Characterization of the diffusion coefficient of blood

PURPOSE: To characterize the diffusion coefficient of human blood for accurate results in intravoxel incoherent motion imaging. METHODS: Diffusion‐weighted MRI of blood samples from 10 healthy volunteers was acquired with a single‐shot echo‐planar‐imaging sequence at body temperature. Effects of gra...

Descripción completa

Detalles Bibliográficos
Autores principales: Funck, Carsten, Laun, Frederik Bernd, Wetscherek, Andreas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5836916/
https://www.ncbi.nlm.nih.gov/pubmed/28940621
http://dx.doi.org/10.1002/mrm.26919
Descripción
Sumario:PURPOSE: To characterize the diffusion coefficient of human blood for accurate results in intravoxel incoherent motion imaging. METHODS: Diffusion‐weighted MRI of blood samples from 10 healthy volunteers was acquired with a single‐shot echo‐planar‐imaging sequence at body temperature. Effects of gradient profile (monopolar or flow‐compensated), diffusion time (40–100 ms), and echo time (60–200 ms) were investigated. RESULTS: Although measured apparent diffusion coefficients of blood were larger for flow‐compensated than for monopolar gradients, no dependence of the apparent diffusion coefficient on the diffusion time was found. Large differences between individual samples were observed, with results ranging from 1.26 to 1.66 µm(2)/ms for flow‐compensated and 0.94 to 1.52 µm(2)/ms for monopolar gradients. Statistical analysis indicates correlations of the flow‐compensated apparent diffusion coefficient with hematocrit (P = 0.007) and hemoglobin (P = 0.017), but not with mean corpuscular volume (P = 0.64). Results of Monte‐Carlo simulations support the experimental observations. CONCLUSIONS: Measured blood apparent diffusion coefficient values depend on hematocrit/hemoglobin concentration and applied gradient profile due to non‐Gaussian diffusion. Because in vivo measurement is delicate, an estimation based on blood count results could be an alternative. For intravoxel incoherent motion modeling, the use of a blood self‐diffusion constant D(b) = 1.54 ± 0.12 µm(2)/ms for flow‐compensated and D(b) = 1.30 ± 0.18 µm(2)/ms for monopolar encoding is suggested. Magn Reson Med 79:2752–2758, 2018. © 2017 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.