A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices

BACKGROUND: Clinical guidelines recommend monitoring of blood pressure at home using an automatic blood pressure device for the management of hypertension. Devices are not often calibrated against direct blood pressure measures, leaving health care providers and patients with less reliable informati...

Descripción completa

Detalles Bibliográficos
Autores principales: Melville, Sarah, Teskey, Robert, Philip, Shona, Simpson, Jeremy A, Lutchmedial, Sohrab, Brunt, Keith R
Formato: Online Artículo Texto
Lenguaje:English
Publicado: JMIR Publications 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5943631/
https://www.ncbi.nlm.nih.gov/pubmed/29695375
http://dx.doi.org/10.2196/jmir.8009
_version_ 1783321667610607616
author Melville, Sarah
Teskey, Robert
Philip, Shona
Simpson, Jeremy A
Lutchmedial, Sohrab
Brunt, Keith R
author_facet Melville, Sarah
Teskey, Robert
Philip, Shona
Simpson, Jeremy A
Lutchmedial, Sohrab
Brunt, Keith R
author_sort Melville, Sarah
collection PubMed
description BACKGROUND: Clinical guidelines recommend monitoring of blood pressure at home using an automatic blood pressure device for the management of hypertension. Devices are not often calibrated against direct blood pressure measures, leaving health care providers and patients with less reliable information than is possible with current technology. Rigorous assessments of medical devices are necessary for establishing clinical utility. OBJECTIVE: The purpose of our study was 2-fold: (1) to assess the validity and perform iterative calibration of indirect blood pressure measurements by a noninvasive wrist cuff blood pressure device in direct comparison with simultaneously recorded peripheral and central intra-arterial blood pressure measurements and (2) to assess the validity of the measurements thereafter of the noninvasive wrist cuff blood pressure device in comparison with measurements by a noninvasive upper arm blood pressure device to the Canadian hypertension guidelines. METHODS: The cloud-based blood pressure algorithms for an oscillometric wrist cuff device were iteratively calibrated to direct pressure measures in 20 consented patient participants. We then assessed measurement validity of the device, using Bland-Altman analysis during routine cardiovascular catheterization. RESULTS: The precalibrated absolute mean difference between direct intra-arterial to wrist cuff pressure measurements were 10.8 (SD 9.7) for systolic and 16.1 (SD 6.3) for diastolic. The postcalibrated absolute mean difference was 7.2 (SD 5.1) for systolic and 4.3 (SD 3.3) for diastolic pressures. This is an improvement in accuracy of 33% systolic and 73% diastolic with a 48% reduction in the variability for both measures. Furthermore, the wrist cuff device demonstrated similar sensitivity in measuring high blood pressure compared with the direct intra-arterial method. The device, when calibrated to direct aortic pressures, demonstrated the potential to reduce a treatment gap in high blood pressure measurements. CONCLUSIONS: The systolic pressure measurements of the wrist cuff have been iteratively calibrated using gold standard central (ascending aortic) pressure. This improves the accuracy of the indirect measures and potentially reduces the treatment gap. Devices that undergo auscultatory (indirect) calibration for licensing can be greatly improved by additional iterative calibration via intra-arterial (direct) measures of blood pressure. Further clinical trials with repeated use of the device over time are needed to assess the reliability of the device in accordance with current and evolving guidelines for informed decision making in the management of hypertension. TRIAL REGISTRATION: ClinicalTrials.gov NCT03015363; https://clinicaltrials.gov/ct2/show/NCT03015363 (Archived by WebCite at http://www.webcitation.org/6xPZgseYS)
format Online
Article
Text
id pubmed-5943631
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher JMIR Publications
record_format MEDLINE/PubMed
