PURPOSE: A Monte Carlo model of a 6 MV medical linear accelerator (linac) unit built indigenously was developed using the BEAMnrc user code of the EGSnrc code system. The model was benchmarked against the measurements. Monte Carlo simulations were carried out for different incident electron beam par...

Descripción completa

Detalles Bibliográficos
Autores principales: Mishra, Subhalaxmi, Dixit, P. K., Selvam, T. Palani, Yavalkar, Sanket S., Deshpande, D. D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5879818/
https://www.ncbi.nlm.nih.gov/pubmed/29628627
http://dx.doi.org/10.4103/jmp.JMP_125_17
_version_ 1783311057009246208
author Mishra, Subhalaxmi
Dixit, P. K.
Selvam, T. Palani
Yavalkar, Sanket S.
Deshpande, D. D.
author_facet Mishra, Subhalaxmi
Dixit, P. K.
Selvam, T. Palani
Yavalkar, Sanket S.
Deshpande, D. D.
author_sort Mishra, Subhalaxmi
collection PubMed
description PURPOSE: A Monte Carlo model of a 6 MV medical linear accelerator (linac) unit built indigenously was developed using the BEAMnrc user code of the EGSnrc code system. The model was benchmarked against the measurements. Monte Carlo simulations were carried out for different incident electron beam parameters in the study. MATERIALS AND METHODS: Simulation of indigenously developed linac unit has been carried out using the Monte Carlo based BEAMnrc user-code of the EGSnrc code system. Using the model, percentage depth dose (PDD), and lateral dose profiles were studied using the DOSXYZnrc user code. To identify appropriate electron parameters, three different distributions of electron beam intensity were investigated. For each case, the kinetic energy of the incident electron was varied from 6 to 6.5 MeV (0.1 MeV increment). The calculated dose data were compared against the measurements using the PTW, Germany make RFA dosimetric system (water tank MP3-M and 0.125 cm(3) ion chamber). RESULTS: The best fit of incident electron beam parameter was found for the combination of beam energy of 6.2 MeV and circular Gaussian distributed source in X and Y with FWHM of 1.0 mm. PDD and beam profiles (along both X and Y directions) were calculated for the field sizes from 5 cm × 5 cm to 25 cm × 25 cm. The dose difference between the calculated and measured PDD and profile values were under 1%, except for the penumbra region where the maximum deviation was found to be around 2%. CONCLUSIONS: A Monte Carlo model of indigenous linac (6 MV) has been developed and benchmarked against the measured data.
format Online
Article
Text
id pubmed-5879818
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-58798182018-04-06 Monte Carlo Investigation of Photon Beam Characteristics and its Variation with Incident Electron Beam Parameters for Indigenous Medical Linear Accelerator Mishra, Subhalaxmi Dixit, P. K. Selvam, T. Palani Yavalkar, Sanket S. Deshpande, D. D. J Med Phys Original Article PURPOSE: A Monte Carlo model of a 6 MV medical linear accelerator (linac) unit built indigenously was developed using the BEAMnrc user code of the EGSnrc code system. The model was benchmarked against the measurements. Monte Carlo simulations were carried out for different incident electron beam parameters in the study. MATERIALS AND METHODS: Simulation of indigenously developed linac unit has been carried out using the Monte Carlo based BEAMnrc user-code of the EGSnrc code system. Using the model, percentage depth dose (PDD), and lateral dose profiles were studied using the DOSXYZnrc user code. To identify appropriate electron parameters, three different distributions of electron beam intensity were investigated. For each case, the kinetic energy of the incident electron was varied from 6 to 6.5 MeV (0.1 MeV increment). The calculated dose data were compared against the measurements using the PTW, Germany make RFA dosimetric system (water tank MP3-M and 0.125 cm(3) ion chamber). RESULTS: The best fit of incident electron beam parameter was found for the combination of beam energy of 6.2 MeV and circular Gaussian distributed source in X and Y with FWHM of 1.0 mm. PDD and beam profiles (along both X and Y directions) were calculated for the field sizes from 5 cm × 5 cm to 25 cm × 25 cm. The dose difference between the calculated and measured PDD and profile values were under 1%, except for the penumbra region where the maximum deviation was found to be around 2%. CONCLUSIONS: A Monte Carlo model of indigenous linac (6 MV) has been developed and benchmarked against the measured data. Medknow Publications & Media Pvt Ltd 2018 /pmc/articles/PMC5879818/ /pubmed/29628627 http://dx.doi.org/10.4103/jmp.JMP_125_17 Text en Copyright: © 2018 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
spellingShingle Original Article
Mishra, Subhalaxmi
Dixit, P. K.
Selvam, T. Palani
Yavalkar, Sanket S.
Deshpande, D. D.
Monte Carlo Investigation of Photon Beam Characteristics and its Variation with Incident Electron Beam Parameters for Indigenous Medical Linear Accelerator
title Monte Carlo Investigation of Photon Beam Characteristics and its Variation with Incident Electron Beam Parameters for Indigenous Medical Linear Accelerator
title_full Monte Carlo Investigation of Photon Beam Characteristics and its Variation with Incident Electron Beam Parameters for Indigenous Medical Linear Accelerator
title_fullStr Monte Carlo Investigation of Photon Beam Characteristics and its Variation with Incident Electron Beam Parameters for Indigenous Medical Linear Accelerator
title_full_unstemmed Monte Carlo Investigation of Photon Beam Characteristics and its Variation with Incident Electron Beam Parameters for Indigenous Medical Linear Accelerator
title_short Monte Carlo Investigation of Photon Beam Characteristics and its Variation with Incident Electron Beam Parameters for Indigenous Medical Linear Accelerator
title_sort monte carlo investigation of photon beam characteristics and its variation with incident electron beam parameters for indigenous medical linear accelerator
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5879818/
https://www.ncbi.nlm.nih.gov/pubmed/29628627
http://dx.doi.org/10.4103/jmp.JMP_125_17
work_keys_str_mv AT mishrasubhalaxmi montecarloinvestigationofphotonbeamcharacteristicsanditsvariationwithincidentelectronbeamparametersforindigenousmedicallinearaccelerator
AT dixitpk montecarloinvestigationofphotonbeamcharacteristicsanditsvariationwithincidentelectronbeamparametersforindigenousmedicallinearaccelerator
AT selvamtpalani montecarloinvestigationofphotonbeamcharacteristicsanditsvariationwithincidentelectronbeamparametersforindigenousmedicallinearaccelerator
AT yavalkarsankets montecarloinvestigationofphotonbeamcharacteristicsanditsvariationwithincidentelectronbeamparametersforindigenousmedicallinearaccelerator
AT deshpandedd montecarloinvestigationofphotonbeamcharacteristicsanditsvariationwithincidentelectronbeamparametersforindigenousmedicallinearaccelerator