Optimization of the dosimetric leaf gap for use in planning VMAT treatments of spine SABR cases

The dosimetric leaf gap (DLG) is a beam configuration parameter used in the Varian Eclipse treatment planning system, to model the effects of rounded MLC leaf ends. Measuring the DLG using the conventional sliding‐slit technique has been shown to be produce questionable results for some volumetric m...

Descripción completa

Detalles Bibliográficos
Autores principales: Middlebrook, Nigel D, Sutherland, Bess, Kairn, Tanya
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874863/
https://www.ncbi.nlm.nih.gov/pubmed/28574219
http://dx.doi.org/10.1002/acm2.12106
Descripción
Sumario:The dosimetric leaf gap (DLG) is a beam configuration parameter used in the Varian Eclipse treatment planning system, to model the effects of rounded MLC leaf ends. Measuring the DLG using the conventional sliding‐slit technique has been shown to be produce questionable results for some volumetric modulated arc therapy (VMAT) treatments. This study therefore investigated the use of radiochromic film measurements to optimize the DLG specifically for the purpose of producing accurate VMAT plans using a flattening‐filter‐free (FFF) beam, for use in treating vertebral targets using a stereotactic (SABR, also known as SBRT) fractionation schedule. Four test treatments were planned using a VMAT technique, to deliver a prescription of 24 Gy in 3 fractions to four different spine SABR treatment sites. Measurements of the doses delivered by these treatments were acquired using an ionization chamber and radiographic film. These measurements were compared with the doses calculated by the treatment planning system using a range of DLG values, including a DLG identified using the conventional sliding‐slit method (1.1 mm). An optimal DLG value was identified, as the value that produced the closest agreement between the planned and measured doses (1.9 mm). The accuracy of the dose calculations produced using the optimized DLG value was verified using additional radiochromic film measurements in a heterogeneous phantom. This study provided a specific initial DLG (1.9 mm) as well as a film‐based optimization method, which may be used by radiotherapy centers when attempting to commission or improve an FFF VMAT‐based SABR treatment programme.