Comparison between electromagnetic transponders and radiographic imaging for prostate localization: A pelvic phantom study with rotations and translations

The aim of this study was to evaluate the differences in target localization between Calypso(®), kV orthogonal imaging and cone‐beam computed tomography (CBCT) for combined translations and rotations of an anthropomorphic pelvic phantom. The phantom was localized using all three systems in 50 differ...

Descripción completa

Detalles Bibliográficos
Autores principales: Hamilton, Daniel G., McKenzie, Dean P., Perkins, Anne E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875817/
https://www.ncbi.nlm.nih.gov/pubmed/28699243
http://dx.doi.org/10.1002/acm2.12119
_version_ 1783310418778783744
author Hamilton, Daniel G.
McKenzie, Dean P.
Perkins, Anne E.
author_facet Hamilton, Daniel G.
McKenzie, Dean P.
Perkins, Anne E.
author_sort Hamilton, Daniel G.
collection PubMed
description The aim of this study was to evaluate the differences in target localization between Calypso(®), kV orthogonal imaging and cone‐beam computed tomography (CBCT) for combined translations and rotations of an anthropomorphic pelvic phantom. The phantom was localized using all three systems in 50 different positions, with applied translational and rotational offsets randomly sampled from representative normal distributions of prostate motion. Lin's concordance correlation coefficient (ρc) and 95% confidence intervals were calculated to assess the agreement between the localization systems. Mean differences and difference vectors between the three systems were also calculated. Agreement between systems for lateral, vertical, and longitudinal translations was excellent, with ρc values of greater than 0.98 between all three systems in all axes. There was excellent agreement between the systems for rotations around the lateral axis (pitch) (ρc > 0.99), and around the vertical axis (yaw) (ρc > 0.97). However, somewhat poorer agreement for rotations around the longitudinal axis (roll) was observed, with the lowest correlation observed between Calypso and kV orthogonal imaging (ρc = 0.895). Mean differences between the phantom position reported by Calypso and the radiographic systems were less than 1 mm and 1° for all translations and rotations. The results for translations are consistent with the publications of previous authors. There is no comparable published data for rotations. While there is lower correlation between the three systems for roll than for the other angles, the mean differences in reported rotations are not clinically significant.
format Online
Article
Text
id pubmed-5875817
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58758172018-04-02 Comparison between electromagnetic transponders and radiographic imaging for prostate localization: A pelvic phantom study with rotations and translations Hamilton, Daniel G. McKenzie, Dean P. Perkins, Anne E. J Appl Clin Med Phys Radiation Oncology Physics The aim of this study was to evaluate the differences in target localization between Calypso(®), kV orthogonal imaging and cone‐beam computed tomography (CBCT) for combined translations and rotations of an anthropomorphic pelvic phantom. The phantom was localized using all three systems in 50 different positions, with applied translational and rotational offsets randomly sampled from representative normal distributions of prostate motion. Lin's concordance correlation coefficient (ρc) and 95% confidence intervals were calculated to assess the agreement between the localization systems. Mean differences and difference vectors between the three systems were also calculated. Agreement between systems for lateral, vertical, and longitudinal translations was excellent, with ρc values of greater than 0.98 between all three systems in all axes. There was excellent agreement between the systems for rotations around the lateral axis (pitch) (ρc > 0.99), and around the vertical axis (yaw) (ρc > 0.97). However, somewhat poorer agreement for rotations around the longitudinal axis (roll) was observed, with the lowest correlation observed between Calypso and kV orthogonal imaging (ρc = 0.895). Mean differences between the phantom position reported by Calypso and the radiographic systems were less than 1 mm and 1° for all translations and rotations. The results for translations are consistent with the publications of previous authors. There is no comparable published data for rotations. While there is lower correlation between the three systems for roll than for the other angles, the mean differences in reported rotations are not clinically significant. John Wiley and Sons Inc. 2017-07-12 /pmc/articles/PMC5875817/ /pubmed/28699243 http://dx.doi.org/10.1002/acm2.12119 Text en © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Hamilton, Daniel G.
McKenzie, Dean P.
Perkins, Anne E.
Comparison between electromagnetic transponders and radiographic imaging for prostate localization: A pelvic phantom study with rotations and translations
title Comparison between electromagnetic transponders and radiographic imaging for prostate localization: A pelvic phantom study with rotations and translations
title_full Comparison between electromagnetic transponders and radiographic imaging for prostate localization: A pelvic phantom study with rotations and translations
title_fullStr Comparison between electromagnetic transponders and radiographic imaging for prostate localization: A pelvic phantom study with rotations and translations
title_full_unstemmed Comparison between electromagnetic transponders and radiographic imaging for prostate localization: A pelvic phantom study with rotations and translations
title_short Comparison between electromagnetic transponders and radiographic imaging for prostate localization: A pelvic phantom study with rotations and translations
title_sort comparison between electromagnetic transponders and radiographic imaging for prostate localization: a pelvic phantom study with rotations and translations
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875817/
https://www.ncbi.nlm.nih.gov/pubmed/28699243
http://dx.doi.org/10.1002/acm2.12119
work_keys_str_mv AT hamiltondanielg comparisonbetweenelectromagnetictranspondersandradiographicimagingforprostatelocalizationapelvicphantomstudywithrotationsandtranslations
AT mckenziedeanp comparisonbetweenelectromagnetictranspondersandradiographicimagingforprostatelocalizationapelvicphantomstudywithrotationsandtranslations
AT perkinsannee comparisonbetweenelectromagnetictranspondersandradiographicimagingforprostatelocalizationapelvicphantomstudywithrotationsandtranslations