Intra‐ and intervariability in beam data commissioning among water phantom scanning systems
Accurate beam data acquisition during commissioning is essential for modeling the treatment planning system and dose calculation in radiotherapy. Although currently several commercial scanning systems are available, there is no report that compared the differences among the systems because most inst...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875503/ https://www.ncbi.nlm.nih.gov/pubmed/25207415 http://dx.doi.org/10.1120/jacmp.v15i4.4850 |
_version_ | 1783310357958230016 |
---|---|
author | Akino, Yuichi Gibbons, John P. Neck, Daniel W. Chu, Connel Das, Indra J. |
author_facet | Akino, Yuichi Gibbons, John P. Neck, Daniel W. Chu, Connel Das, Indra J. |
author_sort | Akino, Yuichi |
collection | PubMed |
description | Accurate beam data acquisition during commissioning is essential for modeling the treatment planning system and dose calculation in radiotherapy. Although currently several commercial scanning systems are available, there is no report that compared the differences among the systems because most institutions do not acquire several scanning systems due to the high cost, storage space, and infrequent usage. In this report, we demonstrate the intra‐ and intervariability of beam profiles measured with four commercial scanning systems. During a recent educational and training workshop, four different vendors of beam scanning water phantoms were invited to demonstrate the operation and data collection of their systems. Systems were set up utilizing vendor‐recommended protocols and were operated with a senior physicist, who was assigned as an instructor along with vendor. During the training sessions, each group was asked to measure beam parameters, and the intravariability in percent depth dose (PDD). At the end of the day, the profile of one linear accelerator was measured with each system to evaluate intervariability. Relatively very small (SD < 0.12%) intervariability in PDD was observed among four systems at a region deeper than peak (1.5 cm). All systems showed almost identical profiles. At the area within 80% of radiation field, the average, and maximum differences were within ± 0.35% and 0.80%, respectively, compared to arbitrarily chosen IBA system as reference. In the penumbrae region, the distance to agreement (DTA) of the region where dose difference exceed ± 1% was less than 1 mm. Repeated PDD measurement showed small intravariability with SD < 0.5%, although large SD was observed in the buildup region. All four water phantom scanning systems demonstrated adequate accuracy for beam data collection (i.e., within 1% of dose difference or 1 mm of DTA among each other). It is concluded that every system is capable of acquiring accurate beam. Thus the selection of a water scanning system should be based on institutional comfort, personal preference of software and hardware, and financial consideration. PACS number: 87.53.Bn |
format | Online Article Text |
id | pubmed-5875503 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-58755032018-04-02 Intra‐ and intervariability in beam data commissioning among water phantom scanning systems Akino, Yuichi Gibbons, John P. Neck, Daniel W. Chu, Connel Das, Indra J. J Appl Clin Med Phys Radiation Oncology Physics Accurate beam data acquisition during commissioning is essential for modeling the treatment planning system and dose calculation in radiotherapy. Although currently several commercial scanning systems are available, there is no report that compared the differences among the systems because most institutions do not acquire several scanning systems due to the high cost, storage space, and infrequent usage. In this report, we demonstrate the intra‐ and intervariability of beam profiles measured with four commercial scanning systems. During a recent educational and training workshop, four different vendors of beam scanning water phantoms were invited to demonstrate the operation and data collection of their systems. Systems were set up utilizing vendor‐recommended protocols and were operated with a senior physicist, who was assigned as an instructor along with vendor. During the training sessions, each group was asked to measure beam parameters, and the intravariability in percent depth dose (PDD). At the end of the day, the profile of one linear accelerator was measured with each system to evaluate intervariability. Relatively very small (SD < 0.12%) intervariability in PDD was observed among four systems at a region deeper than peak (1.5 cm). All systems showed almost identical profiles. At the area within 80% of radiation field, the average, and maximum differences were within ± 0.35% and 0.80%, respectively, compared to arbitrarily chosen IBA system as reference. In the penumbrae region, the distance to agreement (DTA) of the region where dose difference exceed ± 1% was less than 1 mm. Repeated PDD measurement showed small intravariability with SD < 0.5%, although large SD was observed in the buildup region. All four water phantom scanning systems demonstrated adequate accuracy for beam data collection (i.e., within 1% of dose difference or 1 mm of DTA among each other). It is concluded that every system is capable of acquiring accurate beam. Thus the selection of a water scanning system should be based on institutional comfort, personal preference of software and hardware, and financial consideration. PACS number: 87.53.Bn John Wiley and Sons Inc. 2014-07-08 /pmc/articles/PMC5875503/ /pubmed/25207415 http://dx.doi.org/10.1120/jacmp.v15i4.4850 Text en © 2014 The Authors. This is an open access article under the terms of the http://creativecommons.org/licenses/by/3.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Akino, Yuichi Gibbons, John P. Neck, Daniel W. Chu, Connel Das, Indra J. Intra‐ and intervariability in beam data commissioning among water phantom scanning systems |
title | Intra‐ and intervariability in beam data commissioning among water phantom scanning systems |
title_full | Intra‐ and intervariability in beam data commissioning among water phantom scanning systems |
title_fullStr | Intra‐ and intervariability in beam data commissioning among water phantom scanning systems |
title_full_unstemmed | Intra‐ and intervariability in beam data commissioning among water phantom scanning systems |
title_short | Intra‐ and intervariability in beam data commissioning among water phantom scanning systems |
title_sort | intra‐ and intervariability in beam data commissioning among water phantom scanning systems |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875503/ https://www.ncbi.nlm.nih.gov/pubmed/25207415 http://dx.doi.org/10.1120/jacmp.v15i4.4850 |
work_keys_str_mv | AT akinoyuichi intraandintervariabilityinbeamdatacommissioningamongwaterphantomscanningsystems AT gibbonsjohnp intraandintervariabilityinbeamdatacommissioningamongwaterphantomscanningsystems AT neckdanielw intraandintervariabilityinbeamdatacommissioningamongwaterphantomscanningsystems AT chuconnel intraandintervariabilityinbeamdatacommissioningamongwaterphantomscanningsystems AT dasindraj intraandintervariabilityinbeamdatacommissioningamongwaterphantomscanningsystems |