Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System

BACKGROUND: Cardiac magnetic resonance imaging (cMRI) has become the non-invasive reference standard for the evaluation of cardiac function and viability. The introduction of open, high-field, 1.0T (HFO) MR scanners offers advantages for examinations of obese, claustrophobic and paediatric patients....

Descripción completa

Detalles Bibliográficos
Autores principales: Fischbach, Katharina, Kosiek, Otrud, Friebe, Björn, Wybranski, Christian, Schnackenburg, Bernhard, Schmeisser, Alexander, Smid, Jan, Ricke, Jens, Pech, Maciej
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5894020/
https://www.ncbi.nlm.nih.gov/pubmed/29662579
http://dx.doi.org/10.12659/PJR.899822
_version_ 1783313417769058304
author Fischbach, Katharina
Kosiek, Otrud
Friebe, Björn
Wybranski, Christian
Schnackenburg, Bernhard
Schmeisser, Alexander
Smid, Jan
Ricke, Jens
Pech, Maciej
author_facet Fischbach, Katharina
Kosiek, Otrud
Friebe, Björn
Wybranski, Christian
Schnackenburg, Bernhard
Schmeisser, Alexander
Smid, Jan
Ricke, Jens
Pech, Maciej
author_sort Fischbach, Katharina
collection PubMed
description BACKGROUND: Cardiac magnetic resonance imaging (cMRI) has become the non-invasive reference standard for the evaluation of cardiac function and viability. The introduction of open, high-field, 1.0T (HFO) MR scanners offers advantages for examinations of obese, claustrophobic and paediatric patients. The aim of our study was to compare standard cMRI sequences from an HFO scanner and those from a cylindrical, 1.5T MR system. MATERIAL/METHOD: Fifteen volunteers underwent cMRI both in an open HFO and in a cylindrical MR system. The protocol consisted of cine and unenhanced tissue sequences. The signal-to-noise ratio (SNR) for each sequence and blood-myocardium contrast for the cine sequences were assessed. Image quality and artefacts were rated. The location and number of non-diagnostic segments was determined. Volunteers’ tolerance to examinations in both scanners was investigated. RESULTS: SNR was significantly lower in the HFO scanner (all p<0.001). However, the contrast of the cine sequence was significantly higher in the HFO platform compared to the 1.5T MR scanner (0.685±0.41 vs. 0.611±0.54; p<0.001). Image quality was comparable for all sequences (all p>0.05). Overall, only few non-diagnostic myocardial segments were recorded: 6/960 (0.6%) by the HFO and 17/960 (1.8%) segments by the cylindrical system. The volunteers expressed a preference for the open MR system (p<0.01). CONCLUSIONS: Standard cardiac MRI sequences in an HFO platform offer a high image quality that is comparable to the quality of images acquired in a cylindrical 1.5T MR scanner. An open scanner design may potentially improve tolerance of cardiac MRI and therefore allow to examine an even broader patient spectrum.
format Online
Article
Text
id pubmed-5894020
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-58940202018-04-16 Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System Fischbach, Katharina Kosiek, Otrud Friebe, Björn Wybranski, Christian Schnackenburg, Bernhard Schmeisser, Alexander Smid, Jan Ricke, Jens Pech, Maciej Pol J Radiol Original Article BACKGROUND: Cardiac magnetic resonance imaging (cMRI) has become the non-invasive reference standard for the evaluation of cardiac function and viability. The introduction of open, high-field, 1.0T (HFO) MR scanners offers advantages for examinations of obese, claustrophobic and paediatric patients. The aim of our study was to compare standard cMRI sequences from an HFO scanner and those from a cylindrical, 1.5T MR system. MATERIAL/METHOD: Fifteen volunteers underwent cMRI both in an open HFO and in a cylindrical MR system. The protocol consisted of cine and unenhanced tissue sequences. The signal-to-noise ratio (SNR) for each sequence and blood-myocardium contrast for the cine sequences were assessed. Image quality and artefacts were rated. The location and number of non-diagnostic segments was determined. Volunteers’ tolerance to examinations in both scanners was investigated. RESULTS: SNR was significantly lower in the HFO scanner (all p<0.001). However, the contrast of the cine sequence was significantly higher in the HFO platform compared to the 1.5T MR scanner (0.685±0.41 vs. 0.611±0.54; p<0.001). Image quality was comparable for all sequences (all p>0.05). Overall, only few non-diagnostic myocardial segments were recorded: 6/960 (0.6%) by the HFO and 17/960 (1.8%) segments by the cylindrical system. The volunteers expressed a preference for the open MR system (p<0.01). CONCLUSIONS: Standard cardiac MRI sequences in an HFO platform offer a high image quality that is comparable to the quality of images acquired in a cylindrical 1.5T MR scanner. An open scanner design may potentially improve tolerance of cardiac MRI and therefore allow to examine an even broader patient spectrum. Termedia Publishing House 2017-09-15 /pmc/articles/PMC5894020/ /pubmed/29662579 http://dx.doi.org/10.12659/PJR.899822 Text en Copyright © Polish Medical Society of Radiology 2017 https://creativecommons.org/licenses/by-nc-nd/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-Noncommercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0). License allowing third parties to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.
spellingShingle Original Article
Fischbach, Katharina
Kosiek, Otrud
Friebe, Björn
Wybranski, Christian
Schnackenburg, Bernhard
Schmeisser, Alexander
Smid, Jan
Ricke, Jens
Pech, Maciej
Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System
title Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System
title_full Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System
title_fullStr Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System
title_full_unstemmed Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System
title_short Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System
title_sort cardiac magnetic resonance imaging using an open 1.0t mr platform: a comparative study with a 1.5t tunnel system
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5894020/
https://www.ncbi.nlm.nih.gov/pubmed/29662579
http://dx.doi.org/10.12659/PJR.899822
work_keys_str_mv AT fischbachkatharina cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem
AT kosiekotrud cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem
AT friebebjorn cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem
AT wybranskichristian cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem
AT schnackenburgbernhard cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem
AT schmeisseralexander cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem
AT smidjan cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem
AT rickejens cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem
AT pechmaciej cardiacmagneticresonanceimagingusinganopen10tmrplatformacomparativestudywitha15ttunnelsystem