Single-Breath-Hold Whole-heart Unenhanced Coronary MRA Using Multi-shot Gradient Echo EPI at 3T: Comparison with Free-breathing Turbo-field-echo Coronary MRA on Healthy Volunteers

PURPOSE: We investigated the feasibility of single breath hold unenhanced coronary MRA using multi-shot gradient echo planar imaging (MSG-EPI) on a 3T-scanner. METHODS: Fourteen volunteers underwent single breath hold coronary MRA with a MSG-EPI and free-breathing turbo field echo (TFE) coronary MRA...

Descripción completa

Detalles Bibliográficos
Autores principales: Iyama, Yuji, Nakaura, Takeshi, Nagayama, Yasunori, Oda, Seitaro, Utsunomiya, Daisuke, Kidoh, Masafumi, Yuki, Hideaki, Hirata, Kenichiro, Namimoto, Tomohiro, Kitajima, Mika, Morita, Kosuke, Funama, Yoshinori, Takemura, Atsushi, Okuaki, Tomoyuki, Yamashita, Yasuyuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Japanese Society for Magnetic Resonance in Medicine 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5891342/
https://www.ncbi.nlm.nih.gov/pubmed/28652524
http://dx.doi.org/10.2463/mrms.mp.2017-0037
Descripción
Sumario:PURPOSE: We investigated the feasibility of single breath hold unenhanced coronary MRA using multi-shot gradient echo planar imaging (MSG-EPI) on a 3T-scanner. METHODS: Fourteen volunteers underwent single breath hold coronary MRA with a MSG-EPI and free-breathing turbo field echo (TFE) coronary MRA at 3T. The acquisition time, signal to noise ratio (SNR), and the contrast of the sequences were compared with the paired t-test. Readers evaluated the image contrast, noise, sharpness, artifacts, and the overall image quality. RESULTS: The acquisition time was 88.1% shorter for MSG-EPI than TFE (24.7 ± 2.5 vs 206.4 ± 23.1 sec, P < 0.01). The SNR was significantly higher on MSG-EPI than TFE scans (P < 0.01). There was no significant difference in the contrast on MSG-EPI and TFE scans (1.8 ± 0.3 vs 1.9 ± 0.3, P = 0.24). There was no significant difference in image contrast, image sharpness, and overall image quality between two scan techniques. The score of image noise and artifact were significantly higher on MSG-EPI than TFE scans (P < 0.05). CONCLUSION: The single breath hold MSG-EPI sequence is a promising technique for shortening the scan time and for preserving the image quality of unenhanced whole heart coronary MRA on a 3T scanner.