Cargando…
Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit
For passive scattering proton therapy systems, neutron contamination is the main concern both from an occupational and patient safety perspective. The Mevion S250 compact proton therapy system is the first of its kind, offering an in‐room cyclotron design which prompts more concern for shielding ass...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5849833/ https://www.ncbi.nlm.nih.gov/pubmed/29468842 http://dx.doi.org/10.1002/acm2.12256 |
_version_ | 1783306114848260096 |
---|---|
author | Prusator, Michael T. Ahmad, Salahuddin Chen, Yong |
author_facet | Prusator, Michael T. Ahmad, Salahuddin Chen, Yong |
author_sort | Prusator, Michael T. |
collection | PubMed |
description | For passive scattering proton therapy systems, neutron contamination is the main concern both from an occupational and patient safety perspective. The Mevion S250 compact proton therapy system is the first of its kind, offering an in‐room cyclotron design which prompts more concern for shielding assessment. The purpose of this study was to accomplish an in‐depth evaluation of both the shielding design and in‐room neutron production at our facility using both Monte Carlo simulation and measurement. We found that the shielding in place at our facility is adequate, with simulated annual neutron ambient dose equivalents at 30 cm outside wall/door perimeter ranging from background to 0.07 mSv and measured dose equivalents ranging from background to 0.06 mSv. The in‐room measurements reveal that the H*/D decreases when the distance from isocenter and field size increases. Furthermore, the H*/D generally increases when the angle around isocenter increases. Our results from in‐room measurements show consistent trends with our Monte Carlo model of the Mevion system. |
format | Online Article Text |
id | pubmed-5849833 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-58498332018-04-02 Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit Prusator, Michael T. Ahmad, Salahuddin Chen, Yong J Appl Clin Med Phys Radiation Protection & Regulations For passive scattering proton therapy systems, neutron contamination is the main concern both from an occupational and patient safety perspective. The Mevion S250 compact proton therapy system is the first of its kind, offering an in‐room cyclotron design which prompts more concern for shielding assessment. The purpose of this study was to accomplish an in‐depth evaluation of both the shielding design and in‐room neutron production at our facility using both Monte Carlo simulation and measurement. We found that the shielding in place at our facility is adequate, with simulated annual neutron ambient dose equivalents at 30 cm outside wall/door perimeter ranging from background to 0.07 mSv and measured dose equivalents ranging from background to 0.06 mSv. The in‐room measurements reveal that the H*/D decreases when the distance from isocenter and field size increases. Furthermore, the H*/D generally increases when the angle around isocenter increases. Our results from in‐room measurements show consistent trends with our Monte Carlo model of the Mevion system. John Wiley and Sons Inc. 2018-02-22 /pmc/articles/PMC5849833/ /pubmed/29468842 http://dx.doi.org/10.1002/acm2.12256 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Protection & Regulations Prusator, Michael T. Ahmad, Salahuddin Chen, Yong Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit |
title | Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit |
title_full | Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit |
title_fullStr | Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit |
title_full_unstemmed | Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit |
title_short | Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit |
title_sort | shielding verification and neutron dose evaluation of the mevion s250 proton therapy unit |
topic | Radiation Protection & Regulations |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5849833/ https://www.ncbi.nlm.nih.gov/pubmed/29468842 http://dx.doi.org/10.1002/acm2.12256 |
work_keys_str_mv | AT prusatormichaelt shieldingverificationandneutrondoseevaluationofthemevions250protontherapyunit AT ahmadsalahuddin shieldingverificationandneutrondoseevaluationofthemevions250protontherapyunit AT chenyong shieldingverificationandneutrondoseevaluationofthemevions250protontherapyunit |