Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen

BACKGROUND: This study examines the utility of the first prefilled, rapid-acting insulin pen that can be dialed in half-unit increments. Dose accuracy and injection force were examined through a series of design-verification tests, and usability was established by human factors validation testing. M...

Descripción completa

Detalles Bibliográficos
Autores principales: Kappes, Christopher M., Kershner, James R., Morwick, Tina M., Corrigan, Sheila M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5851229/
https://www.ncbi.nlm.nih.gov/pubmed/29084452
http://dx.doi.org/10.1177/1932296817736316
_version_ 1783306357734113280
author Kappes, Christopher M.
Kershner, James R.
Morwick, Tina M.
Corrigan, Sheila M.
author_facet Kappes, Christopher M.
Kershner, James R.
Morwick, Tina M.
Corrigan, Sheila M.
author_sort Kappes, Christopher M.
collection PubMed
description BACKGROUND: This study examines the utility of the first prefilled, rapid-acting insulin pen that can be dialed in half-unit increments. Dose accuracy and injection force were examined through a series of design-verification tests, and usability was established by human factors validation testing. METHODS: Devices were tested for dose accuracy at 3 different doses and temperatures and under free fall, vibration, and cold storage conditioning. Injection force was measured at the maximum dose (30 units). Both experiments used the same semiautomated testing system. Usability was validated in a human factors simulated-use study that included 60 participants (patients with type 1 or type 2 diabetes [aged 10-79 years], adult caregivers, and health care providers). RESULTS: The pen met the International Organization for Standardization (ISO) 11608-1:2014 requirements for dose accuracy at all settings and conditions tested. Furthermore, all individual results were within the ISO specification limits. Mean injection force across temperature settings ranged from 9.25 to 10.85 N at the highest dose. The usability validation study confirmed that use-related risks were reduced to the extent possible and that additional modifications were not likely to afford further reductions. CONCLUSIONS: The results from these studies demonstrated accurate dosing over the dose range (0.5-30 units) at different temperatures and conditions with an injection force that should accommodate the intended users. Use safety and usability in patients with diabetes, caregivers, and health care professionals were validated. The added convenience of this new half-unit, prefilled pen may ease the burden of diabetes management for patients who require smaller incremental dosing.
format Online
Article
Text
id pubmed-5851229
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-58512292018-10-30 Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen Kappes, Christopher M. Kershner, James R. Morwick, Tina M. Corrigan, Sheila M. J Diabetes Sci Technol Original Articles BACKGROUND: This study examines the utility of the first prefilled, rapid-acting insulin pen that can be dialed in half-unit increments. Dose accuracy and injection force were examined through a series of design-verification tests, and usability was established by human factors validation testing. METHODS: Devices were tested for dose accuracy at 3 different doses and temperatures and under free fall, vibration, and cold storage conditioning. Injection force was measured at the maximum dose (30 units). Both experiments used the same semiautomated testing system. Usability was validated in a human factors simulated-use study that included 60 participants (patients with type 1 or type 2 diabetes [aged 10-79 years], adult caregivers, and health care providers). RESULTS: The pen met the International Organization for Standardization (ISO) 11608-1:2014 requirements for dose accuracy at all settings and conditions tested. Furthermore, all individual results were within the ISO specification limits. Mean injection force across temperature settings ranged from 9.25 to 10.85 N at the highest dose. The usability validation study confirmed that use-related risks were reduced to the extent possible and that additional modifications were not likely to afford further reductions. CONCLUSIONS: The results from these studies demonstrated accurate dosing over the dose range (0.5-30 units) at different temperatures and conditions with an injection force that should accommodate the intended users. Use safety and usability in patients with diabetes, caregivers, and health care professionals were validated. The added convenience of this new half-unit, prefilled pen may ease the burden of diabetes management for patients who require smaller incremental dosing. SAGE Publications 2017-10-30 /pmc/articles/PMC5851229/ /pubmed/29084452 http://dx.doi.org/10.1177/1932296817736316 Text en © 2017 Diabetes Technology Society http://www.creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Articles
Kappes, Christopher M.
Kershner, James R.
Morwick, Tina M.
Corrigan, Sheila M.
Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen
title Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen
title_full Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen
title_fullStr Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen
title_full_unstemmed Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen
title_short Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen
title_sort dose accuracy, injection force, and usability assessment of a new half-unit, prefilled insulin pen
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5851229/
https://www.ncbi.nlm.nih.gov/pubmed/29084452
http://dx.doi.org/10.1177/1932296817736316
work_keys_str_mv AT kappeschristopherm doseaccuracyinjectionforceandusabilityassessmentofanewhalfunitprefilledinsulinpen
AT kershnerjamesr doseaccuracyinjectionforceandusabilityassessmentofanewhalfunitprefilledinsulinpen
AT morwicktinam doseaccuracyinjectionforceandusabilityassessmentofanewhalfunitprefilledinsulinpen
AT corrigansheilam doseaccuracyinjectionforceandusabilityassessmentofanewhalfunitprefilledinsulinpen