A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices
BACKGROUND: Clinical guidelines recommend monitoring of blood pressure at home using an automatic blood pressure device for the management of hypertension. Devices are not often calibrated against direct blood pressure measures, leaving health care providers and patients with less reliable informati...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
JMIR Publications
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5943631/ https://www.ncbi.nlm.nih.gov/pubmed/29695375 http://dx.doi.org/10.2196/jmir.8009 |
_version_ | 1783321667610607616 |
---|---|
author | Melville, Sarah Teskey, Robert Philip, Shona Simpson, Jeremy A Lutchmedial, Sohrab Brunt, Keith R |
author_facet | Melville, Sarah Teskey, Robert Philip, Shona Simpson, Jeremy A Lutchmedial, Sohrab Brunt, Keith R |
author_sort | Melville, Sarah |
collection | PubMed |
description | BACKGROUND: Clinical guidelines recommend monitoring of blood pressure at home using an automatic blood pressure device for the management of hypertension. Devices are not often calibrated against direct blood pressure measures, leaving health care providers and patients with less reliable information than is possible with current technology. Rigorous assessments of medical devices are necessary for establishing clinical utility. OBJECTIVE: The purpose of our study was 2-fold: (1) to assess the validity and perform iterative calibration of indirect blood pressure measurements by a noninvasive wrist cuff blood pressure device in direct comparison with simultaneously recorded peripheral and central intra-arterial blood pressure measurements and (2) to assess the validity of the measurements thereafter of the noninvasive wrist cuff blood pressure device in comparison with measurements by a noninvasive upper arm blood pressure device to the Canadian hypertension guidelines. METHODS: The cloud-based blood pressure algorithms for an oscillometric wrist cuff device were iteratively calibrated to direct pressure measures in 20 consented patient participants. We then assessed measurement validity of the device, using Bland-Altman analysis during routine cardiovascular catheterization. RESULTS: The precalibrated absolute mean difference between direct intra-arterial to wrist cuff pressure measurements were 10.8 (SD 9.7) for systolic and 16.1 (SD 6.3) for diastolic. The postcalibrated absolute mean difference was 7.2 (SD 5.1) for systolic and 4.3 (SD 3.3) for diastolic pressures. This is an improvement in accuracy of 33% systolic and 73% diastolic with a 48% reduction in the variability for both measures. Furthermore, the wrist cuff device demonstrated similar sensitivity in measuring high blood pressure compared with the direct intra-arterial method. The device, when calibrated to direct aortic pressures, demonstrated the potential to reduce a treatment gap in high blood pressure measurements. CONCLUSIONS: The systolic pressure measurements of the wrist cuff have been iteratively calibrated using gold standard central (ascending aortic) pressure. This improves the accuracy of the indirect measures and potentially reduces the treatment gap. Devices that undergo auscultatory (indirect) calibration for licensing can be greatly improved by additional iterative calibration via intra-arterial (direct) measures of blood pressure. Further clinical trials with repeated use of the device over time are needed to assess the reliability of the device in accordance with current and evolving guidelines for informed decision making in the management of hypertension. TRIAL REGISTRATION: ClinicalTrials.gov NCT03015363; https://clinicaltrials.gov/ct2/show/NCT03015363 (Archived by WebCite at http://www.webcitation.org/6xPZgseYS) |
format | Online Article Text |
id | pubmed-5943631 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | JMIR Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-59436312018-05-17 A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices Melville, Sarah Teskey, Robert Philip, Shona Simpson, Jeremy A Lutchmedial, Sohrab Brunt, Keith R J Med Internet Res Original Paper BACKGROUND: Clinical guidelines recommend monitoring of blood pressure at home using an automatic blood pressure device for the management of hypertension. Devices are not often calibrated against direct blood pressure measures, leaving health care providers and patients with less reliable information than is possible with current technology. Rigorous assessments of medical devices are necessary for establishing clinical utility. OBJECTIVE: The purpose of our study was 2-fold: (1) to assess the validity and perform iterative calibration of indirect blood pressure measurements by a noninvasive wrist cuff blood pressure