On the tenth value distance of the photon field along the maze of high‐energy linear accelerator vaults
There is a wide range in the reported photon tenth value distance (TVD) in the maze of high‐energy linear accelerator vaults. In order to gain insight into the appropriate use of the TVD value during door design, we performed measurements of the photon dose in the maze of four vaults. In addition, o...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5849845/ https://www.ncbi.nlm.nih.gov/pubmed/29374424 http://dx.doi.org/10.1002/acm2.12250 |
Sumario: | There is a wide range in the reported photon tenth value distance (TVD) in the maze of high‐energy linear accelerator vaults. In order to gain insight into the appropriate use of the TVD value during door design, we performed measurements of the photon dose in the maze of four vaults. In addition, our study represents the first to describe a scenario where an inner borated polyethylene (BPE) door for neutron shielding is installed in the maze downstream to Point A, the point on the maze centerline that is just visible from the isocenter. The measurements were made along the maze centerline at 1 m above the floor. In all cases, the accelerator operated at a nominal energy of 15 MV. Of the four vaults, three were equipped with an inner BPE door at a distance of 1.0–2.1 m downstream to Point A. The door was made of 10.16 cm (4″) BPE sandwiched between two 0.635 cm (1/4″) steel face plates. The photon dose in the maze without a BPE door decreases exponentially with a characteristic TVD of 6 m beyond a distance of 2.5 m from Point A. The presence of a BPE door in an identical vault not only reduces the photon intensity in the maze by about an order of magnitude, but also softens the energy spectrum with a shortened TVD of 4.7 m, significantly lessening the shielding burden at the outer maze entrance. In contrast to the common use of Point A as the reference point to specify distance, the photon dose in the maze with a BPE door located downstream to Point A can be satisfactorily described as exponential functions of the distance measured from the door, which shows good consistency among the three vaults of different room parameters. |
---|