Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique
BACKGROUND: Linac-based stereotactic radiosurgery or fractionated stereotactic radiotherapy (SRS/FSRT) of multiple brain lesions using volumetric modulated arc therapy (VMAT) is typically performed by a multiple-isocenter approach, i.e. one isocenter per lesion, which is time-demanding for the need...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5836328/ https://www.ncbi.nlm.nih.gov/pubmed/29506539 http://dx.doi.org/10.1186/s13014-018-0985-2 |
_version_ | 1783303942294208512 |
---|---|
author | Ruggieri, Ruggero Naccarato, Stefania Mazzola, Rosario Ricchetti, Francesco Corradini, Stefanie Fiorentino, Alba Alongi, Filippo |
author_facet | Ruggieri, Ruggero Naccarato, Stefania Mazzola, Rosario Ricchetti, Francesco Corradini, Stefanie Fiorentino, Alba Alongi, Filippo |
author_sort | Ruggieri, Ruggero |
collection | PubMed |
description | BACKGROUND: Linac-based stereotactic radiosurgery or fractionated stereotactic radiotherapy (SRS/FSRT) of multiple brain lesions using volumetric modulated arc therapy (VMAT) is typically performed by a multiple-isocenter approach, i.e. one isocenter per lesion, which is time-demanding for the need of independent setup verifications of each isocenter. Here, we present our initial experience with a new dedicated mono-isocenter technique with multiple non-coplanar arcs (HyperArc™, Varian Inc.) in terms of a plan comparison with a multiple-isocenter VMAT approach. METHODS: From August 2017 to October 2017, 20 patients with multiple brain metastases (mean 5, range 2–10) have been treated by HyperArc in 1–3 fractions. The prescribed doses (Dp) were 18–25 Gy in single-fraction, and 21–27 Gy in three-fractions. Planning Target Volume (PTV), defined by a 2 mm isotropic margin from each lesion, had mean dimension of 9.6 cm(3) (range 0.5–27.9 cm(3)). Mono-isocenter HyperArc VMAT plans (HA) with 5 non-coplanar 180°-arcs (couch at 0°, ±45°, ±90°) were generated and compared to multiple-isocenter VMAT plans (RA) with 2 coplanar 360°-arcs per isocenter. A dose normalization of 100%Dp at 98%PTV was adopted, while D(2%)(PTV) < 150%D(p) was accepted. All plans had to respect the constraints on maximum dose to the brainstem (D(0.5cm3) < 18 Gy) as well as to the optical nerves/chiasm, eyes and lenses (D(0.5cm3) < 15 Gy). HA and RA plans were compared in terms of dose-volume metrics, by Paddick conformity (CI) and gradient (GI) index and by V(12) and mean dose to the brain-minus-PTV, and in terms of MU and overall treatment time (OTT) per fraction. OTT was measured for HA treatments, whereas for RA plans OTT was estimated by assuming 3 min. For initial patient setup plus 5 min. For each CBCT-guided setup correction per isocenter. RESULTS: Significant variations in favour of HA plans were computed for both target dose indexes, CI (p < .01) and GI (p < .01). The lower GI in HA plans was the likely cause of the significant reduction in V(12) to the brain-minus-PTV (p = .023). Although at low doses, below 2–5 Gy, the sparing of the brain-minus-PTV was in favour of RA plans, no significant difference in terms of mean doses to the brain-minus-PTV was observed between the two groups (p = .31). Finally, both MU (p < .01) and OTT (p < .01) were significantly reduced by HyperArc plans. CONCLUSIONS: For linac-based SRS/FSRT of multiple brain lesions, HyperArc plans assured a higher CI and a lower GI than standard multiple-isocenter VMAT plans. This is consistent with the computed reduction in V(12) to the brain-minus-PTV. Finally, HyperArc treatments were completed within a typical 20 min. time slot, with a significant time reduction with respect to the expected duration of multiple-isocenters VMAT. |
format | Online Article Text |
id | pubmed-5836328 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-58363282018-03-07 Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique Ruggieri, Ruggero Naccarato, Stefania Mazzola, Rosario Ricchetti, Francesco Corradini, Stefanie Fiorentino, Alba Alongi, Filippo Radiat Oncol Research BACKGROUND: Linac-based stereotactic radiosurgery or fractionated stereotactic radiotherapy (SRS/FSRT) of multiple brain lesions using volumetric modulated arc therapy (VMAT) is typically performed by a multiple-isocenter approach, i.e. one isocenter per lesion, which is time-demanding for the need of independent setup verifications of each isocenter. Here, we present our initial experience with a new dedicated mono-isocenter technique