Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans

PURPOSE: This study aimed to evaluate the effect of field-of-view (FOV) size on the gray values derived from conebeam computed tomography (CBCT) compared with the Hounsfield unit values from multidetector computed tomography (MDCT) scans as the gold standard. MATERIALS AND METHODS: A radiographic ph...

Descripción completa

Detalles Bibliográficos
Autores principales: Shokri, Abbas, Ramezani, Leila, Bidgoli, Mohsen, Akbarzadeh, Mahdi, Ghazikhanlu-Sani, Karim, Fallahi-Sichani, Hamed
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Academy of Oral and Maxillofacial Radiology 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5863017/
https://www.ncbi.nlm.nih.gov/pubmed/29581947
http://dx.doi.org/10.5624/isd.2018.48.1.31
_version_ 1783308333023756288
author Shokri, Abbas
Ramezani, Leila
Bidgoli, Mohsen
Akbarzadeh, Mahdi
Ghazikhanlu-Sani, Karim
Fallahi-Sichani, Hamed
author_facet Shokri, Abbas
Ramezani, Leila
Bidgoli, Mohsen
Akbarzadeh, Mahdi
Ghazikhanlu-Sani, Karim
Fallahi-Sichani, Hamed
author_sort Shokri, Abbas
collection PubMed
description PURPOSE: This study aimed to evaluate the effect of field-of-view (FOV) size on the gray values derived from conebeam computed tomography (CBCT) compared with the Hounsfield unit values from multidetector computed tomography (MDCT) scans as the gold standard. MATERIALS AND METHODS: A radiographic phantom was designed with 4 acrylic cylinders. One cylinder was filled with distilled water, and the other 3 were filled with 3 types of bone substitute: namely, Nanobone, Cenobone, and Cerabone. The phantom was scanned with 2 CBCT systems using 2 different FOV sizes, and 1 MDCT system was used as the gold standard. The mean gray values (MGVs) of each cylinder were calculated in each imaging protocol. RESULTS: In both CBCT systems, significant differences were noted in the MGVs of all materials between the 2 FOV sizes (P<.05) except for Cerabone in the Cranex3D system. Significant differences were found in the MGVs of each material compared with the others in both FOV sizes for each CBCT system. No significant difference was seen between the Cranex3D CBCT system and the MDCT system in the MGVs of bone substitutes on images obtained with a small FOV. CONCLUSION: The size of the FOV significantly changed the MGVs of all bone substitutes, except for Cerabone in the Cranex3D system. Both CBCT systems had the ability to distinguish the 3 types of bone substitutes based on a comparison of their MGVs. The Cranex3D CBCT system used with a small FOV had a significant correlation with MDCT results.
format Online
Article
Text
id pubmed-5863017
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Korean Academy of Oral and Maxillofacial Radiology
record_format MEDLINE/PubMed
spelling pubmed-58630172018-03-26 Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans Shokri, Abbas Ramezani, Leila Bidgoli, Mohsen Akbarzadeh, Mahdi Ghazikhanlu-Sani, Karim Fallahi-Sichani, Hamed Imaging Sci Dent Original Article PURPOSE: This study aimed to evaluate the effect of field-of-view (FOV) size on the gray values derived from conebeam computed tomography (CBCT) compared with the Hounsfield unit values from multidetector computed tomography (MDCT) scans as the gold standard. MATERIALS AND METHODS: A radiographic phantom was designed with 4 acrylic cylinders. One cylinder was filled with distilled water, and the other 3 were filled with 3 types of bone substitute: namely, Nanobone, Cenobone, and Cerabone. The phantom was scanned with 2 CBCT systems using 2 different FOV sizes, and 1 MDCT system was used as the gold standard. The mean gray values (MGVs) of each cylinder were calculated in each imaging protocol. RESULTS: In both CBCT systems, significant differences were noted in the MGVs of all materials between the 2 FOV sizes (P<.05) except for Cerabone in the Cranex3D system. Significant differences were found in the MGVs of each material compared with the others in both FOV sizes for each CBCT system. No significant difference was seen between the Cranex3D CBCT system and the MDCT system in the MGVs of bone substitutes on images obtained with a small FOV. CONCLUSION: The size of the FOV significantly changed the MGVs of all bone substitutes, except for Cerabone in the Cranex3D system. Both CBCT systems had the ability to distinguish the 3 types of bone substitutes based on a comparison of their MGVs. The Cranex3D CBCT system used with a small FOV had a significant correlation with MDCT results. Korean Academy of Oral and Maxillofacial Radiology 2018-03 2018-03-19 /pmc/articles/PMC5863017/ /pubmed/29581947 http://dx.doi.org/10.5624/isd.2018.48.1.31 Text en Copyright © 2018 by Korean Academy of Oral and Maxillofacial Radiology http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Shokri, Abbas
Ramezani, Leila
Bidgoli, Mohsen
Akbarzadeh, Mahdi
Ghazikhanlu-Sani, Karim
Fallahi-Sichani, Hamed
Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans
title Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans
title_full Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans
title_fullStr Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans
title_full_unstemmed Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans
title_short Effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the Hounsfield unit values from multidetector computed tomography scans
title_sort effect of field-of-view size on gray values derived from cone-beam computed tomography compared with the hounsfield unit values from multidetector computed tomography scans
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5863017/
https://www.ncbi.nlm.nih.gov/pubmed/29581947
http://dx.doi.org/10.5624/isd.2018.48.1.31
work_keys_str_mv AT shokriabbas effectoffieldofviewsizeongrayvaluesderivedfromconebeamcomputedtomographycomparedwiththehounsfieldunitvaluesfrommultidetectorcomputedtomographyscans
AT ramezanileila effectoffieldofviewsizeongrayvaluesderivedfromconebeamcomputedtomographycomparedwiththehounsfieldunitvaluesfrommultidetectorcomputedtomographyscans
AT bidgolimohsen effectoffieldofviewsizeongrayvaluesderivedfromconebeamcomputedtomographycomparedwiththehounsfieldunitvaluesfrommultidetectorcomputedtomographyscans
AT akbarzadehmahdi effectoffieldofviewsizeongrayvaluesderivedfromconebeamcomputedtomographycomparedwiththehounsfieldunitvaluesfrommultidetectorcomputedtomographyscans
AT ghazikhanlusanikarim effectoffieldofviewsizeongrayvaluesderivedfromconebeamcomputedtomographycomparedwiththehounsfieldunitvaluesfrommultidetectorcomputedtomographyscans
AT fallahisichanihamed effectoffieldofviewsizeongrayvaluesderivedfromconebeamcomputedtomographycomparedwiththehounsfieldunitvaluesfrommultidetectorcomputedtomographyscans