Brachytherapy dose‐volume histogram commissioning with multiple planning systems

The first quality assurance process for validating dose‐volume histogram data involving brachytherapy procedures in radiation therapy is presented. The process is demonstrated using both low dose‐rate and high dose‐rate radionuclide sources. A rectangular cuboid was contoured in five commercially av...

Descripción completa

Detalles Bibliográficos
Autores principales: Gossman, Michael S., Hancock, Samuel S., Kudchadker, Rajat J., Lundahl, Paul R., Cao, Minsong, Melhus, Christopher S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875493/
https://www.ncbi.nlm.nih.gov/pubmed/24710449
http://dx.doi.org/10.1120/jacmp.v15i2.4620
_version_ 1783310355863175168
author Gossman, Michael S.
Hancock, Samuel S.
Kudchadker, Rajat J.
Lundahl, Paul R.
Cao, Minsong
Melhus, Christopher S.
author_facet Gossman, Michael S.
Hancock, Samuel S.
Kudchadker, Rajat J.
Lundahl, Paul R.
Cao, Minsong
Melhus, Christopher S.
author_sort Gossman, Michael S.
collection PubMed
description The first quality assurance process for validating dose‐volume histogram data involving brachytherapy procedures in radiation therapy is presented. The process is demonstrated using both low dose‐rate and high dose‐rate radionuclide sources. A rectangular cuboid was contoured in five commercially available brachytherapy treatment planning systems. A single radioactive source commissioned for QA testing was positioned coplanar and concentric with one end. Using the brachytherapy dosimetry formalism defined in the AAPM Task Group 43 report series, calculations were performed to estimate dose deposition in partial volumes of the cuboid structure. The point‐source approximation was used for a [Formula: see text] source and the line‐source approximation was used for a [Formula: see text] source in simulated permanent and temporary implants, respectively. Hand‐calculated, dose‐volume results were compared to TPS‐generated, dose‐volume histogram (DVH) data to ascertain acceptance. The average disagreement observed between hand calculations and the treatment planning system DVH was less than 1% for the five treatment planning systems and less than 5% for [Formula: see text]. A reproducible method for verifying the accuracy of volumetric statistics from a radiation therapy TPS can be employed. The process satisfies QA requirements for TPS commissioning, upgrading, and annual testing. We suggest that investigations be performed if the [Formula: see text] “actual variance” calculations differ by more than 5% at any specific radial distance with respect to [Formula: see text] , or if the “average variance” DVH [Formula: see text] calculations differ by more than 2% over all radial distances with respect to [Formula: see text]. PACS numbers: 87.10.+e, 87.55.‐x, 87.53.Jw, 07.05.Tp
format Online
Article
Text
id pubmed-5875493
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58754932018-04-02 Brachytherapy dose‐volume histogram commissioning with multiple planning systems Gossman, Michael S. Hancock, Samuel S. Kudchadker, Rajat J. Lundahl, Paul R. Cao, Minsong Melhus, Christopher S. J Appl Clin Med Phys Radiation Oncology Physics The first quality assurance process for validating dose‐volume histogram data involving brachytherapy procedures in radiation therapy is presented. The process is demonstrated using both low dose‐rate and high dose‐rate radionuclide sources. A rectangular cuboid was contoured in five commercially available brachytherapy treatment planning systems. A single radioactive source commissioned for QA testing was positioned coplanar and concentric with one end. Using the brachytherapy dosimetry formalism defined in the AAPM Task Group 43 report series, calculations were performed to estimate dose deposition in partial volumes of the cuboid structure. The point‐source approximation was used for a [Formula: see text] source and the line‐source approximation was used for a [Formula: see text] source in simulated permanent and temporary implants, respectively. Hand‐calculated, dose‐volume results were compared to TPS‐generated, dose‐volume histogram (DVH) data to ascertain acceptance. The average disagreement observed between hand calculations and the treatment planning system DVH was less than 1% for the five treatment planning systems and less than 5% for [Formula: see text]. A reproducible method for verifying the accuracy of volumetric statistics from a radiation therapy TPS can be employed. The process satisfies QA requirements for TPS commissioning, upgrading, and annual testing. We suggest that investigations be performed if the [Formula: see text] “actual variance” calculations differ by more than 5% at any specific radial distance with respect to [Formula: see text] , or if the “average variance” DVH [Formula: see text] calculations differ by more than 2% over all radial distances with respect to [Formula: see text]. PACS numbers: 87.10.+e, 87.55.‐x, 87.53.Jw, 07.05.Tp John Wiley and Sons Inc. 2014-03-06 /pmc/articles/PMC5875493/ /pubmed/24710449 http://dx.doi.org/10.1120/jacmp.v15i2.4620 Text en © 2014 The Authors. This is an open access article under the terms of the http://creativecommons.org/licenses/by/3.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Gossman, Michael S.
Hancock, Samuel S.
Kudchadker, Rajat J.
Lundahl, Paul R.
Cao, Minsong
Melhus, Christopher S.
Brachytherapy dose‐volume histogram commissioning with multiple planning systems
title Brachytherapy dose‐volume histogram commissioning with multiple planning systems
title_full Brachytherapy dose‐volume histogram commissioning with multiple planning systems
title_fullStr Brachytherapy dose‐volume histogram commissioning with multiple planning systems
title_full_unstemmed Brachytherapy dose‐volume histogram commissioning with multiple planning systems
title_short Brachytherapy dose‐volume histogram commissioning with multiple planning systems
title_sort brachytherapy dose‐volume histogram commissioning with multiple planning systems
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875493/
https://www.ncbi.nlm.nih.gov/pubmed/24710449
http://dx.doi.org/10.1120/jacmp.v15i2.4620
work_keys_str_mv AT gossmanmichaels brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems
AT hancocksamuels brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems
AT kudchadkerrajatj brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems
AT lundahlpaulr brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems
AT caominsong brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems
AT melhuschristophers brachytherapydosevolumehistogramcommissioningwithmultipleplanningsystems