Real-Time and Meter-Scale Absolute Distance Measurement by Frequency-Comb-Referenced Multi-Wavelength Interferometry
We report on a frequency-comb-referenced absolute interferometer which instantly measures long distance by integrating multi-wavelength interferometry with direct synthetic wavelength interferometry. The reported interferometer utilizes four different wavelengths, simultaneously calibrated to the fr...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5855924/ https://www.ncbi.nlm.nih.gov/pubmed/29414897 http://dx.doi.org/10.3390/s18020500 |
_version_ | 1783307211277074432 |
---|---|
author | Wang, Guochao Tan, Lilong Yan, Shuhua |
author_facet | Wang, Guochao Tan, Lilong Yan, Shuhua |
author_sort | Wang, Guochao |
collection | PubMed |
description | We report on a frequency-comb-referenced absolute interferometer which instantly measures long distance by integrating multi-wavelength interferometry with direct synthetic wavelength interferometry. The reported interferometer utilizes four different wavelengths, simultaneously calibrated to the frequency comb of a femtosecond laser, to implement subwavelength distance measurement, while direct synthetic wavelength interferometry is elaborately introduced by launching a fifth wavelength to extend a non-ambiguous range for meter-scale measurement. A linearity test performed comparatively with a He–Ne laser interferometer shows a residual error of less than 70.8 nm in peak-to-valley over a 3 m distance, and a 10 h distance comparison is demonstrated to gain fractional deviations of ~3 × 10(−8) versus 3 m distance. Test results reveal that the presented absolute interferometer enables precise, stable, and long-term distance measurements and facilitates absolute positioning applications such as large-scale manufacturing and space missions. |
format | Online Article Text |
id | pubmed-5855924 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-58559242018-03-20 Real-Time and Meter-Scale Absolute Distance Measurement by Frequency-Comb-Referenced Multi-Wavelength Interferometry Wang, Guochao Tan, Lilong Yan, Shuhua Sensors (Basel) Article We report on a frequency-comb-referenced absolute interferometer which instantly measures long distance by integrating multi-wavelength interferometry with direct synthetic wavelength interferometry. The reported interferometer utilizes four different wavelengths, simultaneously calibrated to the frequency comb of a femtosecond laser, to implement subwavelength distance measurement, while direct synthetic wavelength interferometry is elaborately introduced by launching a fifth wavelength to extend a non-ambiguous range for meter-scale measurement. A linearity test performed comparatively with a He–Ne laser interferometer shows a residual error of less than 70.8 nm in peak-to-valley over a 3 m distance, and a 10 h distance comparison is demonstrated to gain fractional deviations of ~3 × 10(−8) versus 3 m distance. Test results reveal that the presented absolute interferometer enables precise, stable, and long-term distance measurements and facilitates absolute positioning applications such as large-scale manufacturing and space missions. MDPI 2018-02-07 /pmc/articles/PMC5855924/ /pubmed/29414897 http://dx.doi.org/10.3390/s18020500 Text en © 2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Wang, Guochao Tan, Lilong Yan, Shuhua Real-Time and Meter-Scale Absolute Distance Measurement by Frequency-Comb-Referenced Multi-Wavelength Interferometry |
title | Real-Time and Meter-Scale Absolute Distance Measurement by Frequency-Comb-Referenced Multi-Wavelength Interferometry |
title_full | Real-Time and Meter-Scale Absolute Distance Measurement by Frequency-Comb-Referenced Multi-Wavelength Interferometry |
title_fullStr | Real-Time and Meter-Scale Absolute Distance Measurement by Frequency-Comb-Referenced Multi-Wavelength Interferometry |
title_full_unstemmed | Real-Time and Meter-Scale Absolute Distance Measurement by Frequency-Comb-Referenced Multi-Wavelength Interferometry |
title_short | Real-Time and Meter-Scale Absolute Distance Measurement by Frequency-Comb-Referenced Multi-Wavelength Interferometry |
title_sort | real-time and meter-scale absolute distance measurement by frequency-comb-referenced multi-wavelength interferometry |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5855924/ https://www.ncbi.nlm.nih.gov/pubmed/29414897 http://dx.doi.org/10.3390/s18020500 |
work_keys_str_mv | AT wangguochao realtimeandmeterscaleabsolutedistancemeasurementbyfrequencycombreferencedmultiwavelengthinterferometry AT tanlilong realtimeandmeterscaleabsolutedistancemeasurementbyfrequencycombreferencedmultiwavelengthinterferometry AT yanshuhua realtimeandmeterscaleabsolutedistancemeasurementbyfrequencycombreferencedmultiwavelengthinterferometry |