Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing

OBJECTIVE: To compare 3-year implant stability, survival, and tolerability of a 4.5-mm-wide (test) and a 3.75-mm-wide (control) percutaneous titanium implant for bone-conduction hearing, loaded with the sound processor after 3 weeks. METHODS: Sixty implants were allocated in a 2:1 ratio (test-contro...

Descripción completa

Detalles Bibliográficos
Autores principales: Kruyt, Ivo J., Nelissen, Rik C., Mylanus, Emmanuel A. M., Hol, Myrthe K. S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5959261/
https://www.ncbi.nlm.nih.gov/pubmed/29561377
http://dx.doi.org/10.1097/MAO.0000000000001761
_version_ 1783324364198903808
author Kruyt, Ivo J.
Nelissen, Rik C.
Mylanus, Emmanuel A. M.
Hol, Myrthe K. S.
author_facet Kruyt, Ivo J.
Nelissen, Rik C.
Mylanus, Emmanuel A. M.
Hol, Myrthe K. S.
author_sort Kruyt, Ivo J.
collection PubMed
description OBJECTIVE: To compare 3-year implant stability, survival, and tolerability of a 4.5-mm-wide (test) and a 3.75-mm-wide (control) percutaneous titanium implant for bone-conduction hearing, loaded with the sound processor after 3 weeks. METHODS: Sixty implants were allocated in a 2:1 ratio (test-control) in 57 adult patients included in this prospective randomized controlled clinical trial. Follow-up visits were performed at 7, 14, 21, and 28 days; 6 and 12 weeks; 6 months; and at 1, 2, and 3 years after implantation. During these visits, the implant stability quotient (ISQ) was measured by means of resonance frequency analysis (RFA). The peri-abutment soft tissue status was assessed according to the Holgers classification. Skin height around the abutment was evaluated. RESULTS: The mean area-under-the-curve (AUC) of ISQ-low was statistically significantly higher for the test implant (65.7 versus 61.4, p = 0.0002). Both implants showed high survival rates (97.4% versus 95.0%, p = 0.6374). Adverse soft tissue reactions were observed sporadically, with no significant inter-group differences. Skin thickening was seen in the majority of the patients, but no correlation with adverse soft tissue reactions or implant type was observed. CONCLUSION: The 4.5-mm-wide implant provides significantly higher ISQ values during the first 3 years after surgery compared with the previous generation 3.75-mm-wide implant. Both implants showed high survival rates and good tolerability. These long-term results indicate that the wider implant, loaded with a sound processor at 3 weeks, is a safe and well-performing option for hearing rehabilitation in specific types of hearing loss.
format Online
Article
Text
id pubmed-5959261
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Lippincott Williams & Wilkins
record_format MEDLINE/PubMed
spelling pubmed-59592612018-06-01 Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing Kruyt, Ivo J. Nelissen, Rik C. Mylanus, Emmanuel A. M. Hol, Myrthe K. S. Otol Neurotol Prosthetic Devices OBJECTIVE: To compare 3-year implant stability, survival, and tolerability of a 4.5-mm-wide (test) and a 3.75-mm-wide (control) percutaneous titanium implant for bone-conduction hearing, loaded with the sound processor after 3 weeks. METHODS: Sixty implants were allocated in a 2:1 ratio (test-control) in 57 adult patients included in this prospective randomized controlled clinical trial. Follow-up visits were performed at 7, 14, 21, and 28 days; 6 and 12 weeks; 6 months; and at 1, 2, and 3 years after implantation. During these visits, the implant stability quotient (ISQ) was measured by means of resonance frequency analysis (RFA). The peri-abutment soft tissue status was assessed according to the Holgers classification. Skin height around the abutment was evaluated. RESULTS: The mean area-under-the-curve (AUC) of ISQ-low was statistically significantly higher for the test implant (65.7 versus 61.4, p = 0.0002). Both implants showed high survival rates (97.4% versus 95.0%, p = 0.6374). Adverse soft tissue reactions were observed sporadically, with no significant inter-group differences. Skin thickening was seen in the majority of the patients, but no correlation with adverse soft tissue reactions or implant type was observed. CONCLUSION: The 4.5-mm-wide implant provides significantly higher ISQ values during the first 3 years after surgery compared with the previous generation 3.75-mm-wide implant. Both implants showed high survival rates and good tolerability. These long-term results indicate that the wider implant, loaded with a sound processor at 3 weeks, is a safe and well-performing option for hearing rehabilitation in specific types of hearing loss. Lippincott Williams & Wilkins 2018-06 2018-05-21 /pmc/articles/PMC5959261/ /pubmed/29561377 http://dx.doi.org/10.1097/MAO.0000000000001761 Text en Copyright © 2018 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of Otology & Neurotology, Inc. http://creativecommons.org/licenses/by-nc-nd/4.0 This is an open access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal. http://creativecommons.org/licenses/by-nc-nd/4.0
spellingShingle Prosthetic Devices
Kruyt, Ivo J.
Nelissen, Rik C.
Mylanus, Emmanuel A. M.
Hol, Myrthe K. S.
Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing
title Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing
title_full Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing
title_fullStr Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing
title_full_unstemmed Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing
title_short Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing
title_sort three-year outcomes of a randomized controlled trial comparing a 4.5-mm-wide to a 3.75-mm-wide titanium implant for bone conduction hearing
topic Prosthetic Devices
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5959261/
https://www.ncbi.nlm.nih.gov/pubmed/29561377
http://dx.doi.org/10.1097/MAO.0000000000001761
work_keys_str_mv AT kruytivoj threeyearoutcomesofarandomizedcontrolledtrialcomparinga45mmwidetoa375mmwidetitaniumimplantforboneconductionhearing
AT nelissenrikc threeyearoutcomesofarandomizedcontrolledtrialcomparinga45mmwidetoa375mmwidetitaniumimplantforboneconductionhearing
AT mylanusemmanuelam threeyearoutcomesofarandomizedcontrolledtrialcomparinga45mmwidetoa375mmwidetitaniumimplantforboneconductionhearing
AT holmyrtheks threeyearoutcomesofarandomizedcontrolledtrialcomparinga45mmwidetoa375mmwidetitaniumimplantforboneconductionhearing