Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments
A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30‐minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this app...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2016
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874969/ https://www.ncbi.nlm.nih.gov/pubmed/27074487 http://dx.doi.org/10.1120/jacmp.v17i2.6040 |
_version_ | 1783310272198344704 |
---|---|
author | Held, Mareike Cremers, Florian Sneed, Penny K. Braunstein, Steve Fogh, Shannon E. Nakamura, Jean Barani, Igor Perez‐Andujar, Angelica Pouliot, Jean Morin, Olivier |
author_facet | Held, Mareike Cremers, Florian Sneed, Penny K. Braunstein, Steve Fogh, Shannon E. Nakamura, Jean Barani, Igor Perez‐Andujar, Angelica Pouliot, Jean Morin, Olivier |
author_sort | Held, Mareike |
collection | PubMed |
description | A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30‐minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this approach available to all clinics equipped with common imaging systems, dose calculation accuracy based on treatment sites was assessed for other imaging units. We evaluated the feasibility of palliative treatment planning using on‐board imaging with respect to image quality and technical challenges. The purpose was to test multiple systems using their commercial setup, disregarding any additional in‐house development. kV CT, kV CBCT, MV CBCT, and MV CT images of water and anthropomorphic phantoms were acquired on five different imaging units (Philips MX8000 CT Scanner, and Varian TrueBeam, Elekta VersaHD, Siemens Artiste, and Accuray Tomotherapy linacs). Image quality (noise, contrast, uniformity, spatial resolution) was evaluated and compared across all machines. Using individual image value to density calibrations, dose calculation accuracies for simple treatment plans were assessed for the same phantom images. Finally, image artifacts on clinical patient images were evaluated and compared among the machines. Image contrast to visualize bony anatomy was sufficient on all machines. Despite a high noise level and low contrast, MV CT images provided the most accurate treatment plans relative to kV CT‐based planning. Spatial resolution was poorest for MV CBCT, but did not limit the visualization of small anatomical structures. A comparison of treatment plans showed that monitor units calculated based on a prescription point were within 5% difference relative to kV CT‐based plans for all machines and all studied treatment sites (brain, neck, and pelvis). Local dose differences [Formula: see text] were found near the phantom edges. The gamma index for 3%/3 mm criteria was [Formula: see text] in most cases. Best dose calculation results were obtained when the treatment isocenter was near the image isocenter for all machines. A large field of view and immediate image export to the treatment planning system were essential for a smooth workflow and were not provided on all devices. Based on this phantom study, image quality of the studied kV CBCT, MV CBCT, and MV CT on‐board imaging devices was sufficient for treatment planning in all tested cases. Treatment plans provided dose calculation accuracies within an acceptable range for simple, urgently planned palliative treatments. However, dose calculation accuracy was compromised towards the edges of an image. Feasibility for clinical implementation should be assessed separately and may be complicated by machine specific features. Image artifacts in patient images and the effect on dose calculation accuracy should be assessed in a separate, machine‐specific study. PACS number(s): 87.55.D‐, 87.57.C‐, 87.57.Q‐ |
format | Online Article Text |
id | pubmed-5874969 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2016 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-58749692018-04-02 Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments Held, Mareike Cremers, Florian Sneed, Penny K. Braunstein, Steve Fogh, Shannon E. Nakamura, Jean Barani, Igor Perez‐Andujar, Angelica Pouliot, Jean Morin, Olivier J Appl Clin Med Phys Radiation Oncology Physics A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30‐minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this approach available to all clinics equipped with common imaging systems, dose calculation accuracy based on treatment sites was assessed for other imaging units. We evaluated the feasibility of palliative treatment planning using on‐board imaging with respect to image quality and technical challenges. The purpose was to test multiple systems using their commercial setup, disregarding any additional in‐house development. kV CT, kV CBCT, MV CBCT, and MV CT images of water and anthropomorphic phantoms were acquired on five different imaging units (Philips MX8000 CT Scanner, and Varian TrueBeam, Elekta VersaHD, Siemens Artiste, and Accuray Tomotherapy linacs). Image quality (noise, contrast, uniformity, spatial resolution) was evaluated and compared across all machines. Using individual image value to density calibrations, dose calculation accuracies for simple treatment plans were assessed for the same phantom images. Finally, image artifacts on clinical patient images were evaluated and compared among the machines. Image contrast to visualize bony anatomy was sufficient on all machines. Despite a high noise level and low contrast, MV CT images provided the most accurate treatment plans relative to kV CT‐based planning. Spatial resolution was poorest for MV CBCT, but did not limit the visualization of small anatomical structures. A comparison of treatment plans showed that monitor units calculated based on a prescription point were within 5% difference relative to kV CT‐based plans for all machines and all studied treatment sites (brain, neck, and pelvis). Local dose differences [Formula: see text] were found near the phantom edges. The gamma index for 3%/3 mm criteria was [Formula: see text] in most cases. Best dose calculation results were obtained when the treatment isocenter was near the image isocenter for all machines. A large field of view and immediate image export to the treatment planning system were essential for a smooth workflow and were not provided on all devices. Based on this phantom study, image quality of the studied kV CBCT, MV CBCT, and MV CT on‐board imaging devices was sufficient for treatment planning in all tested cases. Treatment plans provided dose calculation accuracies within an acceptable range for simple, urgently planned palliative treatments. However, dose calculation accuracy was compromised towards the edges of an image. Feasibility for clinical implementation should be assessed separately and may be complicated by machine specific features. Image artifacts in patient images and the effect on dose calculation accuracy should be assessed in a separate, machine‐specific study. PACS number(s): 87.55.D‐, 87.57.C‐, 87.57.Q‐ John Wiley and Sons Inc. 2016-03-08 /pmc/articles/PMC5874969/ /pubmed/27074487 http://dx.doi.org/10.1120/jacmp.v17i2.6040 Text en © 2016 The Authors. This is an open access article under the terms of the http://creativecommons.org/licenses/by/3.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Held, Mareike Cremers, Florian Sneed, Penny K. Braunstein, Steve Fogh, Shannon E. Nakamura, Jean Barani, Igor Perez‐Andujar, Angelica Pouliot, Jean Morin, Olivier Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments |
title | Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments |
title_full | Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments |
title_fullStr | Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments |
title_full_unstemmed | Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments |
title_short | Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments |
title_sort | assessment of image quality and dose calculation accuracy on kv cbct, mv cbct, and mv ct images for urgent palliative radiotherapy treatments |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874969/ https://www.ncbi.nlm.nih.gov/pubmed/27074487 http://dx.doi.org/10.1120/jacmp.v17i2.6040 |
work_keys_str_mv | AT heldmareike assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT cremersflorian assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT sneedpennyk assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT braunsteinsteve assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT foghshannone assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT nakamurajean assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT baraniigor assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT perezandujarangelica assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT pouliotjean assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments AT morinolivier assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments |