A dosimetric comparison of copper and Cerrobend electron inserts

The purpose of this work was to evaluate differences in dose resulting from the use of copper aperture inserts compared to lead‐alloy (Cerrobend) aperture inserts for electron beam therapy. Specifically, this study examines if copper aperture inserts can be used clinically with the same commissionin...

Descripción completa

Detalles Bibliográficos
Autores principales: Rusk, Benjamin D., Carver, Robert L., Gibbons, John P., Hogstrom, Kenneth R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874111/
https://www.ncbi.nlm.nih.gov/pubmed/27685126
http://dx.doi.org/10.1120/jacmp.v17i5.6282
_version_ 1783310106906066944
author Rusk, Benjamin D.
Carver, Robert L.
Gibbons, John P.
Hogstrom, Kenneth R.
author_facet Rusk, Benjamin D.
Carver, Robert L.
Gibbons, John P.
Hogstrom, Kenneth R.
author_sort Rusk, Benjamin D.
collection PubMed
description The purpose of this work was to evaluate differences in dose resulting from the use of copper aperture inserts compared to lead‐alloy (Cerrobend) aperture inserts for electron beam therapy. Specifically, this study examines if copper aperture inserts can be used clinically with the same commissioning data measured using lead‐alloy aperture inserts. The copper inserts were acquired from .decimal, LLC and matching lead‐alloy, Cerrobend inserts were constructed in‐house for 32 combinations of nine square insert field sizes ([Formula: see text] to [Formula: see text]) and five applicator sizes ([Formula: see text] to [Formula: see text]). Percent depth‐dose and off‐axis relative dose profiles were measured using an electron diode in water for select copper and Cerrobend inserts for a subset of applicators ([Formula: see text]) and energies (6, 12, 20 MeV) at 100 and 110 cm source‐to‐surface distances (SSD) on a Varian Clinac 21EX accelerator. Dose outputs were measured for all field size‐insert combinations and five available energies ([Formula: see text]) at 100 cm SSD and for a smaller subset at 110 cm SSD. Using these data, 2D planar absolute dose distributions were generated and compared. Criteria for agreement were [Formula: see text] of maximum dose or 1 mm distance‐to‐agreement for 99% of points. A gamma analysis of the beam dosimetry showed 94 of 96 combinations of insert size, applicator, energy, and SSD were within the [Formula: see text] criteria for [Formula: see text] of points. Outside the field, copper inserts showed less bremsstrahlung dose under the insert compared to Cerrobend (greatest difference was 2.5% at 20 MeV and 100 cm SSD). This effect was most prominent at the highest energies for combinations of large applicators with small field sizes, causing some gamma analysis failures. Inside the field, more electrons scattered from the collimator edge of copper compared to Cerrobend, resulting in an increased dose at the field edge for copper at shallow depths (greatest increase was 1% at 20 MeV and 100 cm SSD). Dose differences decreased as the SSD increased, with no gamma failures at 110 cm SSD. Inserts for field sizes [Formula: see text] at any energy, or for small fields ([Formula: see text]) at energies [Formula: see text] , showed dosimetric differences less than [Formula: see text] for more than 99% of points. All areas of comparison criteria failures were from lower out‐of‐field dose under copper inserts due to a reduction in bremsstrahlung production, which is clinically beneficial in reducing dose to healthy tissue outside of the planned treatment volume. All field size‐applicator size‐energy combinations passed [Formula: see text] criteria for 100% of points. Therefore, it should be clinically acceptable to utilize copper insets with dose distributions measured with Cerrobend inserts for treatment planning dose calculations and monitor unit calculations. PACS number(s): 87.56.jk
format Online
Article
Text
id pubmed-5874111
