Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments

A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30‐minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this app...

Descripción completa

Detalles Bibliográficos
Autores principales: Held, Mareike, Cremers, Florian, Sneed, Penny K., Braunstein, Steve, Fogh, Shannon E., Nakamura, Jean, Barani, Igor, Perez‐Andujar, Angelica, Pouliot, Jean, Morin, Olivier
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2016
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874969/
https://www.ncbi.nlm.nih.gov/pubmed/27074487
http://dx.doi.org/10.1120/jacmp.v17i2.6040
_version_ 1783310272198344704
author Held, Mareike
Cremers, Florian
Sneed, Penny K.
Braunstein, Steve
Fogh, Shannon E.
Nakamura, Jean
Barani, Igor
Perez‐Andujar, Angelica
Pouliot, Jean
Morin, Olivier
author_facet Held, Mareike
Cremers, Florian
Sneed, Penny K.
Braunstein, Steve
Fogh, Shannon E.
Nakamura, Jean
Barani, Igor
Perez‐Andujar, Angelica
Pouliot, Jean
Morin, Olivier
author_sort Held, Mareike
collection PubMed
description A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30‐minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this approach available to all clinics equipped with common imaging systems, dose calculation accuracy based on treatment sites was assessed for other imaging units. We evaluated the feasibility of palliative treatment planning using on‐board imaging with respect to image quality and technical challenges. The purpose was to test multiple systems using their commercial setup, disregarding any additional in‐house development. kV CT, kV CBCT, MV CBCT, and MV CT images of water and anthropomorphic phantoms were acquired on five different imaging units (Philips MX8000 CT Scanner, and Varian TrueBeam, Elekta VersaHD, Siemens Artiste, and Accuray Tomotherapy linacs). Image quality (noise, contrast, uniformity, spatial resolution) was evaluated and compared across all machines. Using individual image value to density calibrations, dose calculation accuracies for simple treatment plans were assessed for the same phantom images. Finally, image artifacts on clinical patient images were evaluated and compared among the machines. Image contrast to visualize bony anatomy was sufficient on all machines. Despite a high noise level and low contrast, MV CT images provided the most accurate treatment plans relative to kV CT‐based planning. Spatial resolution was poorest for MV CBCT, but did not limit the visualization of small anatomical structures. A comparison of treatment plans showed that monitor units calculated based on a prescription point were within 5% difference relative to kV CT‐based plans for all machines and all studied treatment sites (brain, neck, and pelvis). Local dose differences [Formula: see text] were found near the phantom edges. The gamma index for 3%/3 mm criteria was [Formula: see text] in most cases. Best dose calculation results were obtained when the treatment isocenter was near the image isocenter for all machines. A large field of view and immediate image export to the treatment planning system were essential for a smooth workflow and were not provided on all devices. Based on this phantom study, image quality of the studied kV CBCT, MV CBCT, and MV CT on‐board imaging devices was sufficient for treatment planning in all tested cases. Treatment plans provided dose calculation accuracies within an acceptable range for simple, urgently planned palliative treatments. However, dose calculation accuracy was compromised towards the edges of an image. Feasibility for clinical implementation should be assessed separately and may be complicated by machine specific features. Image artifacts in patient images and the effect on dose calculation accuracy should be assessed in a separate, machine‐specific study. PACS number(s): 87.55.D‐, 87.57.C‐, 87.57.Q‐
format Online
Article
Text
id pubmed-5874969
institution National Center for Biotechnology Information
language English
publishDate 2016
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58749692018-04-02 Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments Held, Mareike Cremers, Florian Sneed, Penny K. Braunstein, Steve Fogh, Shannon E. Nakamura, Jean Barani, Igor Perez‐Andujar, Angelica Pouliot, Jean Morin, Olivier J Appl Clin Med Phys Radiation Oncology Physics A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30‐minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this approach available to all clinics equipped with common imaging systems, dose calculation accuracy based on treatment sites was assessed for other imaging units. We evaluated the feasibility of palliative treatment planning using on‐board imaging with respect to image quality and technical challenges. The purpose was to test multiple systems using their commercial setup, disregarding any additional in‐house development. kV CT, kV CBCT, MV CBCT, and MV CT images of water and anthropomorphic phantoms were acquired on five different imaging units (Philips MX8000 CT Scanner, and Varian TrueBeam, Elekta VersaHD, Siemens Artiste, and Accuray Tomotherapy linacs). Image quality (noise, contrast, uniformity, spatial resolution) was evaluated and compared across all machines. Using individual image value to density calibrations, dose calculation accuracies for simple treatment plans were assessed for the same phantom images. Finally, image artifacts on clinical patient images were evaluated and compared among the machines. Image contrast to visualize bony anatomy was sufficient on all machines. Despite a high noise level and low contrast, MV CT images provided the most accurate treatment plans relative to kV CT‐based planning. Spatial resolution was poorest for MV CBCT, but did not limit the visualization of small anatomical structures. A comparison of treatment plans showed that monitor units calculated based on a prescription point were within 5% difference relative to kV CT‐based plans for all machines and all studied treatment sites (brain, neck, and pelvis). Local dose differences [Formula: see text] were found near the phantom edges. The gamma index for 3%/3 mm criteria was [Formula: see text] in most cases. Best dose calculation results were obtained when the treatment isocenter was near the image isocenter for all machines. A large field of view and immediate image export to the treatment planning system were essential for a smooth workflow and were not provided on all devices. Based on this phantom study, image quality of the studied kV CBCT, MV CBCT, and MV CT on‐board imaging devices was sufficient for treatment planning in all tested cases. Treatment plans provided dose calculation accuracies within an acceptable range for simple, urgently planned palliative treatments. However, dose calculation accuracy was compromised towards the edges of an image. Feasibility for clinical implementation should be assessed separately and may be complicated by machine specific features. Image artifacts in patient images and the effect on dose calculation accuracy should be assessed in a separate, machine‐specific study. PACS number(s): 87.55.D‐, 87.57.C‐, 87.57.Q‐ John Wiley and Sons Inc. 2016-03-08 /pmc/articles/PMC5874969/ /pubmed/27074487 http://dx.doi.org/10.1120/jacmp.v17i2.6040 Text en © 2016 The Authors. This is an open access article under the terms of the http://creativecommons.org/licenses/by/3.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Held, Mareike
Cremers, Florian
Sneed, Penny K.
Braunstein, Steve
Fogh, Shannon E.
Nakamura, Jean
Barani, Igor
Perez‐Andujar, Angelica
Pouliot, Jean
Morin, Olivier
Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments
title Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments
title_full Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments
title_fullStr Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments
title_full_unstemmed Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments
title_short Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments
title_sort assessment of image quality and dose calculation accuracy on kv cbct, mv cbct, and mv ct images for urgent palliative radiotherapy treatments
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5874969/
https://www.ncbi.nlm.nih.gov/pubmed/27074487
http://dx.doi.org/10.1120/jacmp.v17i2.6040
work_keys_str_mv AT heldmareike assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT cremersflorian assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT sneedpennyk assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT braunsteinsteve assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT foghshannone assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT nakamurajean assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT baraniigor assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT perezandujarangelica assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT pouliotjean assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments
AT morinolivier assessmentofimagequalityanddosecalculationaccuracyonkvcbctmvcbctandmvctimagesforurgentpalliativeradiotherapytreatments