spelling pubmed-59436312018-05-17 A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices Melville, Sarah Teskey, Robert Philip, Shona Simpson, Jeremy A Lutchmedial, Sohrab Brunt, Keith R J Med Internet Res Original Paper BACKGROUND: Clinical guidelines recommend monitoring of blood pressure at home using an automatic blood pressure device for the management of hypertension. Devices are not often calibrated against direct blood pressure measures, leaving health care providers and patients with less reliable information than is possible with current technology. Rigorous assessments of medical devices are necessary for establishing clinical utility. OBJECTIVE: The purpose of our study was 2-fold: (1) to assess the validity and perform iterative calibration of indirect blood pressure measurements by a noninvasive wrist cuff blood pressure device in direct comparison with simultaneously recorded peripheral and central intra-arterial blood pressure measurements and (2) to assess the validity of the measurements thereafter of the noninvasive wrist cuff blood pressure device in comparison with measurements by a noninvasive upper arm blood pressure device to the Canadian hypertension guidelines. METHODS: The cloud-based blood pressure algorithms for an oscillometric wrist cuff device were iteratively calibrated to direct pressure measures in 20 consented patient participants. We then assessed measurement validity of the device, using Bland-Altman analysis during routine cardiovascular catheterization. RESULTS: The precalibrated absolute mean difference between direct intra-arterial to wrist cuff pressure measurements were 10.8 (SD 9.7) for systolic and 16.1 (SD 6.3) for diastolic. The postcalibrated absolute mean difference was 7.2 (SD 5.1) for systolic and 4.3 (SD 3.3) for diastolic pressures. This is an improvement in accuracy of 33% systolic and 73% diastolic with a 48% reduction in the variability for both measures. Furthermore, the wrist cuff device demonstrated similar sensitivity in measuring high blood pressure compared with the direct intra-arterial method. The device, when calibrated to direct aortic pressures, demonstrated the potential to reduce a treatment gap in high blood pressure measurements. CONCLUSIONS: The systolic pressure measurements of the wrist cuff have been iteratively calibrated using gold standard central (ascending aortic) pressure. This improves the accuracy of the indirect measures and potentially reduces the treatment gap. Devices that undergo auscultatory (indirect) calibration for licensing can be greatly improved by additional iterative calibration via intra-arterial (direct) measures of blood pressure. Further clinical trials with repeated use of the device over time are needed to assess the reliability of the device in accordance with current and evolving guidelines for informed decision making in the management of hypertension. TRIAL REGISTRATION: ClinicalTrials.gov NCT03015363; https://clinicaltrials.gov/ct2/show/NCT03015363 (Archived by WebCite at http://www.webcitation.org/6xPZgseYS) JMIR Publications 2018-04-25 /pmc/articles/PMC5943631/ /pubmed/29695375 http://dx.doi.org/10.2196/jmir.8009 Text en ©Sarah Melville, Robert Teskey, Shona Philip, Jeremy A Simpson, Sohrab Lutchmedial, Keith R Brunt. Originally published in the Journal of Medical Internet Research (http://www.jmir.org), 25.04.2018. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research, is properly cited. The complete bibliographic information, a link to the original publication on http://www.jmir.org/, as well as this copyright and license information must be included.
spellingShingle Original Paper
Melville, Sarah
Teskey, Robert
Philip, Shona
Simpson, Jeremy A
Lutchmedial, Sohrab
Brunt, Keith R
A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices
title A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices
title_full A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices
title_fullStr A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices
title_full_unstemmed A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices
title_short A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices
title_sort comparison and calibration of a wrist-worn blood pressure monitor for patient management: assessing the reliability of innovative blood pressure devices
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5943631/
https://www.ncbi.nlm.nih.gov/pubmed/29695375
http://dx.doi.org/10.2196/jmir.8009
work_keys_str_mv AT melvillesarah acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT teskeyrobert acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT philipshona acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT simpsonjeremya acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT lutchmedialsohrab acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT bruntkeithr acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT melvillesarah comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT teskeyrobert comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT philipshona comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT simpsonjeremya comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT lutchmedialsohrab comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices
AT bruntkeithr comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices