Brachytherapy dose‐volume histogram commissioning with multiple planning systems
The first quality assurance process for validating dose‐volume histogram data involving brachytherapy procedures in radiation therapy is presented. The process is demonstrated using both low dose‐rate and high dose‐rate radionuclide sources. A rectangular cuboid was contoured in five commercially av...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875493/ https://www.ncbi.nlm.nih.gov/pubmed/24710449 http://dx.doi.org/10.1120/jacmp.v15i2.4620 |
_version_ | 1783310355863175168 |
---|---|
author | Gossman, Michael S. Hancock, Samuel S. Kudchadker, Rajat J. Lundahl, Paul R. Cao, Minsong Melhus, Christopher S. |
author_facet | Gossman, Michael S. Hancock, Samuel S. Kudchadker, Rajat J. Lundahl, Paul R. Cao, Minsong Melhus, Christopher S. |
author_sort | Gossman, Michael S. |
collection | PubMed |
description | The first quality assurance process for validating dose‐volume histogram data involving brachytherapy procedures in radiation therapy is presented. The process is demonstrated using both low dose‐rate and high dose‐rate radionuclide sources. A rectangular cuboid was contoured in five commercially available brachytherapy treatment planning systems. A single radioactive source commissioned for QA testing was positioned coplanar and concentric with one end. Using the brachytherapy dosimetry formalism defined in the AAPM Task Group 43 report series, calculations were performed to estimate dose deposition in partial volumes of the cuboid structure. The point‐source approximation was used for a [Formula: see text] source and the line‐source approximation was used for a [Formula: see text] source in simulated permanent and temporary implants, respectively. Hand‐calculated, dose‐volume results were compared to TPS‐generated, dose‐volume histogram (DVH) data to ascertain acceptance. The average disagreement observed between hand calculations and the treatment planning system DVH was less than 1% for the five treatment planning systems and less than 5% for [Formula: see text]. A reproducible method for verifying the accuracy of volumetric statistics from a radiation therapy TPS can be employed. The process satisfies QA requirements for TPS commissioning, upgrading, and annual testing. We suggest that investigations be performed if the [Formula: see text] “actual variance” calculations differ by more than 5% at any specific radial distance with respect to [Formula: see text] , or if the “average variance” DVH [Formula: see text] calculations differ by more than 2% over all radial distances with respect to [Formula: see text]. PACS numbers: 87.10.+e, 87.55.‐x, 87.53.Jw, 07.05.Tp |
format | Online Article Text |
id | pubmed-5875493 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-58754932018-04-02 Brachytherapy dose‐volume histogram commissioning with multiple planning systems Gossman, Michael S. Hancock, Samuel S. Kudchadker, Rajat J. Lundahl, Paul R. Cao, Minsong Melhus, Christopher S. J Appl Clin Med Phys Radiation Oncology Physics The first quality assurance process for validating dose‐volume histogram data involving brachytherapy procedures in radiation therapy is presented. The process is demonstrated using both low dose‐rate and high dose‐rate radionuclide sources. A rectangular cuboid was contoured in five commercially available brachytherapy treatment planning systems. A single radioactive source commissioned for QA testing was positioned coplanar and concentric with one end. Using the brachytherapy dosimetry formalism defined in the AAPM Task Group 43 report series, calculations were performed to estimate dose deposition in partial volumes of the cuboid structure. The point‐source approximation was used for a [Formula: see text] source and the line‐source approximation was used for a [Formula: see text] source in simulated permanent and temporary implants, respectively. Hand‐calculated, dose‐volume results were compared to TPS‐generated, dose‐volume histogram (DVH) data to ascertain acceptance. The average disagreement observed between hand calculations and the treatment planning system DVH was less than 1% for the five treatment planning systems and less than 5% for [Formula: see text]. A reproducible method for verifying the accuracy of volumetric statistics from a radiation therapy TPS can be employed. The process satisfies QA requirements for TPS commissioning, upgrading, and annual testing. We suggest that investigations be performed if the [Formula: see text] “actual variance” calculations differ by more than 5% at any specific radial distance with respect to [Formula: see text] , or if the “average variance” DVH [Formula: see text] calculations differ by more than 2% over all radial distances with respect to [Formula: see text]. PACS numbers: 87.10.+e, 87.55.‐x, 87.53.Jw, 07.05.Tp John Wiley and Sons Inc. 2014-03-06 /pmc/articles/PMC5875493/ /pubmed/24710449 http://dx.doi.org/10.1120/jacmp.v15i2.4620 Text en © 2014 The Authors. This is an open access article under the terms of the http://creativecommons.org/licenses/by/3.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Gossman, Michael S. Hancock, Samuel S. Kudchadker, Rajat J. Lundahl, Paul R. Cao, Minsong Melhus, Christopher S. Brachytherapy dose‐volume histogram commissioning with multiple planning systems |
title | Brachytherapy dose‐volume histogram commissioning with multiple planning systems |
title_full | Brachytherapy dose‐volume histogram commissioning with multiple planning systems |
title_fullStr | Brachytherapy dose‐volume histogram commissioning with multiple planning systems |
title_full_unstemmed | Brachytherapy dose‐volume histogram commissioning with multiple planning systems |
title_short | Brachytherapy dose‐volume histogram commissioning with multiple planning systems |
title_sort | brachytherapy dose‐volume histogram commissioning with multiple planning systems |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875493/ https://www.ncbi.nlm.nih.gov/pubmed/24710449 http://dx.doi.org/10.1120/jacmp.v15i2.4620 |
work_keys_str_mv | AT gossmanmichaels brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems AT hancocksamuels brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems AT kudchadkerrajatj brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems AT lundahlpaulr brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems AT caominsong brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems AT melhuschristophers brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems |