Repeatability of Retinal Sensitivity Measurements Using a Medmont Dark-Adapted Chromatic Perimeter in Healthy and Age-Related Macular Degeneration Cases
PURPOSE: To determine the intrasession and intersession test–retest repeatability of retinal sensitivity measurements using a dark-adapted chromatic perimeter (DACP). METHODS: For intrasession testing, retinal sensitivity within the central 24° for the 505-nm stimulus was measured after 20, 30, and...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Association for Research in Vision and Ophthalmology
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5931259/ https://www.ncbi.nlm.nih.gov/pubmed/29736324 http://dx.doi.org/10.1167/tvst.7.3.3 |
_version_ | 1783319617156939776 |
---|---|
author | Tan, Rose S. Guymer, Robyn H. Luu, Chi D. |
author_facet | Tan, Rose S. Guymer, Robyn H. Luu, Chi D. |
author_sort | Tan, Rose S. |
collection | PubMed |
description | PURPOSE: To determine the intrasession and intersession test–retest repeatability of retinal sensitivity measurements using a dark-adapted chromatic perimeter (DACP). METHODS: For intrasession testing, retinal sensitivity within the central 24° for the 505-nm stimulus was measured after 20, 30, and 40 minutes of dark adaptation (DA) and for the 625-nm stimulus was measured after the first and second 505-nm tests. For intersession testing, retinal sensitivity for both stimuli was measured after 30 minutes of DA at baseline and 1 month. The point-wise sensitivity (PWS) difference and coefficient of repeatability (CoR) of each stimulus and group were determined. RESULTS: For intrasession testing, 10 age-related macular degeneration (AMD) and eight control subjects were recruited. The overall CoR for the 505-nm stimulus was 8.4 dB for control subjects and 9.1 dB for AMD cases, and for the 625-nm stimulus was 6.7 dB for control subjects and 9.5 dB for AMD cases. For intersession testing, seven AMD cases and 13 control subjects returned an overall CoR for the 505-nm stimulus of 8.2 dB for the control and 11.7 dB for the AMD group. For the 625-nm stimulus the CoR was 6.2 dB for the control group and 8.4 dB for the AMD group. Approximately 80% of all test points had a PWS difference of ±5 dB between the two intrasession or intersession measurements for both stimuli. CONCLUSIONS: The CoR for the DACP is larger than that reported for scotopic perimeters; however, the majority of test points had a PWS difference of ±5 dB between tests. TRANSLATIONAL RELEVANCE: The DACP offers an opportunity to measure static and dynamic rod function at multiple locations with an acceptable reproducibility level. |
format | Online Article Text |
id | pubmed-5931259 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | The Association for Research in Vision and Ophthalmology |
record_format | MEDLINE/PubMed |
spelling | pubmed-59312592018-05-07 Repeatability of Retinal Sensitivity Measurements Using a Medmont Dark-Adapted Chromatic Perimeter in Healthy and Age-Related Macular Degeneration Cases Tan, Rose S. Guymer, Robyn H. Luu, Chi D. Transl Vis Sci Technol Articles PURPOSE: To determine the intrasession and intersession test–retest repeatability of retinal sensitivity measurements using a dark-adapted chromatic perimeter (DACP). METHODS: For intrasession testing, retinal sensitivity within the central 24° for the 505-nm stimulus was measured after 20, 30, and 40 minutes of dark adaptation (DA) and for the 625-nm stimulus was measured after the first and second 505-nm tests. For intersession testing, retinal sensitivity for both stimuli was measured after 30 minutes of DA at baseline and 1 month. The point-wise sensitivity (PWS) difference and coefficient of repeatability (CoR) of each stimulus and group were determined. RESULTS: For intrasession testing, 10 age-related macular degeneration (AMD) and eight control subjects were recruited. The overall CoR for the 505-nm stimulus was 8.4 dB for control subjects and 9.1 dB for AMD cases, and for the 625-nm stimulus was 6.7 dB for control subjects and 9.5 dB for AMD cases. For intersession testing, seven AMD cases and 13 control subjects returned an overall CoR for the 505-nm stimulus of 8.2 dB for the control and 11.7 dB for the AMD group. For the 625-nm stimulus the CoR was 6.2 dB for the control group and 8.4 dB for the AMD group. Approximately 80% of all test points had a PWS difference of ±5 dB between the two intrasession or intersession measurements for both stimuli. CONCLUSIONS: The CoR for the DACP is larger than that reported for scotopic perimeters; however, the majority of test points had a PWS difference of ±5 dB between tests. TRANSLATIONAL RELEVANCE: The DACP offers an opportunity to measure static and dynamic rod function at multiple locations with an acceptable reproducibility level. The Association for Research in Vision and Ophthalmology 2018-05-01 /pmc/articles/PMC5931259/ /pubmed/29736324 http://dx.doi.org/10.1167/tvst.7.3.3 Text en Copyright 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. |
spellingShingle | Articles Tan, Rose S. Guymer, Robyn H. Luu, Chi D. Repeatability of Retinal Sensitivity Measurements Using a Medmont Dark-Adapted Chromatic Perimeter in Healthy and Age-Related Macular Degeneration Cases |
title | Repeatability of Retinal Sensitivity Measurements Using a Medmont Dark-Adapted Chromatic Perimeter in Healthy and Age-Related Macular Degeneration Cases |
title_full | Repeatability of Retinal Sensitivity Measurements Using a Medmont Dark-Adapted Chromatic Perimeter in Healthy and Age-Related Macular Degeneration Cases |
title_fullStr | Repeatability of Retinal Sensitivity Measurements Using a Medmont Dark-Adapted Chromatic Perimeter in Healthy and Age-Related Macular Degeneration Cases |
title_full_unstemmed | Repeatability of Retinal Sensitivity Measurements Using a Medmont Dark-Adapted Chromatic Perimeter in Healthy and Age-Related Macular Degeneration Cases |
title_short | Repeatability of Retinal Sensitivity Measurements Using a Medmont Dark-Adapted Chromatic Perimeter in Healthy and Age-Related Macular Degeneration Cases |
title_sort | repeatability of retinal sensitivity measurements using a medmont dark-adapted chromatic perimeter in healthy and age-related macular degeneration cases |
topic | Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5931259/ https://www.ncbi.nlm.nih.gov/pubmed/29736324 http://dx.doi.org/10.1167/tvst.7.3.3 |
work_keys_str_mv | AT tanroses repeatabilityofretinalsensitivitymeasurementsusingamedmontdarkadaptedchromaticperimeterinhealthyandagerelatedmaculardegenerationcases AT guymerrobynh repeatabilityofretinalsensitivitymeasurementsusingamedmontdarkadaptedchromaticperimeterinhealthyandagerelatedmaculardegenerationcases AT luuchid repeatabilityofretinalsensitivitymeasurementsusingamedmontdarkadaptedchromaticperimeterinhealthyandagerelatedmaculardegenerationcases |