Radiation Reduction in Low Dose Pulsed Fluoroscopy versus Standard Dose Continuous Fluoroscopy during Fluoroscopically-Guided Lumbar Punctures: A Prospective Controlled Study

PURPOSE: The purpose of this study was to evaluate radiation dose reduction in fluoroscopically guided lumbar punctures (FGLP) using “pulsed fluoroscopy in a low dose mode” compared with the commonly used “continuous fluoroscopy in a standard dose mode” while maintaining the technical success. MATER...

Descripción completa

Detalles Bibliográficos
Autores principales: Sabat, Shyam, Slonimsky, Einat
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5868228/
https://www.ncbi.nlm.nih.gov/pubmed/29619280
http://dx.doi.org/10.4103/jcis.JCIS_94_17
_version_ 1783309117438296064
author Sabat, Shyam
Slonimsky, Einat
author_facet Sabat, Shyam
Slonimsky, Einat
author_sort Sabat, Shyam
collection PubMed
description PURPOSE: The purpose of this study was to evaluate radiation dose reduction in fluoroscopically guided lumbar punctures (FGLP) using “pulsed fluoroscopy in a low dose mode” compared with the commonly used “continuous fluoroscopy in a standard dose mode” while maintaining the technical success. MATERIALS AND METHODS: Thirty-five consecutive patients who underwent FGLP divided randomly to seventeen patients in the control group with standard dose continuous FGLP and eighteen patients in the study group with low-dose low-frame-rate of 3 frames per second (fps) FGLP. Entrance surface dose measurements from a dosimeter device were recorded as well as the dose area product (DAP). RESULTS: A total of 35 patients with average age of 52 years (range: 15–87 [±17 standard deviation [SD]]) were evaluated. Average entrance surface dose of the study group was significantly lower (3.81 mGy [range: 0.21–11.14, [±2.8 SD]]) compared with the control group (22.45 mGy [range: 1.23–73.44, [±19.41 [SD]]). The average DAP of the study group (10 mGy·cm(2) [range: 1–41, [±9.8 SD]]) was also significantly lower than the control group (65 mGy·cm(2) [range: 5–199, [±53 SD]]). Success rate was similar between the study and control groups. CONCLUSION: Low dose pulsed fluoroscopy of 3 fps significantly reduces radiation exposure by about 600% compared with standard dose continuous fluoroscopy in FGLP. Utilizing this radiation saving strategy will allow to dramatically reduce radiation exposure, without impacting the technical success rate.
format Online
Article
Text
id pubmed-5868228
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-58682282018-04-04 Radiation Reduction in Low Dose Pulsed Fluoroscopy versus Standard Dose Continuous Fluoroscopy during Fluoroscopically-Guided Lumbar Punctures: A Prospective Controlled Study Sabat, Shyam Slonimsky, Einat J Clin Imaging Sci Original Article PURPOSE: The purpose of this study was to evaluate radiation dose reduction in fluoroscopically guided lumbar punctures (FGLP) using “pulsed fluoroscopy in a low dose mode” compared with the commonly used “continuous fluoroscopy in a standard dose mode” while maintaining the technical success. MATERIALS AND METHODS: Thirty-five consecutive patients who underwent FGLP divided randomly to seventeen patients in the control group with standard dose continuous FGLP and eighteen patients in the study group with low-dose low-frame-rate of 3 frames per second (fps) FGLP. Entrance surface dose measurements from a dosimeter device were recorded as well as the dose area product (DAP). RESULTS: A total of 35 patients with average age of 52 years (range: 15–87 [±17 standard deviation [SD]]) were evaluated. Average entrance surface dose of the study group was significantly lower (3.81 mGy [range: 0.21–11.14, [±2.8 SD]]) compared with the control group (22.45 mGy [range: 1.23–73.44, [±19.41 [SD]]). The average DAP of the study group (10 mGy·cm(2) [range: 1–41, [±9.8 SD]]) was also significantly lower than the control group (65 mGy·cm(2) [range: 5–199, [±53 SD]]). Success rate was similar between the study and control groups. CONCLUSION: Low dose pulsed fluoroscopy of 3 fps significantly reduces radiation exposure by about 600% compared with standard dose continuous fluoroscopy in FGLP. Utilizing this radiation saving strategy will allow to dramatically reduce radiation exposure, without impacting the technical success rate. Medknow Publications & Media Pvt Ltd 2018-03-09 /pmc/articles/PMC5868228/ /pubmed/29619280 http://dx.doi.org/10.4103/jcis.JCIS_94_17 Text en Copyright: © 2018 Journal of Clinical Imaging Science http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Article
Sabat, Shyam
Slonimsky, Einat
Radiation Reduction in Low Dose Pulsed Fluoroscopy versus Standard Dose Continuous Fluoroscopy during Fluoroscopically-Guided Lumbar Punctures: A Prospective Controlled Study
title Radiation Reduction in Low Dose Pulsed Fluoroscopy versus Standard Dose Continuous Fluoroscopy during Fluoroscopically-Guided Lumbar Punctures: A Prospective Controlled Study
title_full Radiation Reduction in Low Dose Pulsed Fluoroscopy versus Standard Dose Continuous Fluoroscopy during Fluoroscopically-Guided Lumbar Punctures: A Prospective Controlled Study
title_fullStr Radiation Reduction in Low Dose Pulsed Fluoroscopy versus Standard Dose Continuous Fluoroscopy during Fluoroscopically-Guided Lumbar Punctures: A Prospective Controlled Study
title_full_unstemmed Radiation Reduction in Low Dose Pulsed Fluoroscopy versus Standard Dose Continuous Fluoroscopy during Fluoroscopically-Guided Lumbar Punctures: A Prospective Controlled Study
title_short Radiation Reduction in Low Dose Pulsed Fluoroscopy versus Standard Dose Continuous Fluoroscopy during Fluoroscopically-Guided Lumbar Punctures: A Prospective Controlled Study
title_sort radiation reduction in low dose pulsed fluoroscopy versus standard dose continuous fluoroscopy during fluoroscopically-guided lumbar punctures: a prospective controlled study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5868228/
https://www.ncbi.nlm.nih.gov/pubmed/29619280
http://dx.doi.org/10.4103/jcis.JCIS_94_17
work_keys_str_mv AT sabatshyam radiationreductioninlowdosepulsedfluoroscopyversusstandarddosecontinuousfluoroscopyduringfluoroscopicallyguidedlumbarpuncturesaprospectivecontrolledstudy
AT slonimskyeinat radiationreductioninlowdosepulsedfluoroscopyversusstandarddosecontinuousfluoroscopyduringfluoroscopicallyguidedlumbarpuncturesaprospectivecontrolledstudy