Clinical evaluation of an anatomy‐based patient specific quality assurance system

The Delta(4DVH) Anatomy 3D quality assurance (QA) system (ScandiDos), which converts the measured detector dose into the dose distribution in the patient geometry was evaluated. It allows a direct comparison of the calculated 3D dose with the measured back‐projected dose. In total, 16 static and 16...

Descripción completa

Detalles Bibliográficos
Autores principales: Hauri, Pascal, Verlaan, Sarah, Graydon, Shaun, Ahnen, Linda, Klöck, Stephan, Lang, Stephanie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875461/
https://www.ncbi.nlm.nih.gov/pubmed/24710453
http://dx.doi.org/10.1120/jacmp.v15i2.4647
Descripción
Sumario:The Delta(4DVH) Anatomy 3D quality assurance (QA) system (ScandiDos), which converts the measured detector dose into the dose distribution in the patient geometry was evaluated. It allows a direct comparison of the calculated 3D dose with the measured back‐projected dose. In total, 16 static and 16 volumetric‐modulated arc therapy (VMAT) fields were planned using four different energies. Isocenter dose was measured with a pinpoint chamber in homogeneous phantoms to investigate the dose prediction by the Delta(4DVH) Anatomy algorithm for static fields. Dose distributions of VMAT fields were measured using GAFCHROMIC film. Gravitational gantry errors up to 10° were introduced into all VMAT plans to study the potential of detecting errors. Additionally, 20 clinical treatment plans were verified. For static fields, the Delta(4DVH) Anatomy predicted the isocenter dose accurately, with a deviation to the measured phantom dose of [Formula: see text]. For VMAT fields the predicted Delta(4DVH) Anatomy dose in the isocenter plane corresponded to the measured dose in the phantom, with an average gamma agreement index (GAI) [Formula: see text] of [Formula: see text]. The Delta(4DVH) Anatomy detected the induced systematic gantry error of 10° with a relative GAI [Formula: see text] change of [Formula: see text]. The conventional Delta(4PT) QA system detected a GAI change of [Formula: see text]. The conventional Delta(4PT) GAI [Formula: see text] was [Formula: see text] for the clinical treatment plans. The mean body and PTV‐GAI [Formula: see text] for the Delta(4DVH) Anatomy were [Formula: see text] and [Formula: see text]; however, this dropped to [Formula: see text] and [Formula: see text] for passing criteria of [Formula: see text]. The anatomy‐based patient specific quality assurance system predicts the dose distribution correctly for a homogeneous case. The limiting factor for the error detection is the large variability in the error‐free plans. The dose calculation algorithm is inferior to that used in the TPS (Eclipse). PACS numbers: 87.56.Fc, 87.56.‐v