects |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5833126/ https://www.ncbi.nlm.nih.gov/pubmed/29499706 http://dx.doi.org/10.1186/s12968-018-0432-4 |
work_keys_str_mv | AT stollvictoriam testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT loudonmargaret testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT erikssonjonatan testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT bissellmalenkam testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT dyverfeldtpetter testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT ebberstino testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT myersonsaulg testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT neubauerstefan testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT carlhallcarljohan testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects AT hessaaront testretestvariabilityofleftventricular4dflowcardiovascularmagneticresonancemeasurementsinhealthysubjects |