Maximizing the fixation strength of modular components by impaction without tissue damage
OBJECTIVES: Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of thi...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5895944/ https://www.ncbi.nlm.nih.gov/pubmed/29682286 http://dx.doi.org/10.1302/2046-3758.72.BJR-2017-0078.R2 |
_version_ | 1783313746329862144 |
---|---|
author | Krull, A. Morlock, M. M. Bishop, N. E. |
author_facet | Krull, A. Morlock, M. M. Bishop, N. E. |
author_sort | Krull, A. |
collection | PubMed |
description | OBJECTIVES: Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues. METHODS: A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues. RESULTS: The pull-off force of the head increased as the stiffness of the impactor tip increased but without increasing the force transmitted through the springs (patient). Increasing the impaction energy increased the pull-off force but also increased the force transmitted through the springs. CONCLUSIONS: To limit wear and corrosion, manufacturers should maximize the stiffness of the impactor tool but without damaging the surface of the head. This strategy will maximize the stability of the head on the stem for a given applied energy, without influencing the force transmitted through the patient’s tissues. Current impactor designs already appear to approach this limit. Increasing the applied energy (which is dependent on the mass of the hammer and square of the contact speed) increases the stability of the modular connection but proportionally increases the force transmitted through the patient’s tissues, as well as to the surface of the head, and should be restricted to safe levels. Cite this article: A. Krull, M. M. Morlock, N. E. Bishop. Maximizing the fixation strength of modular components by impaction without tissue damage. Bone Joint Res 2018;7:196–204. DOI: 10.1302/2046-3758.72.BJR-2017-0078.R2. |
format | Online Article Text |
id | pubmed-5895944 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
record_format | MEDLINE/PubMed |
spelling | pubmed-58959442018-04-20 Maximizing the fixation strength of modular components by impaction without tissue damage Krull, A. Morlock, M. M. Bishop, N. E. Bone Joint Res Research OBJECTIVES: Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues. METHODS: A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues. RESULTS: The pull-off force of the head increased as the stiffness of the impactor tip increased but without increasing the force transmitted through the springs (patient). Increasing the impaction energy increased the pull-off force but also increased the force transmitted through the springs. CONCLUSIONS: To limit wear and corrosion, manufacturers should maximize the stiffness of the impactor tool but without damaging the surface of the head. This strategy will maximize the stability of the head on the stem for a given applied energy, without influencing the force transmitted through the patient’s tissues. Current impactor designs already appear to approach this limit. Increasing the applied energy (which is dependent on the mass of the hammer and square of the contact speed) increases the stability of the modular connection but proportionally increases the force transmitted through the patient’s tissues, as well as to the surface of the head, and should be restricted to safe levels. Cite this article: A. Krull, M. M. Morlock, N. E. Bishop. Maximizing the fixation strength of modular components by impaction without tissue damage. Bone Joint Res 2018;7:196–204. DOI: 10.1302/2046-3758.72.BJR-2017-0078.R2. 2018-04-12 /pmc/articles/PMC5895944/ /pubmed/29682286 http://dx.doi.org/10.1302/2046-3758.72.BJR-2017-0078.R2 Text en © 2018 Krull et al. This is an open-access article distributed under the terms of the Creative Commons Attributions licence (CC-BY-NC), which permits unrestricted use, distribution, and reproduction in any medium, but not for commercial gain, provided the original author and source are credited. |
spellingShingle | Research Krull, A. Morlock, M. M. Bishop, N. E. Maximizing the fixation strength of modular components by impaction without tissue damage |
title | Maximizing the fixation strength of modular components by impaction without tissue damage |
title_full | Maximizing the fixation strength of modular components by impaction without tissue damage |
title_fullStr | Maximizing the fixation strength of modular components by impaction without tissue damage |
title_full_unstemmed | Maximizing the fixation strength of modular components by impaction without tissue damage |
title_short | Maximizing the fixation strength of modular components by impaction without tissue damage |
title_sort | maximizing the fixation strength of modular components by impaction without tissue damage |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5895944/ https://www.ncbi.nlm.nih.gov/pubmed/29682286 http://dx.doi.org/10.1302/2046-3758.72.BJR-2017-0078.R2 |
work_keys_str_mv | AT krulla maximizingthefixationstrengthofmodularcomponentsbyimpactionwithouttissuedamage AT morlockmm maximizingthefixationstrengthofmodularcomponentsbyimpactionwithouttissuedamage AT bishopne maximizingthefixationstrengthofmodularcomponentsbyimpactionwithouttissuedamage |