Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study

BACKGROUND AND OBJECTIVES: Intermittent vital signs measurements are the current standard on hospital wards, typically recorded once every 8 hours. Early signs of deterioration may therefore be missed. Recent innovations have resulted in ‘wearable’ sensors, which may capture patient deterioration at...

Descripción completa

Detalles Bibliográficos
Autores principales: Breteler, Martine J M, Huizinga, Erik, van Loon, Kim, Leenen, Luke P H, Dohmen, Daan A J, Kalkman, Cor J, Blokhuis, Taco J
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5855309/
https://www.ncbi.nlm.nih.gov/pubmed/29487076
http://dx.doi.org/10.1136/bmjopen-2017-020162
_version_ 1783307073665105920
author Breteler, Martine J M
Huizinga, Erik
van Loon, Kim
Leenen, Luke P H
Dohmen, Daan A J
Kalkman, Cor J
Blokhuis, Taco J
author_facet Breteler, Martine J M
Huizinga, Erik
van Loon, Kim
Leenen, Luke P H
Dohmen, Daan A J
Kalkman, Cor J
Blokhuis, Taco J
author_sort Breteler, Martine J M
collection PubMed
description BACKGROUND AND OBJECTIVES: Intermittent vital signs measurements are the current standard on hospital wards, typically recorded once every 8 hours. Early signs of deterioration may therefore be missed. Recent innovations have resulted in ‘wearable’ sensors, which may capture patient deterioration at an earlier stage. The objective of this study was to determine whether a wireless ‘patch’ sensor is able to reliably measure respiratory and heart rate continuously in high-risk surgical patients. The secondary objective was to explore the potential of the wireless sensor to serve as a safety monitor. DESIGN: In an observational methods comparisons study, patients were measured with both the wireless sensor and bedside routine standard for at least 24 hours. SETTING: University teaching hospital, single centre. PARTICIPANTS: Twenty-five postoperative surgical patients admitted to a step-down unit. OUTCOME MEASURES: Primary outcome measures were limits of agreement and bias of heart rate and respiratory rate. Secondary outcome measures were sensor reliability, defined as time until first occurrence of data loss. RESULTS: 1568 hours of vital signs data were analysed. Bias and 95% limits of agreement for heart rate were −1.1 (−8.8 to 6.5) beats per minute. For respiration rate, bias was −2.3 breaths per minute with wide limits of agreement (−15.8 to 11.2 breaths per minute). Median filtering over a 15 min period improved limits of agreement of both respiration and heart rate. 63% of the measurements were performed without data loss greater than 2 min. Overall data loss was limited (6% of time). CONCLUSIONS: The wireless sensor is capable of accurately measuring heart rate, but accuracy for respiratory rate was outside acceptable limits. Remote monitoring has the potential to contribute to early recognition of physiological decline in high-risk patients. Future studies should focus on the ability to detect patient deterioration on low care environments and at home after discharge.
format Online
Article
Text
id pubmed-5855309
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher BMJ Publishing Group
record_format MEDLINE/PubMed
spelling pubmed-58553092018-03-19 Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study Breteler, Martine J M Huizinga, Erik van Loon, Kim Leenen, Luke P H Dohmen, Daan A J Kalkman, Cor J Blokhuis, Taco J BMJ Open Surgery BACKGROUND AND OBJECTIVES: Intermittent vital signs measurements are the current standard on hospital wards, typically recorded once every 8 hours. Early signs of deterioration may therefore be missed. Recent innovations have resulted in ‘wearable’ sensors, which may capture patient deterioration at an earlier stage. The objective of this study was to determine whether a wireless ‘patch’ sensor is able to reliably measure respiratory and heart rate continuously in high-risk surgical patients. The secondary objective was to explore the potential of the wireless sensor to serve as a safety monitor. DESIGN: In an observational methods comparisons study, patients were measured with both the wireless sensor and bedside routine standard for at least 24 hours. SETTING: University teaching hospital, single centre. PARTICIPANTS: Twenty-five postoperative surgical patients admitted to a step-down unit. OUTCOME MEASURES: Primary outcome measures were limits of agreement and bias of heart rate and respiratory rate. Secondary outcome measures were sensor reliability, defined as time until first occurrence of data loss. RESULTS: 1568 hours of vital signs data were analysed. Bias and 95% limits of agreement for heart rate were −1.1 (−8.8 to 6.5) beats per minute. For respiration rate, bias was −2.3 breaths per minute with wide limits of agreement (−15.8 to 11.2 breaths per minute). Median filtering over a 15 min period improved limits of agreement of both respiration and heart rate. 63% of the measurements were performed without data loss greater than 2 min. Overall data loss was limited (6% of time). CONCLUSIONS: The wireless sensor is capable of accurately measuring heart rate, but accuracy for respiratory rate was outside acceptable limits. Remote monitoring has the potential to contribute to early recognition of physiological decline in high-risk patients. Future studies should focus on the ability to detect patient deterioration on low care environments and at home after discharge. BMJ Publishing Group 2018-02-27 /pmc/articles/PMC5855309/ /pubmed/29487076 http://dx.doi.org/10.1136/bmjopen-2017-020162 Text en © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted. This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
spellingShingle Surgery
Breteler, Martine J M
Huizinga, Erik
van Loon, Kim
Leenen, Luke P H
Dohmen, Daan A J
Kalkman, Cor J
Blokhuis, Taco J
Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study
title Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study
title_full Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study
title_fullStr Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study
title_full_unstemmed Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study
title_short Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study
title_sort reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the netherlands: a clinical validation study
topic Surgery
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5855309/
https://www.ncbi.nlm.nih.gov/pubmed/29487076
http://dx.doi.org/10.1136/bmjopen-2017-020162
work_keys_str_mv AT bretelermartinejm reliabilityofwirelessmonitoringusingawearablepatchsensorinhighrisksurgicalpatientsatastepdownunitinthenetherlandsaclinicalvalidationstudy
AT huizingaerik reliabilityofwirelessmonitoringusingawearablepatchsensorinhighrisksurgicalpatientsatastepdownunitinthenetherlandsaclinicalvalidationstudy
AT vanloonkim reliabilityofwirelessmonitoringusingawearablepatchsensorinhighrisksurgicalpatientsatastepdownunitinthenetherlandsaclinicalvalidationstudy
AT leenenlukeph reliabilityofwirelessmonitoringusingawearablepatchsensorinhighrisksurgicalpatientsatastepdownunitinthenetherlandsaclinicalvalidationstudy
AT dohmendaanaj reliabilityofwirelessmonitoringusingawearablepatchsensorinhighrisksurgicalpatientsatastepdownunitinthenetherlandsaclinicalvalidationstudy
AT kalkmancorj reliabilityofwirelessmonitoringusingawearablepatchsensorinhighrisksurgicalpatientsatastepdownunitinthenetherlandsaclinicalvalidationstudy
AT blokhuistacoj reliabilityofwirelessmonitoringusingawearablepatchsensorinhighrisksurgicalpatientsatastepdownunitinthenetherlandsaclinicalvalidationstudy