Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T
BACKGROUND: To evaluate the performance of a chemiluminescence detection kit for cardiac troponin T (cTnT). METHODS: According to the “Guiding principles on performance analysis of diagnostic reagents in vitro” and the Clinical and Laboratory Standards Institute (CLSI) Guidelines, we assessed the de...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5873270/ https://www.ncbi.nlm.nih.gov/pubmed/28940712 http://dx.doi.org/10.1002/jcla.22311 |
_version_ | 1783310011459436544 |
---|---|
author | Wang, Jingkun Fang, Yizhen Li, Peihua Lin, Huayue Jin, Hongwei Gao, Lei Ge, Shengxiang Zhang, Zhong‐Ying |
author_facet | Wang, Jingkun Fang, Yizhen Li, Peihua Lin, Huayue Jin, Hongwei Gao, Lei Ge, Shengxiang Zhang, Zhong‐Ying |
author_sort | Wang, Jingkun |
collection | PubMed |
description | BACKGROUND: To evaluate the performance of a chemiluminescence detection kit for cardiac troponin T (cTnT). METHODS: According to the “Guiding principles on performance analysis of diagnostic reagents in vitro” and the Clinical and Laboratory Standards Institute (CLSI) Guidelines, we assessed the detection limit, linear range, reportable range, accuracy, precision, cross reactivity, interference factors, and matrix effect of the kit, and compared these parameters with that of the commercial electrochemiluminescence detection kit for cTnT (Roche). RESULTS: The minimum detection limit of the kit was 0.01 ng/mL. The linear range was 0.01 ng/mL‐25 ng/mL. The reportable range was from 0.01 ng/mL to 100 ng/mL. Precision within the batch was 2.9%‐6.4%, and precision between batches was 6.0%; the accuracy was good and the recovery rate reached 96.2%. The cross‐reaction test demonstrated that there was no reactivity between cTnT and a variety of troponins. Test results deviated by less than ±10% in the presence of hemoglobin ≤1000 μg/mL, bilirubin ≤250 μg/mL, triglycerides ≤11.25 mg/mL, and rheumatoid factor ≤206 U/mL, suggesting that kit results were not significantly interfered with these factors. Results from the matrix‐effect assessment revealed absence of a matrix effect in the tested serum samples. Correlation study revealed that the performance of the kit was very consistent with that of the Roche electrochemiluminescence detection kit (Kappa = 0.900, P < .001) with a high correlation (r = .903, P < .001) and a total matching rate of 95.00%. CONCLUSION: The performance of the evaluated chemiluminescence immunoquantitation kit for cTnT detection was acceptable in all tested parameters, fulfilling requirements for clinical applications. |
format | Online Article Text |
id | pubmed-5873270 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-58732702018-03-31 Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T Wang, Jingkun Fang, Yizhen Li, Peihua Lin, Huayue Jin, Hongwei Gao, Lei Ge, Shengxiang Zhang, Zhong‐Ying J Clin Lab Anal Research Articles BACKGROUND: To evaluate the performance of a chemiluminescence detection kit for cardiac troponin T (cTnT). METHODS: According to the “Guiding principles on performance analysis of diagnostic reagents in vitro” and the Clinical and Laboratory Standards Institute (CLSI) Guidelines, we assessed the detection limit, linear range, reportable range, accuracy, precision, cross reactivity, interference factors, and matrix effect of the kit, and compared these parameters with that of the commercial electrochemiluminescence detection kit for cTnT (Roche). RESULTS: The minimum detection limit of the kit was 0.01 ng/mL. The linear range was 0.01 ng/mL‐25 ng/mL. The reportable range was from 0.01 ng/mL to 100 ng/mL. Precision within the batch was 2.9%‐6.4%, and precision between batches was 6.0%; the accuracy was good and the recovery rate reached 96.2%. The cross‐reaction test demonstrated that there was no reactivity between cTnT and a variety of troponins. Test results deviated by less than ±10% in the presence of hemoglobin ≤1000 μg/mL, bilirubin ≤250 μg/mL, triglycerides ≤11.25 mg/mL, and rheumatoid factor ≤206 U/mL, suggesting that kit results were not significantly interfered with these factors. Results from the matrix‐effect assessment revealed absence of a matrix effect in the tested serum samples. Correlation study revealed that the performance of the kit was very consistent with that of the Roche electrochemiluminescence detection kit (Kappa = 0.900, P < .001) with a high correlation (r = .903, P < .001) and a total matching rate of 95.00%. CONCLUSION: The performance of the evaluated chemiluminescence immunoquantitation kit for cTnT detection was acceptable in all tested parameters, fulfilling requirements for clinical applications. John Wiley and Sons Inc. 2017-09-22 /pmc/articles/PMC5873270/ /pubmed/28940712 http://dx.doi.org/10.1002/jcla.22311 Text en © 2017 The Authors Journal of Clinical Laboratory Analysis Published by Wiley Periodicals, Inc. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes. |
spellingShingle | Research Articles Wang, Jingkun Fang, Yizhen Li, Peihua Lin, Huayue Jin, Hongwei Gao, Lei Ge, Shengxiang Zhang, Zhong‐Ying Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T |
title | Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T |
title_full | Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T |
title_fullStr | Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T |
title_full_unstemmed | Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T |
title_short | Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T |
title_sort | evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin t |
topic | Research Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5873270/ https://www.ncbi.nlm.nih.gov/pubmed/28940712 http://dx.doi.org/10.1002/jcla.22311 |
work_keys_str_mv | AT wangjingkun evaluationofanewlydevelopedchemiluminescenceimmunoassayfordetectingcardiactroponint AT fangyizhen evaluationofanewlydevelopedchemiluminescenceimmunoassayfordetectingcardiactroponint AT lipeihua evaluationofanewlydevelopedchemiluminescenceimmunoassayfordetectingcardiactroponint AT linhuayue evaluationofanewlydevelopedchemiluminescenceimmunoassayfordetectingcardiactroponint AT jinhongwei evaluationofanewlydevelopedchemiluminescenceimmunoassayfordetectingcardiactroponint AT gaolei evaluationofanewlydevelopedchemiluminescenceimmunoassayfordetectingcardiactroponint AT geshengxiang evaluationofanewlydevelopedchemiluminescenceimmunoassayfordetectingcardiactroponint AT zhangzhongying evaluationofanewlydevelopedchemiluminescenceimmunoassayfordetectingcardiactroponint |