A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution
A cuprous oxide (Cu(2)O) thin layer served as the base for a non-enzymatic glucose sensor in an alkaline medium, 0.1 NaOH solution, with a linear range of 50–200 mg/dL using differential pulse voltammetry (DPV) measurement. An X-ray photoelectron spectroscopy (XPS) study confirmed the formation of t...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5872052/ https://www.ncbi.nlm.nih.gov/pubmed/29316652 http://dx.doi.org/10.3390/bios8010004 |
_version_ | 1783309752577556480 |
---|---|
author | Dai, Yifan Molazemhosseini, Alireza Abbasi, Kevin Liu, Chung Chiun |
author_facet | Dai, Yifan Molazemhosseini, Alireza Abbasi, Kevin Liu, Chung Chiun |
author_sort | Dai, Yifan |
collection | PubMed |
description | A cuprous oxide (Cu(2)O) thin layer served as the base for a non-enzymatic glucose sensor in an alkaline medium, 0.1 NaOH solution, with a linear range of 50–200 mg/dL using differential pulse voltammetry (DPV) measurement. An X-ray photoelectron spectroscopy (XPS) study confirmed the formation of the cuprous oxide layer on the thin gold film sensor prototype. Quantitative detection of glucose in both phosphate-buffered saline (PBS) and undiluted human serum was carried out. Neither ascorbic acid nor uric acid, even at a relatively high concentration level (100 mg/dL in serum), interfered with the glucose detection, demonstrating the excellent selectivity of this non-enzymatic cuprous oxide thin layer-based glucose sensor. Chronoamperometry and single potential amperometric voltammetry were used to verify the measurements obtained by DPV, and the positive results validated that the detection of glucose in a 0.1 M NaOH alkaline medium by DPV measurement was effective. Nickel, platinum, and copper are commonly used metals for non-enzymatic glucose detection. The performance of these metal-based sensors for glucose detection using DPV were also evaluated. The cuprous oxide (Cu(2)O) thin layer-based sensor showed the best sensitivity for glucose detection among the sensors evaluated. |
format | Online Article Text |
id | pubmed-5872052 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-58720522018-03-29 A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution Dai, Yifan Molazemhosseini, Alireza Abbasi, Kevin Liu, Chung Chiun Biosensors (Basel) Article A cuprous oxide (Cu(2)O) thin layer served as the base for a non-enzymatic glucose sensor in an alkaline medium, 0.1 NaOH solution, with a linear range of 50–200 mg/dL using differential pulse voltammetry (DPV) measurement. An X-ray photoelectron spectroscopy (XPS) study confirmed the formation of the cuprous oxide layer on the thin gold film sensor prototype. Quantitative detection of glucose in both phosphate-buffered saline (PBS) and undiluted human serum was carried out. Neither ascorbic acid nor uric acid, even at a relatively high concentration level (100 mg/dL in serum), interfered with the glucose detection, demonstrating the excellent selectivity of this non-enzymatic cuprous oxide thin layer-based glucose sensor. Chronoamperometry and single potential amperometric voltammetry were used to verify the measurements obtained by DPV, and the positive results validated that the detection of glucose in a 0.1 M NaOH alkaline medium by DPV measurement was effective. Nickel, platinum, and copper are commonly used metals for non-enzymatic glucose detection. The performance of these metal-based sensors for glucose detection using DPV were also evaluated. The cuprous oxide (Cu(2)O) thin layer-based sensor showed the best sensitivity for glucose detection among the sensors evaluated. MDPI 2018-01-06 /pmc/articles/PMC5872052/ /pubmed/29316652 http://dx.doi.org/10.3390/bios8010004 Text en © 2018 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Dai, Yifan Molazemhosseini, Alireza Abbasi, Kevin Liu, Chung Chiun A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution |
title | A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution |
title_full | A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution |
title_fullStr | A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution |
title_full_unstemmed | A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution |
title_short | A Cuprous Oxide Thin Film Non-Enzymatic Glucose Sensor Using Differential Pulse Voltammetry and Other Voltammetry Methods and a Comparison to Different Thin Film Electrodes on the Detection of Glucose in an Alkaline Solution |
title_sort | cuprous oxide thin film non-enzymatic glucose sensor using differential pulse voltammetry and other voltammetry methods and a comparison to different thin film electrodes on the detection of glucose in an alkaline solution |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5872052/ https://www.ncbi.nlm.nih.gov/pubmed/29316652 http://dx.doi.org/10.3390/bios8010004 |
work_keys_str_mv | AT daiyifan acuprousoxidethinfilmnonenzymaticglucosesensorusingdifferentialpulsevoltammetryandothervoltammetrymethodsandacomparisontodifferentthinfilmelectrodesonthedetectionofglucoseinanalkalinesolution AT molazemhosseinialireza acuprousoxidethinfilmnonenzymaticglucosesensorusingdifferentialpulsevoltammetryandothervoltammetrymethodsandacomparisontodifferentthinfilmelectrodesonthedetectionofglucoseinanalkalinesolution AT abbasikevin acuprousoxidethinfilmnonenzymaticglucosesensorusingdifferentialpulsevoltammetryandothervoltammetrymethodsandacomparisontodifferentthinfilmelectrodesonthedetectionofglucoseinanalkalinesolution AT liuchungchiun acuprousoxidethinfilmnonenzymaticglucosesensorusingdifferentialpulsevoltammetryandothervoltammetrymethodsandacomparisontodifferentthinfilmelectrodesonthedetectionofglucoseinanalkalinesolution AT daiyifan cuprousoxidethinfilmnonenzymaticglucosesensorusingdifferentialpulsevoltammetryandothervoltammetrymethodsandacomparisontodifferentthinfilmelectrodesonthedetectionofglucoseinanalkalinesolution AT molazemhosseinialireza cuprousoxidethinfilmnonenzymaticglucosesensorusingdifferentialpulsevoltammetryandothervoltammetrymethodsandacomparisontodifferentthinfilmelectrodesonthedetectionofglucoseinanalkalinesolution AT abbasikevin cuprousoxidethinfilmnonenzymaticglucosesensorusingdifferentialpulsevoltammetryandothervoltammetrymethodsandacomparisontodifferentthinfilmelectrodesonthedetectionofglucoseinanalkalinesolution AT liuchungchiun cuprousoxidethinfilmnonenzymaticglucosesensorusingdifferentialpulsevoltammetryandothervoltammetrymethodsandacomparisontodifferentthinfilmelectrodesonthedetectionofglucoseinanalkalinesolution |