Cargando…

Hemoptysis After CardioMEMS Implantation: Case Report and Review

Patient: Female, 79 Final Diagnosis: Hemoptysis Symptoms: Hemoptysis Medication: — Clinical Procedure: — Specialty: Cardiology OBJECTIVE: Unusual or unexpected effect of treatment BACKGROUND: The CardioMEMS heart failure system is a small sensor that is placed in a branch pulmonary artery for ambula...

Descripción completa

Detalles Bibliográficos
Autores principales: Rali, Aniket S., Shah, Zubair, Sauer, Andrew J., Gupta, Kamal
Formato: Online Artículo Texto
Lenguaje:English
Publicado: International Scientific Literature, Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5890615/
https://www.ncbi.nlm.nih.gov/pubmed/29602943
http://dx.doi.org/10.12659/AJCR.908508
_version_ 1783312896721158144
author Rali, Aniket S.
Shah, Zubair
Sauer, Andrew J.
Gupta, Kamal
author_facet Rali, Aniket S.
Shah, Zubair
Sauer, Andrew J.
Gupta, Kamal
author_sort Rali, Aniket S.
collection PubMed
description Patient: Female, 79 Final Diagnosis: Hemoptysis Symptoms: Hemoptysis Medication: — Clinical Procedure: — Specialty: Cardiology OBJECTIVE: Unusual or unexpected effect of treatment BACKGROUND: The CardioMEMS heart failure system is a small sensor that is placed in a branch pulmonary artery for ambulatory monitoring of pulmonary artery pressures. CardioMEMS has been approved for use in the United States in patients with New York Heart Association (NYHA) class III heart failure and frequent hospitalizations. In this report we describe a patient who had hemoptysis after CardioMEMS implantation. Further, we discuss possible etiologies for the occurrence of hemoptysis and suggest strategies to minimize this risk. CASE REPORT: The patient was a 79-year-old female with NYHA class III heart failure with non-ischemic cardiomyopathy (LVEF 40%) and chronic atrial fibrillation who was referred for CardioMEMS implantation. The procedure was completed uneventfully. The patient was transferred out of the procedure suite to the recovery area where she developed a slight cough approximately 20 minutes after the implantation. Within a few coughs the patient started having hemoptysis. She was transferred to the cardiac intensive care unit for observation. She was kept off warfarin and aspirin and her hemoptysis resolved 3 days later. While the exact etiology of hemoptysis in this patient was unclear, we felt that it may have been precipitated by a minor wire-induced distal branch pulmonary artery injury. CONCLUSIONS: Our report discusses hemoptysis as a potential life-threatening complication of CardioMEMS sensor implantation while suggesting possible etiologies and avoidance strategies. As the utilization of this technology expands in the years to come, a more comprehensive national registry for surveillance of device related complications will be crucial.
format Online
Article
Text
id pubmed-5890615
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher International Scientific Literature, Inc.
record_format MEDLINE/PubMed
spelling pubmed-58906152018-04-10 Hemoptysis After CardioMEMS Implantation: Case Report and Review Rali, Aniket S. Shah, Zubair Sauer, Andrew J. Gupta, Kamal Am J Case Rep Articles Patient: Female, 79 Final Diagnosis: Hemoptysis Symptoms: Hemoptysis Medication: — Clinical Procedure: — Specialty: Cardiology OBJECTIVE: Unusual or unexpected effect of treatment BACKGROUND: The CardioMEMS heart failure system is a small sensor that is placed in a branch pulmonary artery for ambulatory monitoring of pulmonary artery pressures. CardioMEMS has been approved for use in the United States in patients with New York Heart Association (NYHA) class III heart failure and frequent hospitalizations. In this report we describe a patient who had hemoptysis after CardioMEMS implantation. Further, we discuss possible etiologies for the occurrence of hemoptysis and suggest strategies to minimize this risk. CASE REPORT: The patient was a 79-year-old female with NYHA class III heart failure with non-ischemic cardiomyopathy (LVEF 40%) and chronic atrial fibrillation who was referred for CardioMEMS implantation. The procedure was completed uneventfully. The patient was transferred out of the procedure suite to the recovery area where she developed a slight cough approximately 20 minutes after the implantation. Within a few coughs the patient started having hemoptysis. She was transferred to the cardiac intensive care unit for observation. She was kept off warfarin and aspirin and her hemoptysis resolved 3 days later. While the exact etiology of hemoptysis in this patient was unclear, we felt that it may have been precipitated by a minor wire-induced distal branch pulmonary artery injury. CONCLUSIONS: Our report discusses hemoptysis as a potential life-threatening complication of CardioMEMS sensor implantation while suggesting possible etiologies and avoidance strategies. As the utilization of this technology expands in the years to come, a more comprehensive national registry for surveillance of device related complications will be crucial. International Scientific Literature, Inc. 2018-03-31 /pmc/articles/PMC5890615/ /pubmed/29602943 http://dx.doi.org/10.12659/AJCR.908508 Text en © Am J Case Rep, 2018 This work is licensed under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0 (https://creativecommons.org/licenses/by-nc-nd/4.0/) )
spellingShingle Articles
Rali, Aniket S.
Shah, Zubair
Sauer, Andrew J.
Gupta, Kamal
Hemoptysis After CardioMEMS Implantation: Case Report and Review
title Hemoptysis After CardioMEMS Implantation: Case Report and Review
title_full Hemoptysis After CardioMEMS Implantation: Case Report and Review
title_fullStr Hemoptysis After CardioMEMS Implantation: Case Report and Review
title_full_unstemmed Hemoptysis After CardioMEMS Implantation: Case Report and Review
title_short Hemoptysis After CardioMEMS Implantation: Case Report and Review
title_sort hemoptysis after cardiomems implantation: case report and review
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5890615/
https://www.ncbi.nlm.nih.gov/pubmed/29602943
http://dx.doi.org/10.12659/AJCR.908508
work_keys_str_mv AT ralianikets hemoptysisaftercardiomemsimplantationcasereportandreview
AT shahzubair hemoptysisaftercardiomemsimplantationcasereportandreview
AT sauerandrewj hemoptysisaftercardiomemsimplantationcasereportandreview
AT guptakamal hemoptysisaftercardiomemsimplantationcasereportandreview