Sensitivity of array detector measurements in determining shifts of MLC leaf positions

Using a MatriXX 2D ionization chamber array, we evaluated the detection sensitivity of systematically introduced MLC leaf positioning shifts to test whether the conventional IMRT QA method can be used for quality assurance of an MLC tracking algorithm. Because of finite special resolution, we first...

Descripción completa

Detalles Bibliográficos
Autores principales: Shang, Qingyang, Godley, Andrew, Huang, Long, Qi, Peng, Xia, Ping
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874934/
https://www.ncbi.nlm.nih.gov/pubmed/28799273
http://dx.doi.org/10.1002/acm2.12148
_version_ 1783310264292081664
author Shang, Qingyang
Godley, Andrew
Huang, Long
Qi, Peng
Xia, Ping
author_facet Shang, Qingyang
Godley, Andrew
Huang, Long
Qi, Peng
Xia, Ping
author_sort Shang, Qingyang
collection PubMed
description Using a MatriXX 2D ionization chamber array, we evaluated the detection sensitivity of systematically introduced MLC leaf positioning shifts to test whether the conventional IMRT QA method can be used for quality assurance of an MLC tracking algorithm. Because of finite special resolution, we first tested whether the detection sensitivity was dependent of the locations of leaf shifts and positions of ionization chambers. We then introduced the same systematic leaf shifts in two clinical intensity modulated radiotherapy plans (prostate and head and neck cancer). Our results reported differences between the measured planar doses with and without MLC shifts (errors). Independent of the locations of the leaf position shifts and positions of the detectors, for the simple rectangular fields, the MatriXX was able to detect ±2 mm MLC leaf positioning shifts with Gamma index of 3%/3 mm and ±1 mm MLC leaf position shifts with Gamma index of 2%/2 mm. For the clinical plans, measuring the fields individually, leaf positioning shifts of ±2 mm were detected using Gamma index of 3%/3 mm and a passing rate of 95%. When the fields were measured compositely, the Gamma index exhibited less sensitivity for the detection of leaf positioning shifts than when the fields were measured individually. In conclusion, if more than 2 mm MLC leaf shifts were required, the commercial detector array (MatriXX) is able to detect such MLC positioning shifts, otherwise a more sensitive quality assurance method should be used.
format Online
Article
Text
id pubmed-5874934
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58749342018-04-02 Sensitivity of array detector measurements in determining shifts of MLC leaf positions Shang, Qingyang Godley, Andrew Huang, Long Qi, Peng Xia, Ping J Appl Clin Med Phys Radiation Oncology Physics Using a MatriXX 2D ionization chamber array, we evaluated the detection sensitivity of systematically introduced MLC leaf positioning shifts to test whether the conventional IMRT QA method can be used for quality assurance of an MLC tracking algorithm. Because of finite special resolution, we first tested whether the detection sensitivity was dependent of the locations of leaf shifts and positions of ionization chambers. We then introduced the same systematic leaf shifts in two clinical intensity modulated radiotherapy plans (prostate and head and neck cancer). Our results reported differences between the measured planar doses with and without MLC shifts (errors). Independent of the locations of the leaf position shifts and positions of the detectors, for the simple rectangular fields, the MatriXX was able to detect ±2 mm MLC leaf positioning shifts with Gamma index of 3%/3 mm and ±1 mm MLC leaf position shifts with Gamma index of 2%/2 mm. For the clinical plans, measuring the fields individually, leaf positioning shifts of ±2 mm were detected using Gamma index of 3%/3 mm and a passing rate of 95%. When the fields were measured compositely, the Gamma index exhibited less sensitivity for the detection of leaf positioning shifts than when the fields were measured individually. In conclusion, if more than 2 mm MLC leaf shifts were required, the commercial detector array (MatriXX) is able to detect such MLC positioning shifts, otherwise a more sensitive quality assurance method should be used. John Wiley and Sons Inc. 2017-08-11 /pmc/articles/PMC5874934/ /pubmed/28799273 http://dx.doi.org/10.1002/acm2.12148 Text en © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Shang, Qingyang
Godley, Andrew
Huang, Long
Qi, Peng
Xia, Ping
Sensitivity of array detector measurements in determining shifts of MLC leaf positions
title Sensitivity of array detector measurements in determining shifts of MLC leaf positions
title_full Sensitivity of array detector measurements in determining shifts of MLC leaf positions
title_fullStr Sensitivity of array detector measurements in determining shifts of MLC leaf positions
title_full_unstemmed Sensitivity of array detector measurements in determining shifts of MLC leaf positions
title_short Sensitivity of array detector measurements in determining shifts of MLC leaf positions
title_sort sensitivity of array detector measurements in determining shifts of mlc leaf positions
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874934/
https://www.ncbi.nlm.nih.gov/pubmed/28799273
http://dx.doi.org/10.1002/acm2.12148
work_keys_str_mv AT shangqingyang sensitivityofarraydetectormeasurementsindeterminingshiftsofmlcleafpositions
AT godleyandrew sensitivityofarraydetectormeasurementsindeterminingshiftsofmlcleafpositions
AT huanglong sensitivityofarraydetectormeasurementsindeterminingshiftsofmlcleafpositions
AT qipeng sensitivityofarraydetectormeasurementsindeterminingshiftsofmlcleafpositions
AT xiaping sensitivityofarraydetectormeasurementsindeterminingshiftsofmlcleafpositions