Cargando…
A review of nonstandardized applicators digitization in Nucletron™ HDR procedures
The major errors in HDR procedures were failures to enter the correct treatment distance, which could be caused by either entering wrong transmission lengths or imprecisely digitizing the dwelling positions. Most of those errors were not easily avoidable by enhancing the HDR management level because...
Autores principales: | Wang, Kelin, Ferenci, Michele S., de la Zerda, Alberto, Padgett, Kyle R., Bossart, Elizabeth L., Chao, Ming, Shao, Hua, Zhang, Mutain |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5875818/ https://www.ncbi.nlm.nih.gov/pubmed/28857433 http://dx.doi.org/10.1002/acm2.12156 |
Ejemplares similares
-
A technical evaluation of the Nucletron FIRST system: Conformance of a remote afterloading brachytherapy seed implantation system to manufacturer specifications and AAPM Task Group report recommendations
por: Rivard, Mark J., et al.
Publicado: (2005) -
Varian HDR surface applicators — commissioning and clinical implementation
por: Iftimia, Ileana, et al.
Publicado: (2016) -
A fast, independent dose check of HDR plans
por: Lachaine, Martin E., et al.
Publicado: (2003) -
Comparison of two planning systems for HDR brachytherapy gynecological application
por: Elhanafy, Osman A., et al.
Publicado: (2001) -
Development of an HDR‐BT QA tool for source position verification
por: Kumazaki, Yu, et al.
Publicado: (2020)