Cargando…

Comparative bench study evaluation of different infant interfaces for non-invasive ventilation

BACKGROUND: To compare, in terms of patient-ventilator interaction and performance, a new nasal mask (Respireo, AirLiquide, FR) with the Endotracheal tube (ET) and a commonly used nasal mask (FPM, Fisher and Paykel, NZ) for delivering Pressure Support Ventilation (PSV) in an infant model of Acute Re...

Descripción completa

Detalles Bibliográficos
Autores principales: Conti, Giorgio, Spinazzola, Giorgia, Gregoretti, Cesare, Ferrone, Giuliano, Cortegiani, Andrea, Festa, Olimpia, Piastra, Marco, Tortorolo, Luca, Costa, Roberta
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5889592/
https://www.ncbi.nlm.nih.gov/pubmed/29625596
http://dx.doi.org/10.1186/s12890-018-0620-x
Descripción
Sumario:BACKGROUND: To compare, in terms of patient-ventilator interaction and performance, a new nasal mask (Respireo, AirLiquide, FR) with the Endotracheal tube (ET) and a commonly used nasal mask (FPM, Fisher and Paykel, NZ) for delivering Pressure Support Ventilation (PSV) in an infant model of Acute Respiratory Failure (ARF). METHODS: An active test lung (ASL 5000) connected to an infant mannequin through 3 different interfaces (Respireo, ET and FPM), was ventilated with a standard ICU ventilator set in PSV. The test lung was set to simulate a 5.5 kg infant with ARF, breathing at 50 and 60 breaths/min). Non-invasive ventilation (NIV) mode was not used and the leaks were nearly zero. RESULTS: The ET showed the shortest inspiratory trigger delay and pressurization time compared to FPM and Respireo (p < 0.01). At each respiratory rate tested, the FPM showed the shortest Expiratory trigger delay compared to ET and Respireo (p < 0.01). The Respireo presented a lower value of Inspiratory pressure–time product and trigger pressure drop than ET (p < 0.01), while no significant difference was found in terms of pressure-time product at 300 and 500 ms. During all tests, compared with the FPM, ET showed a significantly higher tidal volume (V(T)) delivered (p < 0.01), while Respireo showed a trend toward an increase of tidal volume delivered compared with FPM. CONCLUSIONS: The ET showed a better patient-ventilator interaction and performance compared to both the nasal masks. Despite the higher internal volume, Respireo showed a trend toward an increase of the delivered tidal volume; globally, its efficiency in terms of patient-ventilator interaction was comparable to the FPM, which is the infant NIV mask characterized by the smaller internal volume among the (few) models on the market.