Cargando…
Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study
PURPOSE: The purpose of this study is to compare the effect of implant-abutment connections, abutment angulations, and screw lengths on screw loosening (SL) of preloaded abutment using three dimensional (3D) finite element analysis. MATERIALS AND METHODS: 3D models of implants (conical connection wi...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5903180/ https://www.ncbi.nlm.nih.gov/pubmed/29692570 http://dx.doi.org/10.4103/jips.jips_219_17 |
_version_ | 1783314903966154752 |
---|---|
author | Kanneganti, Krishna Chaitanya Vinnakota, Dileep Nag Pottem, Srinivas Rao Pulagam, Mahesh |
author_facet | Kanneganti, Krishna Chaitanya Vinnakota, Dileep Nag Pottem, Srinivas Rao Pulagam, Mahesh |
author_sort | Kanneganti, Krishna Chaitanya |
collection | PubMed |
description | PURPOSE: The purpose of this study is to compare the effect of implant-abutment connections, abutment angulations, and screw lengths on screw loosening (SL) of preloaded abutment using three dimensional (3D) finite element analysis. MATERIALS AND METHODS: 3D models of implants (conical connection with hex/trilobed connections), abutments (straight/angulated), abutment screws (short/long), and crown and bone were designed using software Parametric Technology Corporation Creo and assembled to form 8 simulations. After discretization, the contact stresses developed for 150 N vertical and 100 N oblique load applications were analyzed, using ABAQUS. By assessing damage initiation and shortest fatigue load on screw threads, the SL for 2.5, 5, and 10 lakh cyclic loads were estimated, using fe-safe program. The obtained values were compared for influence of connection design, abutment angulation, and screw length. RESULTS: In straight abutment models, conical connection showed more damage (14.3%–72.3%) when compared to trilobe (10.1%–65.73%) at 2.5, 5, and 10 lakh cycles for both vertical and oblique loads, whereas in angulated abutments, trilobe (16.1%–76.9%) demonstrated more damage compared to conical (13.5%–70%). Irrespective of the connection type and abutment angulation, short screws showed more percentage of damage compared to long screws. CONCLUSIONS: The present study suggests selecting appropriate implant-abutment connection based on the abutment angulation, as well as preferring long screws with more number of threads for effective preload retention by the screws. |
format | Online Article Text |
id | pubmed-5903180 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-59031802019-04-01 Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study Kanneganti, Krishna Chaitanya Vinnakota, Dileep Nag Pottem, Srinivas Rao Pulagam, Mahesh J Indian Prosthodont Soc Original Article PURPOSE: The purpose of this study is to compare the effect of implant-abutment connections, abutment angulations, and screw lengths on screw loosening (SL) of preloaded abutment using three dimensional (3D) finite element analysis. MATERIALS AND METHODS: 3D models of implants (conical connection with hex/trilobed connections), abutments (straight/angulated), abutment screws (short/long), and crown and bone were designed using software Parametric Technology Corporation Creo and assembled to form 8 simulations. After discretization, the contact stresses developed for 150 N vertical and 100 N oblique load applications were analyzed, using ABAQUS. By assessing damage initiation and shortest fatigue load on screw threads, the SL for 2.5, 5, and 10 lakh cyclic loads were estimated, using fe-safe program. The obtained values were compared for influence of connection design, abutment angulation, and screw length. RESULTS: In straight abutment models, conical connection showed more damage (14.3%–72.3%) when compared to trilobe (10.1%–65.73%) at 2.5, 5, and 10 lakh cycles for both vertical and oblique loads, whereas in angulated abutments, trilobe (16.1%–76.9%) demonstrated more damage compared to conical (13.5%–70%). Irrespective of the connection type and abutment angulation, short screws showed more percentage of damage compared to long screws. CONCLUSIONS: The present study suggests selecting appropriate implant-abutment connection based on the abutment angulation, as well as preferring long screws with more number of threads for effective preload retention by the screws. Medknow Publications & Media Pvt Ltd 2018 /pmc/articles/PMC5903180/ /pubmed/29692570 http://dx.doi.org/10.4103/jips.jips_219_17 Text en Copyright: © 2018 The Journal of Indian Prosthodontic Society http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Kanneganti, Krishna Chaitanya Vinnakota, Dileep Nag Pottem, Srinivas Rao Pulagam, Mahesh Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study |
title | Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study |
title_full | Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study |
title_fullStr | Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study |
title_full_unstemmed | Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study |
title_short | Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study |
title_sort | comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: an in vitro study |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5903180/ https://www.ncbi.nlm.nih.gov/pubmed/29692570 http://dx.doi.org/10.4103/jips.jips_219_17 |
work_keys_str_mv | AT kannegantikrishnachaitanya comparativeeffectofimplantabutmentconnectionsabutmentangulationsandscrewlengthsonpreloadedabutmentscrewusingthreedimensionalfiniteelementanalysisaninvitrostudy AT vinnakotadileepnag comparativeeffectofimplantabutmentconnectionsabutmentangulationsandscrewlengthsonpreloadedabutmentscrewusingthreedimensionalfiniteelementanalysisaninvitrostudy AT pottemsrinivasrao comparativeeffectofimplantabutmentconnectionsabutmentangulationsandscrewlengthsonpreloadedabutmentscrewusingthreedimensionalfiniteelementanalysisaninvitrostudy AT pulagammahesh comparativeeffectofimplantabutmentconnectionsabutmentangulationsandscrewlengthsonpreloadedabutmentscrewusingthreedimensionalfiniteelementanalysisaninvitrostudy |