High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer

A 46‐year‐old female with a body mass index of 38.9 kg/m(2) and no organic heart disease underwent a subcutaneous implantable cardioverter‐defibrillator implantation for secondary prevention of sudden cardiac death in the setting of idiopathic ventricular fibrillation. Defibrillation threshold (DFT)...

Descripción completa

Detalles Bibliográficos
Autores principales: Hirao, Tatsuhiko, Nitta, Junichi, Sato, Akira, Takahashi, Yoshihide, Goya, Masahiko, Hirao, Kenzo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5891406/
https://www.ncbi.nlm.nih.gov/pubmed/29657596
http://dx.doi.org/10.1002/joa3.12033
_version_ 1783313001506406400
author Hirao, Tatsuhiko
Nitta, Junichi
Sato, Akira
Takahashi, Yoshihide
Goya, Masahiko
Hirao, Kenzo
author_facet Hirao, Tatsuhiko
Nitta, Junichi
Sato, Akira
Takahashi, Yoshihide
Goya, Masahiko
Hirao, Kenzo
author_sort Hirao, Tatsuhiko
collection PubMed
description A 46‐year‐old female with a body mass index of 38.9 kg/m(2) and no organic heart disease underwent a subcutaneous implantable cardioverter‐defibrillator implantation for secondary prevention of sudden cardiac death in the setting of idiopathic ventricular fibrillation. Defibrillation threshold (DFT) testing during implantation detected high shock impedance and high DFT. Fluoroscopy revealed subcoil fat between the lead and the sternum, which we suspected was the reason for the high shock impedance and high DFT. We repositioned the lead to a site just above the sternum and the shock impedance and DFT improved to within the respective normal ranges.
format Online
Article
Text
id pubmed-5891406
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-58914062018-04-13 High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer Hirao, Tatsuhiko Nitta, Junichi Sato, Akira Takahashi, Yoshihide Goya, Masahiko Hirao, Kenzo J Arrhythm Case Reports A 46‐year‐old female with a body mass index of 38.9 kg/m(2) and no organic heart disease underwent a subcutaneous implantable cardioverter‐defibrillator implantation for secondary prevention of sudden cardiac death in the setting of idiopathic ventricular fibrillation. Defibrillation threshold (DFT) testing during implantation detected high shock impedance and high DFT. Fluoroscopy revealed subcoil fat between the lead and the sternum, which we suspected was the reason for the high shock impedance and high DFT. We repositioned the lead to a site just above the sternum and the shock impedance and DFT improved to within the respective normal ranges. John Wiley and Sons Inc. 2018-02-08 /pmc/articles/PMC5891406/ /pubmed/29657596 http://dx.doi.org/10.1002/joa3.12033 Text en © 2018 The Authors. Journal of Arrhythmia published by John Wiley & Sons Australia, Ltd on behalf of the Japanese Heart Rhythm Society. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Case Reports
Hirao, Tatsuhiko
Nitta, Junichi
Sato, Akira
Takahashi, Yoshihide
Goya, Masahiko
Hirao, Kenzo
High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
title High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
title_full High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
title_fullStr High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
title_full_unstemmed High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
title_short High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
title_sort high defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
topic Case Reports
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5891406/
https://www.ncbi.nlm.nih.gov/pubmed/29657596
http://dx.doi.org/10.1002/joa3.12033
work_keys_str_mv AT hiraotatsuhiko highdefibrillationthresholdwithasubcutaneousimplantablecardiacdefibrillatorduetotheleadhavingbeenpositionedinthefatlayer
AT nittajunichi highdefibrillationthresholdwithasubcutaneousimplantablecardiacdefibrillatorduetotheleadhavingbeenpositionedinthefatlayer
AT satoakira highdefibrillationthresholdwithasubcutaneousimplantablecardiacdefibrillatorduetotheleadhavingbeenpositionedinthefatlayer
AT takahashiyoshihide highdefibrillationthresholdwithasubcutaneousimplantablecardiacdefibrillatorduetotheleadhavingbeenpositionedinthefatlayer
AT goyamasahiko highdefibrillationthresholdwithasubcutaneousimplantablecardiacdefibrillatorduetotheleadhavingbeenpositionedinthefatlayer
AT hiraokenzo highdefibrillationthresholdwithasubcutaneousimplantablecardiacdefibrillatorduetotheleadhavingbeenpositionedinthefatlayer