Evaluation of techniques for slice sensitivity profile measurement and analysis

The purpose of this study was to compare the resulting full width at half maximum of slice sensitivity profiles (SSP) generated by several commercially available point response phantoms, and determine an appropriate imaging technique and analysis method. Four CT phantoms containing point response ob...

Descripción completa

Detalles Bibliográficos
Autores principales: Greene, Travis C., Rong, X. John
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875475/
https://www.ncbi.nlm.nih.gov/pubmed/24710429
http://dx.doi.org/10.1120/jacmp.v15i2.4042
_version_ 1783310352094593024
author Greene, Travis C.
Rong, X. John
author_facet Greene, Travis C.
Rong, X. John
author_sort Greene, Travis C.
collection PubMed
description The purpose of this study was to compare the resulting full width at half maximum of slice sensitivity profiles (SSP) generated by several commercially available point response phantoms, and determine an appropriate imaging technique and analysis method. Four CT phantoms containing point response objects designed to produce a delta impulse signal used in this study: a Fluke CT‐SSP phantom, a Gammex 464, a CatPhan 600, and a Kagaku Micro Disc phantom. Each phantom was imaged using 120 kVp, 325 mAs, head scan field of view, [Formula: see text] helical scan with a 20 mm beam width and a pitch of 0.969. The acquired images were then reconstructed into all available slice thicknesses [Formula: see text]. A computer program was developed to analyze the images of each dataset for generating a SSP from which the full width at half maximum (FWHM) was determined. Two methods for generating SSPs were evaluated and compared by choosing the mean vs. maximum value in the ROI, along with two methods for evaluating the FWHM of the SSP, linear interpolation and Gaussian curve fitting. FWHMs were compared with the manufacturer's specifications using percent error and z‐test with a significance value of [Formula: see text]. The FWHMs from each phantom were not significantly different [Formula: see text] with an average error of 3.5%. The FWHMs from SSPs generated from the mean value were statistically different [Formula: see text]. The FWHMs from the different FWHM methods were not statistically different [Formula: see text]. Evaluation of the SSP is dependent on the ROI value used. The maximum value from the ROI should be used to generate the SSP whenever possible. SSP measurement is independent of the phantoms used in this study. PACS number: 87.
format Online
Article
Text
id pubmed-5875475
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58754752018-04-02 Evaluation of techniques for slice sensitivity profile measurement and analysis Greene, Travis C. Rong, X. John J Appl Clin Med Phys Medical Imaging The purpose of this study was to compare the resulting full width at half maximum of slice sensitivity profiles (SSP) generated by several commercially available point response phantoms, and determine an appropriate imaging technique and analysis method. Four CT phantoms containing point response objects designed to produce a delta impulse signal used in this study: a Fluke CT‐SSP phantom, a Gammex 464, a CatPhan 600, and a Kagaku Micro Disc phantom. Each phantom was imaged using 120 kVp, 325 mAs, head scan field of view, [Formula: see text] helical scan with a 20 mm beam width and a pitch of 0.969. The acquired images were then reconstructed into all available slice thicknesses [Formula: see text]. A computer program was developed to analyze the images of each dataset for generating a SSP from which the full width at half maximum (FWHM) was determined. Two methods for generating SSPs were evaluated and compared by choosing the mean vs. maximum value in the ROI, along with two methods for evaluating the FWHM of the SSP, linear interpolation and Gaussian curve fitting. FWHMs were compared with the manufacturer's specifications using percent error and z‐test with a significance value of [Formula: see text]. The FWHMs from each phantom were not significantly different [Formula: see text] with an average error of 3.5%. The FWHMs from SSPs generated from the mean value were statistically different [Formula: see text]. The FWHMs from the different FWHM methods were not statistically different [Formula: see text]. Evaluation of the SSP is dependent on the ROI value used. The maximum value from the ROI should be used to generate the SSP whenever possible. SSP measurement is independent of the phantoms used in this study. PACS number: 87. John Wiley and Sons Inc. 2014-03-06 /pmc/articles/PMC5875475/ /pubmed/24710429 http://dx.doi.org/10.1120/jacmp.v15i2.4042 Text en © 2014 The Authors. This is an open access article under the terms of the http://creativecommons.org/licenses/by/3.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Greene, Travis C.
Rong, X. John
Evaluation of techniques for slice sensitivity profile measurement and analysis
title Evaluation of techniques for slice sensitivity profile measurement and analysis
title_full Evaluation of techniques for slice sensitivity profile measurement and analysis
title_fullStr Evaluation of techniques for slice sensitivity profile measurement and analysis
title_full_unstemmed Evaluation of techniques for slice sensitivity profile measurement and analysis
title_short Evaluation of techniques for slice sensitivity profile measurement and analysis
title_sort evaluation of techniques for slice sensitivity profile measurement and analysis
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875475/
https://www.ncbi.nlm.nih.gov/pubmed/24710429
http://dx.doi.org/10.1120/jacmp.v15i2.4042
work_keys_str_mv AT greenetravisc evaluationoftechniquesforslicesensitivityprofilemeasurementandanalysis
AT rongxjohn evaluationoftechniquesforslicesensitivityprofilemeasurementandanalysis