Cargando…
Difference in dose‐volumetric data between the analytical anisotropic algorithm, the dose‐to‐medium, and the dose‐to‐water reporting modes of the Acuros XB for lung stereotactic body radiation therapy
The purpose of this study was to evaluate the difference in dose‐volumetric data between the analytical anisotropic algorithms (AAA) and the two dose reporting modes of the Acuros XB, namely, the dose to water ([Formula: see text]) and dose to medium ([Formula: see text]) in lung stereotactic body r...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874099/ https://www.ncbi.nlm.nih.gov/pubmed/27685138 http://dx.doi.org/10.1120/jacmp.v17i5.6338 |
Sumario: | The purpose of this study was to evaluate the difference in dose‐volumetric data between the analytical anisotropic algorithms (AAA) and the two dose reporting modes of the Acuros XB, namely, the dose to water ([Formula: see text]) and dose to medium ([Formula: see text]) in lung stereotactic body radiotherapy (SBRT). Thirty‐eight plans were generated using the [Formula: see text] in Eclipse Treatment Planning System (TPS) and then recalculated with the [Formula: see text] and AAA, using identical beam setup. A dose of 50 Gy in 4 fractions was prescribed to the isocenter and the planning target volume (PTV) D95%. The isocenter was always inside the PTV. The following dose‐volumetric parameters were evaluated; D2%, D50%, D95%, and D98% for the internal target volume (ITV) and the PTV. Two‐tailed paired Student's t‐tests determined the statistical significance. Although for most of the parameters evaluated, the mean differences observed between the AAA, [Formula: see text] and [Formula: see text] were statistically significant ([Formula: see text]), absolute differences were rather small, in general less than 5% points. The maximum mean difference was observed in the ITV D50% between the [Formula: see text] and the AAA and was 1.7% points under the isocenter prescription and 3.3% points under the D95 prescription. [Formula: see text] produced higher values than [Formula: see text] with differences ranging from 0.4 to 1.1% points under isocenter prescription and 0.0 to 0.7% points under the PTV D95% prescription. The differences observed under the PTV D95% prescription were larger compared to those observed for the isocenter prescription between [Formula: see text] and AAA, [Formula: see text] and [Formula: see text] , and [Formula: see text] and AAA. Although statistically significant, the mean differences between the three algorithms are within 3.3% points. PACS number(s): 87.55.x, 87.55.D‐, 87.55.dk |
---|