with multiple non-coplanar arcs (HyperArc™, Varian Inc.) in terms of a plan comparison with a multiple-isocenter VMAT approach. METHODS: From August 2017 to October 2017, 20 patients with multiple brain metastases (mean 5, range 2–10) have been treated by HyperArc in 1–3 fractions. The prescribed doses (Dp) were 18–25 Gy in single-fraction, and 21–27 Gy in three-fractions. Planning Target Volume (PTV), defined by a 2 mm isotropic margin from each lesion, had mean dimension of 9.6 cm(3) (range 0.5–27.9 cm(3)). Mono-isocenter HyperArc VMAT plans (HA) with 5 non-coplanar 180°-arcs (couch at 0°, ±45°, ±90°) were generated and compared to multiple-isocenter VMAT plans (RA) with 2 coplanar 360°-arcs per isocenter. A dose normalization of 100%Dp at 98%PTV was adopted, while D(2%)(PTV) < 150%D(p) was accepted. All plans had to respect the constraints on maximum dose to the brainstem (D(0.5cm3) < 18 Gy) as well as to the optical nerves/chiasm, eyes and lenses (D(0.5cm3) < 15 Gy). HA and RA plans were compared in terms of dose-volume metrics, by Paddick conformity (CI) and gradient (GI) index and by V(12) and mean dose to the brain-minus-PTV, and in terms of MU and overall treatment time (OTT) per fraction. OTT was measured for HA treatments, whereas for RA plans OTT was estimated by assuming 3 min. For initial patient setup plus 5 min. For each CBCT-guided setup correction per isocenter. RESULTS: Significant variations in favour of HA plans were computed for both target dose indexes, CI (p < .01) and GI (p < .01). The lower GI in HA plans was the likely cause of the significant reduction in V(12) to the brain-minus-PTV (p = .023). Although at low doses, below 2–5 Gy, the sparing of the brain-minus-PTV was in favour of RA plans, no significant difference in terms of mean doses to the brain-minus-PTV was observed between the two groups (p = .31). Finally, both MU (p < .01) and OTT (p < .01) were significantly reduced by HyperArc plans. CONCLUSIONS: For linac-based SRS/FSRT of multiple brain lesions, HyperArc plans assured a higher CI and a lower GI than standard multiple-isocenter VMAT plans. This is consistent with the computed reduction in V(12) to the brain-minus-PTV. Finally, HyperArc treatments were completed within a typical 20 min. time slot, with a significant time reduction with respect to the expected duration of multiple-isocenters VMAT. BioMed Central 2018-03-05 /pmc/articles/PMC5836328/ /pubmed/29506539 http://dx.doi.org/10.1186/s13014-018-0985-2 Text en © The Author(s). 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Ruggieri, Ruggero Naccarato, Stefania Mazzola, Rosario Ricchetti, Francesco Corradini, Stefanie Fiorentino, Alba Alongi, Filippo Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique |
title | Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique |
title_full | Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique |
title_fullStr | Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique |
title_full_unstemmed | Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique |
title_short | Linac-based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique |
title_sort | linac-based vmat radiosurgery for multiple brain lesions: comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5836328/ https://www.ncbi.nlm.nih.gov/pubmed/29506539 http://dx.doi.org/10.1186/s13014-018-0985-2 |
work_keys_str_mv | AT ruggieriruggero linacbasedvmatradiosurgeryformultiplebrainlesionscomparisonbetweenaconventionalmultiisocenterapproachandanewdedicatedmonoisocentertechnique AT naccaratostefania linacbasedvmatradiosurgeryformultiplebrainlesionscomparisonbetweenaconventionalmultiisocenterapproachandanewdedicatedmonoisocentertechnique AT mazzolarosario linacbasedvmatradiosurgeryformultiplebrainlesionscomparisonbetweenaconventionalmultiisocenterapproachandanewdedicatedmonoisocentertechnique AT ricchettifrancesco linacbasedvmatradiosurgeryformultiplebrainlesionscomparisonbetweenaconventionalmultiisocenterapproachandanewdedicatedmonoisocentertechnique AT corradinistefanie linacbasedvmatradiosurgeryformultiplebrainlesionscomparisonbetweenaconventionalmultiisocenterapproachandanewdedicatedmonoisocentertechnique AT fiorentinoalba linacbasedvmatradiosurgeryformultiplebrainlesionscomparisonbetweenaconventionalmultiisocenterapproachandanewdedicatedmonoisocentertechnique AT alongifilippo linacbasedvmatradiosurgeryformultiplebrainlesionscomparisonbetweenaconventionalmultiisocenterapproachandanewdedicatedmonoisocentertechnique |