device in direct comparison with simultaneously recorded peripheral and central intra-arterial blood pressure measurements and (2) to assess the validity of the measurements thereafter of the noninvasive wrist cuff blood pressure device in comparison with measurements by a noninvasive upper arm blood pressure device to the Canadian hypertension guidelines. METHODS: The cloud-based blood pressure algorithms for an oscillometric wrist cuff device were iteratively calibrated to direct pressure measures in 20 consented patient participants. We then assessed measurement validity of the device, using Bland-Altman analysis during routine cardiovascular catheterization. RESULTS: The precalibrated absolute mean difference between direct intra-arterial to wrist cuff pressure measurements were 10.8 (SD 9.7) for systolic and 16.1 (SD 6.3) for diastolic. The postcalibrated absolute mean difference was 7.2 (SD 5.1) for systolic and 4.3 (SD 3.3) for diastolic pressures. This is an improvement in accuracy of 33% systolic and 73% diastolic with a 48% reduction in the variability for both measures. Furthermore, the wrist cuff device demonstrated similar sensitivity in measuring high blood pressure compared with the direct intra-arterial method. The device, when calibrated to direct aortic pressures, demonstrated the potential to reduce a treatment gap in high blood pressure measurements. CONCLUSIONS: The systolic pressure measurements of the wrist cuff have been iteratively calibrated using gold standard central (ascending aortic) pressure. This improves the accuracy of the indirect measures and potentially reduces the treatment gap. Devices that undergo auscultatory (indirect) calibration for licensing can be greatly improved by additional iterative calibration via intra-arterial (direct) measures of blood pressure. Further clinical trials with repeated use of the device over time are needed to assess the reliability of the device in accordance with current and evolving guidelines for informed decision making in the management of hypertension. TRIAL REGISTRATION: ClinicalTrials.gov NCT03015363; https://clinicaltrials.gov/ct2/show/NCT03015363 (Archived by WebCite at http://www.webcitation.org/6xPZgseYS) JMIR Publications 2018-04-25 /pmc/articles/PMC5943631/ /pubmed/29695375 http://dx.doi.org/10.2196/jmir.8009 Text en ©Sarah Melville, Robert Teskey, Shona Philip, Jeremy A Simpson, Sohrab Lutchmedial, Keith R Brunt. Originally published in the Journal of Medical Internet Research (http://www.jmir.org), 25.04.2018. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Journal of Medical Internet Research, is properly cited. The complete bibliographic information, a link to the original publication on http://www.jmir.org/, as well as this copyright and license information must be included. |
spellingShingle | Original Paper Melville, Sarah Teskey, Robert Philip, Shona Simpson, Jeremy A Lutchmedial, Sohrab Brunt, Keith R A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices |
title | A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices |
title_full | A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices |
title_fullStr | A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices |
title_full_unstemmed | A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices |
title_short | A Comparison and Calibration of a Wrist-Worn Blood Pressure Monitor for Patient Management: Assessing the Reliability of Innovative Blood Pressure Devices |
title_sort | comparison and calibration of a wrist-worn blood pressure monitor for patient management: assessing the reliability of innovative blood pressure devices |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5943631/ https://www.ncbi.nlm.nih.gov/pubmed/29695375 http://dx.doi.org/10.2196/jmir.8009 |
work_keys_str_mv | AT melvillesarah acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT teskeyrobert acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT philipshona acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT simpsonjeremya acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT lutchmedialsohrab acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT bruntkeithr acomparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT melvillesarah comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT teskeyrobert comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT philipshona comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT simpsonjeremya comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT lutchmedialsohrab comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices AT bruntkeithr comparisonandcalibrationofawristwornbloodpressuremonitorforpatientmanagementassessingthereliabilityofinnovativebloodpressuredevices |