Comparison of activated clotting times measured using the Hemochron Jr. Signature and Medtronic ACT Plus during cardiopulmonary bypass with acute normovolemic haemodilution

OBJECTIVE: This study compared the activated clotting time (ACT) measured using the Hemochron Jr. Signature (HACT) with the ACT measured using the Medtronic ACT Plus (MACT) during cardiopulmonary bypass (CPB) with acute normovolemic haemodilution (ANH) in patients undergoing cardiac surgery. METHODS...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Jung Min, Park, Eun Young, Kim, Kyung Mi, Won, Jong Chan, Jung, Tack Koon, Lee, Soo Kyung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5971518/
https://www.ncbi.nlm.nih.gov/pubmed/28974132
http://dx.doi.org/10.1177/0300060517731952
Descripción
Sumario:OBJECTIVE: This study compared the activated clotting time (ACT) measured using the Hemochron Jr. Signature (HACT) with the ACT measured using the Medtronic ACT Plus (MACT) during cardiopulmonary bypass (CPB) with acute normovolemic haemodilution (ANH) in patients undergoing cardiac surgery. METHODS: The ACT was checked at baseline with both devices after inducing anaesthesia, and 400 to 800 mL of whole blood was withdrawn to induce moderate ANH. Before initiating CPB, a 300-IU/kg bolus dose of heparin was administered to maintain the HACT at >400 s; protamine was later given to reverse the anticoagulation. The ACT was checked using both devices at baseline, during heparinisation, and after protamine administration. RESULTS: In total, 106 pairs of samples from 29 patients were analysed. The ACT showed a good correlation between the two devices (r = 0.956). However, Bland–Altman analysis showed that the MACT was higher, particularly at baseline and during heparinisation. Multiple regression analysis showed that the blood glucose concentration significantly influenced the differences between the two ACT devices. CONCLUSIONS: The HACT was lower than the MACT during CPB with ANH in patients undergoing cardiac surgery. Clinicians should be cautious when using each ACT device within generally accepted reference ACT values.