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58741112018-04-02 A dosimetric comparison of copper and Cerrobend electron inserts Rusk, Benjamin D. Carver, Robert L. Gibbons, John P. Hogstrom, Kenneth R. J Appl Clin Med Phys Radiation Oncology Physics The purpose of this work was to evaluate differences in dose resulting from the use of copper aperture inserts compared to lead‐alloy (Cerrobend) aperture inserts for electron beam therapy. Specifically, this study examines if copper aperture inserts can be used clinically with the same commissioning data measured using lead‐alloy aperture inserts. The copper inserts were acquired from .decimal, LLC and matching lead‐alloy, Cerrobend inserts were constructed in‐house for 32 combinations of nine square insert field sizes ([Formula: see text] to [Formula: see text]) and five applicator sizes ([Formula: see text] to [Formula: see text]). Percent depth‐dose and off‐axis relative dose profiles were measured using an electron diode in water for select copper and Cerrobend inserts for a subset of applicators ([Formula: see text]) and energies (6, 12, 20 MeV) at 100 and 110 cm source‐to‐surface distances (SSD) on a Varian Clinac 21EX accelerator. Dose outputs were measured for all field size‐insert combinations and five available energies ([Formula: see text]) at 100 cm SSD and for a smaller subset at 110 cm SSD. Using these data, 2D planar absolute dose distributions were generated and compared. Criteria for agreement were [Formula: see text] of maximum dose or 1 mm distance‐to‐agreement for 99% of points. A gamma analysis of the beam dosimetry showed 94 of 96 combinations of insert size, applicator, energy, and SSD were within the [Formula: see text] criteria for [Formula: see text] of points. Outside the field, copper inserts showed less bremsstrahlung dose under the insert compared to Cerrobend (greatest difference was 2.5% at 20 MeV and 100 cm SSD). This effect was most prominent at the highest energies for combinations of large applicators with small field sizes, causing some gamma analysis failures. Inside the field, more electrons scattered from the collimator edge of copper compared to Cerrobend, resulting in an increased dose at the field edge for copper at shallow depths (greatest increase was 1% at 20 MeV and 100 cm SSD). Dose differences decreased as the SSD increased, with no gamma failures at 110 cm SSD. Inserts for field sizes [Formula: see text] at any energy, or for small fields ([Formula: see text]) at energies [Formula: see text] , showed dosimetric differences less than [Formula: see text] for more than 99% of points. All areas of comparison criteria failures were from lower out‐of‐field dose under copper inserts due to a reduction in bremsstrahlung production, which is clinically beneficial in reducing dose to healthy tissue outside of the planned treatment volume. All field size‐applicator size‐energy combinations passed [Formula: see text] criteria for 100% of points. Therefore, it should be clinically acceptable to utilize copper insets with dose distributions measured with Cerrobend inserts for treatment planning dose calculations and monitor unit calculations. PACS number(s): 87.56.jk John Wiley and Sons Inc. 2016-09-08 /pmc/articles/PMC5874111/ /pubmed/27685126 http://dx.doi.org/10.1120/jacmp.v17i5.6282 Text en © 2016 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Rusk, Benjamin D.
Carver, Robert L.
Gibbons, John P.
Hogstrom, Kenneth R.
A dosimetric comparison of copper and Cerrobend electron inserts
title A dosimetric comparison of copper and Cerrobend electron inserts
title_full A dosimetric comparison of copper and Cerrobend electron inserts
title_fullStr A dosimetric comparison of copper and Cerrobend electron inserts
title_full_unstemmed A dosimetric comparison of copper and Cerrobend electron inserts
title_short A dosimetric comparison of copper and Cerrobend electron inserts
title_sort dosimetric comparison of copper and cerrobend electron inserts
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874111/
https://www.ncbi.nlm.nih.gov/pubmed/27685126
http://dx.doi.org/10.1120/jacmp.v17i5.6282
work_keys_str_mv AT ruskbenjamind adosimetriccomparisonofcopperandcerrobendelectroninserts
AT carverrobertl adosimetriccomparisonofcopperandcerrobendelectroninserts
AT gibbonsjohnp adosimetriccomparisonofcopperandcerrobendelectroninserts
AT hogstromkennethr adosimetriccomparisonofcopperandcerrobendelectroninserts
AT ruskbenjamind dosimetriccomparisonofcopperandcerrobendelectroninserts
AT carverrobertl dosimetriccomparisonofcopperandcerrobendelectroninserts
AT gibbonsjohnp dosimetriccomparisonofcopperandcerrobendelectroninserts
AT hogstromkennethr dosimetriccomparisonofcopperandcerrobendelectroninserts