Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy

The measurement of dose distributions in clinical brachytherapy, for the purpose of quality control, commissioning or dosimetric audit, is challenging and requires development. Radiochromic film dosimetry with a commercial flatbed scanner may be suitable, but careful methodologies are required to co...

Descripción completa

Detalles Bibliográficos
Autores principales: Palmer, Antony L, Bradley, David, Nisbet, Andrew
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875501/
https://www.ncbi.nlm.nih.gov/pubmed/25207417
http://dx.doi.org/10.1120/jacmp.v15i4.4854
_version_ 1783310357500002304
author Palmer, Antony L
Bradley, David
Nisbet, Andrew
author_facet Palmer, Antony L
Bradley, David
Nisbet, Andrew
author_sort Palmer, Antony L
collection PubMed
description The measurement of dose distributions in clinical brachytherapy, for the purpose of quality control, commissioning or dosimetric audit, is challenging and requires development. Radiochromic film dosimetry with a commercial flatbed scanner may be suitable, but careful methodologies are required to control various sources of uncertainty. Triple‐channel dosimetry has recently been utilized in external beam radiotherapy to improve the accuracy of film dosimetry, but its use in brachytherapy, with characteristic high maximum doses, steep dose gradients, and small scales, has been less well researched. We investigate the use of advanced film dosimetry techniques for brachytherapy dosimetry, evaluating uncertainties and assessing the mitigation afforded by triple‐channel dosimetry. We present results on postirradiation film darkening, lateral scanner effect, film surface perturbation, film active layer thickness, film curling, and examples of the measurement of clinical brachytherapy dose distributions. The lateral scanner effect in brachytherapy film dosimetry can be very significant, up to 23% dose increase at 14 Gy, at ± 9 cm lateral from the scanner axis for simple single‐channel dosimetry. Triple‐channel dosimetry mitigates the effect, but still limits the useable width of a typical scanner to less than 8 cm at high dose levels to give dose uncertainty to within 1%. Triple‐channel dosimetry separates dose and dose‐independent signal components, and effectively removes disturbances caused by film thickness variation and surface perturbations in the examples considered in this work. The use of reference dose films scanned simultaneously with brachytherapy test films is recommended to account for scanner variations from calibration conditions. Postirradiation darkening, which is a continual logarithmic function with time, must be taken into account between the reference and test films. Finally, films must be flat when scanned to avoid the Callier‐like effects and to provide reliable dosimetric results. We have demonstrated that radiochromic film dosimetry with GAFCHROMIC EBT3 film and a commercial flatbed scanner is a viable method for brachytherapy dose distribution measurement, and uncertainties may be reduced with triple‐channel dosimetry and specific film scan and evaluation methodologies. PACS numbers: 87.55.Qr, 87.56.bg, 87.55.km
format Online
Article
Text
id pubmed-5875501
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58755012018-04-02 Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy Palmer, Antony L Bradley, David Nisbet, Andrew J Appl Clin Med Phys Radiation Oncology Physics The measurement of dose distributions in clinical brachytherapy, for the purpose of quality control, commissioning or dosimetric audit, is challenging and requires development. Radiochromic film dosimetry with a commercial flatbed scanner may be suitable, but careful methodologies are required to control various sources of uncertainty. Triple‐channel dosimetry has recently been utilized in external beam radiotherapy to improve the accuracy of film dosimetry, but its use in brachytherapy, with characteristic high maximum doses, steep dose gradients, and small scales, has been less well researched. We investigate the use of advanced film dosimetry techniques for brachytherapy dosimetry, evaluating uncertainties and assessing the mitigation afforded by triple‐channel dosimetry. We present results on postirradiation film darkening, lateral scanner effect, film surface perturbation, film active layer thickness, film curling, and examples of the measurement of clinical brachytherapy dose distributions. The lateral scanner effect in brachytherapy film dosimetry can be very significant, up to 23% dose increase at 14 Gy, at ± 9 cm lateral from the scanner axis for simple single‐channel dosimetry. Triple‐channel dosimetry mitigates the effect, but still limits the useable width of a typical scanner to less than 8 cm at high dose levels to give dose uncertainty to within 1%. Triple‐channel dosimetry separates dose and dose‐independent signal components, and effectively removes disturbances caused by film thickness variation and surface perturbations in the examples considered in this work. The use of reference dose films scanned simultaneously with brachytherapy test films is recommended to account for scanner variations from calibration conditions. Postirradiation darkening, which is a continual logarithmic function with time, must be taken into account between the reference and test films. Finally, films must be flat when scanned to avoid the Callier‐like effects and to provide reliable dosimetric results. We have demonstrated that radiochromic film dosimetry with GAFCHROMIC EBT3 film and a commercial flatbed scanner is a viable method for brachytherapy dose distribution measurement, and uncertainties may be reduced with triple‐channel dosimetry and specific film scan and evaluation methodologies. PACS numbers: 87.55.Qr, 87.56.bg, 87.55.km John Wiley and Sons Inc. 2014-07-08 /pmc/articles/PMC5875501/ /pubmed/25207417 http://dx.doi.org/10.1120/jacmp.v15i4.4854 Text en © 2014 The Authors. This is an open access article under the terms of the http://creativecommons.org/licenses/by/3.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Palmer, Antony L
Bradley, David
Nisbet, Andrew
Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy
title Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy
title_full Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy
title_fullStr Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy
title_full_unstemmed Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy
title_short Evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy
title_sort evaluation and implementation of triple‐channel radiochromic film dosimetry in brachytherapy
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875501/
https://www.ncbi.nlm.nih.gov/pubmed/25207417
http://dx.doi.org/10.1120/jacmp.v15i4.4854
work_keys_str_mv AT palmerantonyl evaluationandimplementationoftriplechannelradiochromicfilmdosimetryinbrachytherapy
AT bradleydavid evaluationandimplementationoftriplechannelradiochromicfilmdosimetryinbrachytherapy
AT nisbetandrew evaluationandimplementationoftriplechannelradiochromicfilmdosimetryinbrachytherapy