扫描及重建条件对肺结节三维容积及CT值定量影响的体模研究

BACKGROUND AND OBJECTIVE: The computed tomography (CT) follow-up of indeterminate pulmonary nodules aiming to evaluate the change of the volume and CT value is the common strategy in clinic. The CT dose needs to considered on serious CT scans in addition to the measurement accuracy. The purpose of t...

Descripción completa

Detalles Bibliográficos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: 中国肺癌杂志编辑部 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5973001/
https://www.ncbi.nlm.nih.gov/pubmed/28855038
http://dx.doi.org/10.3779/j.issn.1009-3419.2017.08.10
_version_ 1783326520472764416
collection PubMed
description BACKGROUND AND OBJECTIVE: The computed tomography (CT) follow-up of indeterminate pulmonary nodules aiming to evaluate the change of the volume and CT value is the common strategy in clinic. The CT dose needs to considered on serious CT scans in addition to the measurement accuracy. The purpose of this study is to quantify the precision of pulmonary nodule volumetric measurement and CT value measurement with various tube currents and reconstruction algorithms in a phantom study with dual-energy CT. METHODS: A chest phantom containing 9 artificial spherical solid nodules with known diameter (D=2.5 mm, 5 mm, 10 mm) and density (-100 HU, 60 HU and 100 HU) was scanned using a 64-row detector CT canner at 120 Kilovolt & various currents (10 mA, 20 mA, 50 mA, 80 mA, 100 mA, 150 mA and 350 mA). Raw data were reconstructed with filtered back projection and three levels of adaptive statistical iterative reconstruction algorithm (FBP, ASIR; 30%, 50% and 80%). Automatic volumetric measurements were performed using commercially available software. The relative volume error (RVE) and the absolute attenuation error (AAE) between the software measures and the referencestandard were calculated. Analyses of the variance were performed to evaluate the effect of reconstruction methods, different scan parameters, nodule size and attenuation on the RPE. RESULTS: The software substantially overestimated the very small (D=2.5 mm) nodule's volume [mean RVE: (100.8%±28%)] and underestimated it attenuation [mean AAE: (-756±80) HU]. The mean RVEs of nodule with diameter as 5 mm and 10 mm were small [(-0.9%±1.1%) vs (0.9%±1.4%)], however, the mean AAEs [(-243±26) HU vs (-129±7) HU)] were large. The ANOVA analysis for repeated measurements showed that different tube current and reconstruction algorithm had no significant effect on the volumetric measurements for nodules with diameter of 5 mm and 10 mm (F=5.60, P=0.10 vs F=11.13, P=0.08), but significant effects on the measurement of CT value (F=34.79, P < 0.001 vs F=156.14, P < 0.001). CONCLUSION: An infinitesimally small errors of volumetric measurement of 5 mm or 10 mm nodule could achieved with very low current and ASIR reconstruction, suggesting a possibility of remarkable radiation dose reductions, while it is not applicable for 5 mm nodule. The attenuation acquired through three dimensional software has large measurement error and can not applied in clinical currently.
format Online
Article
Text
id pubmed-5973001
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher 中国肺癌杂志编辑部
record_format MEDLINE/PubMed
spelling pubmed-59730012018-07-06 扫描及重建条件对肺结节三维容积及CT值定量影响的体模研究 Zhongguo Fei Ai Za Zhi 临床研究 BACKGROUND AND OBJECTIVE: The computed tomography (CT) follow-up of indeterminate pulmonary nodules aiming to evaluate the change of the volume and CT value is the common strategy in clinic. The CT dose needs to considered on serious CT scans in addition to the measurement accuracy. The purpose of this study is to quantify the precision of pulmonary nodule volumetric measurement and CT value measurement with various tube currents and reconstruction algorithms in a phantom study with dual-energy CT. METHODS: A chest phantom containing 9 artificial spherical solid nodules with known diameter (D=2.5 mm, 5 mm, 10 mm) and density (-100 HU, 60 HU and 100 HU) was scanned using a 64-row detector CT canner at 120 Kilovolt & various currents (10 mA, 20 mA, 50 mA, 80 mA, 100 mA, 150 mA and 350 mA). Raw data were reconstructed with filtered back projection and three levels of adaptive statistical iterative reconstruction algorithm (FBP, ASIR; 30%, 50% and 80%). Automatic volumetric measurements were performed using commercially available software. The relative volume error (RVE) and the absolute attenuation error (AAE) between the software measures and the referencestandard were calculated. Analyses of the variance were performed to evaluate the effect of reconstruction methods, different scan parameters, nodule size and attenuation on the RPE. RESULTS: The software substantially overestimated the very small (D=2.5 mm) nodule's volume [mean RVE: (100.8%±28%)] and underestimated it attenuation [mean AAE: (-756±80) HU]. The mean RVEs of nodule with diameter as 5 mm and 10 mm were small [(-0.9%±1.1%) vs (0.9%±1.4%)], however, the mean AAEs [(-243±26) HU vs (-129±7) HU)] were large. The ANOVA analysis for repeated measurements showed that different tube current and reconstruction algorithm had no significant effect on the volumetric measurements for nodules with diameter of 5 mm and 10 mm (F=5.60, P=0.10 vs F=11.13, P=0.08), but significant effects on the measurement of CT value (F=34.79, P < 0.001 vs F=156.14, P < 0.001). CONCLUSION: An infinitesimally small errors of volumetric measurement of 5 mm or 10 mm nodule could achieved with very low current and ASIR reconstruction, suggesting a possibility of remarkable radiation dose reductions, while it is not applicable for 5 mm nodule. The attenuation acquired through three dimensional software has large measurement error and can not applied in clinical currently. 中国肺癌杂志编辑部 2017-08-20 /pmc/articles/PMC5973001/ /pubmed/28855038 http://dx.doi.org/10.3779/j.issn.1009-3419.2017.08.10 Text en 版权所有©《中国肺癌杂志》编辑部2017 https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed in accordance with the terms of the Creative Commons Attribution (CC BY 3.0) License. See: https://creativecommons.org/licenses/by/3.0/
spellingShingle 临床研究
扫描及重建条件对肺结节三维容积及CT值定量影响的体模研究
title 扫描及重建条件对肺结节三维容积及CT值定量影响的体模研究
title_full 扫描及重建条件对肺结节三维容积及CT值定量影响的体模研究
title_fullStr 扫描及重建条件对肺结节三维容积及CT值定量影响的体模研究
title_full_unstemmed 扫描及重建条件对肺结节三维容积及CT值定量影响的体模研究
title_short 扫描及重建条件对肺结节三维容积及CT值定量影响的体模研究
title_sort 扫描及重建条件对肺结节三维容积及ct值定量影响的体模研究
topic 临床研究
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5973001/
https://www.ncbi.nlm.nih.gov/pubmed/28855038
http://dx.doi.org/10.3779/j.issn.1009-3419.2017.08.10
work_keys_str_mv AT sǎomiáojízhòngjiàntiáojiànduìfèijiéjiésānwéiróngjījíctzhídìngliàngyǐngxiǎngdetǐmóyánjiū
AT sǎomiáojízhòngjiàntiáojiànduìfèijiéjiésānwéiróngjījíctzhídìngliàngyǐngxiǎngdetǐmóyánjiū
AT sǎomiáojízhòngjiàntiáojiànduìfèijiéjiésānwéiróngjījíctzhídìngliàngyǐngxiǎngdetǐmóyánjiū
AT sǎomiáojízhòngjiàntiáojiànduìfèijiéjiésānwéiróngjījíctzhídìngliàngyǐngxiǎngdetǐmóyánjiū