Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software

This study reports on our experience with the in-vivo dose verification software, EPIgray® (DOSIsoft, Cachan, France). After the initial commissioning process, clinical experiments on phantom treatments were evaluated to assess the level of accuracy of the electronic portal imaging device (EPID) bas...

Descripción completa

Detalles Bibliográficos
Autores principales: Held, Mareike, Cheung, Joey, Perez Andujar, Angelica, Husson, François, Morin, Olivier
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5880591/
https://www.ncbi.nlm.nih.gov/pubmed/29632749
http://dx.doi.org/10.7759/cureus.2139
_version_ 1783311182021525504
author Held, Mareike
Cheung, Joey
Perez Andujar, Angelica
Husson, François
Morin, Olivier
author_facet Held, Mareike
Cheung, Joey
Perez Andujar, Angelica
Husson, François
Morin, Olivier
author_sort Held, Mareike
collection PubMed
description This study reports on our experience with the in-vivo dose verification software, EPIgray® (DOSIsoft, Cachan, France). After the initial commissioning process, clinical experiments on phantom treatments were evaluated to assess the level of accuracy of the electronic portal imaging device (EPID) based in-vivo dose verification. EPIgray was commissioned based on the company’s instructions. This involved ion chamber measurements and portal imaging of solid water blocks of various thicknesses between 5 and 35 cm. Field sizes varied between 2 x 2 cm(2) and 20 x 20 cm(2). The determined conversion factors were adjusted through an additional iterative process using treatment planning system calculations. Subsequently, evaluation was performed using treatment plans of single and opposed beams, as well as intensity modulated radiotherapy (IMRT) plans, based on recommendations from the task group report TG-119 to test for dose reconstruction accuracy. All tests were performed using blocks of solid water slabs as a phantom. For single square fields, the dose at isocenter was reconstructed within 3% accuracy in EPIgray compared to the treatment planning system dose. Similarly, the relative deviation of the total dose was accurately reconstructed within 3% for all IMRT plans with points placed inside a high-dose region near the isocenter. Predictions became less accurate than < 5% when the evaluation point was outside the treatment target. Dose at points 5 cm or more away from the isocenter or within an avoidance structure was reconstructed less reliably. EPIgray formalism accuracy is adequate for an efficient error detection system with verifications performed in high-dose volumes. It provides immediate intra-fractional feedback on the delivery of treatment plans without affecting the treatment beam. Besides the EPID, no additional hardware is required. The software evaluates local point dose measurements to verify treatment plan delivery and patient positioning within 5% accuracy, depending on the placement of evaluation points.
format Online
Article
Text
id pubmed-5880591
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-58805912018-04-09 Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software Held, Mareike Cheung, Joey Perez Andujar, Angelica Husson, François Morin, Olivier Cureus Medical Physics This study reports on our experience with the in-vivo dose verification software, EPIgray® (DOSIsoft, Cachan, France). After the initial commissioning process, clinical experiments on phantom treatments were evaluated to assess the level of accuracy of the electronic portal imaging device (EPID) based in-vivo dose verification. EPIgray was commissioned based on the company’s instructions. This involved ion chamber measurements and portal imaging of solid water blocks of various thicknesses between 5 and 35 cm. Field sizes varied between 2 x 2 cm(2) and 20 x 20 cm(2). The determined conversion factors were adjusted through an additional iterative process using treatment planning system calculations. Subsequently, evaluation was performed using treatment plans of single and opposed beams, as well as intensity modulated radiotherapy (IMRT) plans, based on recommendations from the task group report TG-119 to test for dose reconstruction accuracy. All tests were performed using blocks of solid water slabs as a phantom. For single square fields, the dose at isocenter was reconstructed within 3% accuracy in EPIgray compared to the treatment planning system dose. Similarly, the relative deviation of the total dose was accurately reconstructed within 3% for all IMRT plans with points placed inside a high-dose region near the isocenter. Predictions became less accurate than < 5% when the evaluation point was outside the treatment target. Dose at points 5 cm or more away from the isocenter or within an avoidance structure was reconstructed less reliably. EPIgray formalism accuracy is adequate for an efficient error detection system with verifications performed in high-dose volumes. It provides immediate intra-fractional feedback on the delivery of treatment plans without affecting the treatment beam. Besides the EPID, no additional hardware is required. The software evaluates local point dose measurements to verify treatment plan delivery and patient positioning within 5% accuracy, depending on the placement of evaluation points. Cureus 2018-02-02 /pmc/articles/PMC5880591/ /pubmed/29632749 http://dx.doi.org/10.7759/cureus.2139 Text en Copyright © 2018, Held et al. http://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Medical Physics
Held, Mareike
Cheung, Joey
Perez Andujar, Angelica
Husson, François
Morin, Olivier
Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software
title Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software
title_full Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software
title_fullStr Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software
title_full_unstemmed Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software
title_short Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software
title_sort commissioning and evaluation of an electronic portal imaging device-based in-vivo dosimetry software
topic Medical Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5880591/
https://www.ncbi.nlm.nih.gov/pubmed/29632749
http://dx.doi.org/10.7759/cureus.2139
work_keys_str_mv AT heldmareike commissioningandevaluationofanelectronicportalimagingdevicebasedinvivodosimetrysoftware
AT cheungjoey commissioningandevaluationofanelectronicportalimagingdevicebasedinvivodosimetrysoftware
AT perezandujarangelica commissioningandevaluationofanelectronicportalimagingdevicebasedinvivodosimetrysoftware
AT hussonfrancois commissioningandevaluationofanelectronicportalimagingdevicebasedinvivodosimetrysoftware
AT morinolivier commissioningandevaluationofanelectronicportalimagingdevicebasedinvivodosimetrysoftware