Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition

Mobile fluoroscopy (c‐arm) units offering 3D image reconstruction are becoming more common in surgical settings. Although these images are “CT‐like” and sometimes replace the postoperative CT, the acquisition is technically very different from a traditional CT acquisition. Dose assessment is complic...

Descripción completa

Detalles Bibliográficos
Autor principal: Leon, Stephanie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874898/
https://www.ncbi.nlm.nih.gov/pubmed/28585724
http://dx.doi.org/10.1002/acm2.12108
_version_ 1783310256075440128
author Leon, Stephanie
author_facet Leon, Stephanie
author_sort Leon, Stephanie
collection PubMed
description Mobile fluoroscopy (c‐arm) units offering 3D image reconstruction are becoming more common in surgical settings. Although these images are “CT‐like” and sometimes replace the postoperative CT, the acquisition is technically very different from a traditional CT acquisition. Dose assessment is complicated by a large beam width, automatic exposure rate control, and a rotation of less than 360°. The purpose of this work was to explore the impact of these factors on the volumetric dose calculation and to provide practical recommendations for clinical physicists assessing dose from these units using commonly available equipment. CTDI(W) was calculated using the IAEA method for dosimetry of wide beams and compared to scans of the 32‐cm CTDI phantom using the full beam width and a 20‐mm collimated beam width. The impact of the partial rotation on the CTDI(W) calculation was assessed by acquiring measurements at four and twelve positions on the phantom periphery. For the system tested, the CTDI(W) was calculated to be 16.1 mGy using the IAEA method with default clinical protocol. Results showed that measuring CTDI(W) with the full beam width or a collimated beam width alone resulted in CTDI values of 19.0 mGy and 19.5 mGy, respectively. Using four peripheral measurements instead of 12 resulted in a difference of 4% for a collimated beam and 6% for an open beam. Variations in positioning on the order of a few centimeters resulted in a variation of only 4% with an open beam. The excellent reproducibility of the measurements using the full beam width suggests that this simple method is adequate for year‐to‐year comparisons. In contrast, the IAEA method is difficult to employ, particularly with 180° acquisitions. Use of peripheral measurements in excess of the usual four is time‐consuming and not necessary for most applications obtained with the geometry specific to this system.
format Online
Article
Text
id pubmed-5874898
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58748982018-04-02 Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition Leon, Stephanie J Appl Clin Med Phys Medical Imaging Mobile fluoroscopy (c‐arm) units offering 3D image reconstruction are becoming more common in surgical settings. Although these images are “CT‐like” and sometimes replace the postoperative CT, the acquisition is technically very different from a traditional CT acquisition. Dose assessment is complicated by a large beam width, automatic exposure rate control, and a rotation of less than 360°. The purpose of this work was to explore the impact of these factors on the volumetric dose calculation and to provide practical recommendations for clinical physicists assessing dose from these units using commonly available equipment. CTDI(W) was calculated using the IAEA method for dosimetry of wide beams and compared to scans of the 32‐cm CTDI phantom using the full beam width and a 20‐mm collimated beam width. The impact of the partial rotation on the CTDI(W) calculation was assessed by acquiring measurements at four and twelve positions on the phantom periphery. For the system tested, the CTDI(W) was calculated to be 16.1 mGy using the IAEA method with default clinical protocol. Results showed that measuring CTDI(W) with the full beam width or a collimated beam width alone resulted in CTDI values of 19.0 mGy and 19.5 mGy, respectively. Using four peripheral measurements instead of 12 resulted in a difference of 4% for a collimated beam and 6% for an open beam. Variations in positioning on the order of a few centimeters resulted in a variation of only 4% with an open beam. The excellent reproducibility of the measurements using the full beam width suggests that this simple method is adequate for year‐to‐year comparisons. In contrast, the IAEA method is difficult to employ, particularly with 180° acquisitions. Use of peripheral measurements in excess of the usual four is time‐consuming and not necessary for most applications obtained with the geometry specific to this system. John Wiley and Sons Inc. 2017-06-06 /pmc/articles/PMC5874898/ /pubmed/28585724 http://dx.doi.org/10.1002/acm2.12108 Text en © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Leon, Stephanie
Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition
title Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition
title_full Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition
title_fullStr Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition
title_full_unstemmed Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition
title_short Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition
title_sort assessment of volumetric absorbed dose for mobile fluoroscopic 3d image acquisition
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874898/
https://www.ncbi.nlm.nih.gov/pubmed/28585724
http://dx.doi.org/10.1002/acm2.12108
work_keys_str_mv AT leonstephanie assessmentofvolumetricabsorbeddoseformobilefluoroscopic3dimageacquisition