Cargando…

Study of new practical ESR dosimeter based on carbonated hydroxyapatite and its dosimetric properties

The development of new dosimeters with good dosimetric properties is important for quality control in radiation applications. A new practical electron spin resonance (ESR) dosimeter based on carbonated hydroxyapatite that simulated the composition and structure of tooth enamel was specially synthesi...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Ye, Ma, Lei, Guo, Junwang, Dong, Guofu, Cong, Jianbo, Ji, Yunlong, Ning, Jing, Yang, Guoshan, Wu, Ke
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5973591/
https://www.ncbi.nlm.nih.gov/pubmed/29813104
http://dx.doi.org/10.1371/journal.pone.0197953
Descripción
Sumario:The development of new dosimeters with good dosimetric properties is important for quality control in radiation applications. A new practical electron spin resonance (ESR) dosimeter based on carbonated hydroxyapatite that simulated the composition and structure of tooth enamel was specially synthesized. The synthesized material was investigated by transmission electron microscope, X-ray diffraction, fourier transform infrared spectroscopy and X-ray photo electron spectroscopy to confirm to the main composition of carbonated hydroxyapatite with CO(3)(2-) successfully doped into the crystal lattice through optimizing the synthesis process of C/P molar ratio, pH value dynamical adjustment, annealing temperature and time. The dosimetric properties were systematically investigated by ESR spectroscopy. The results indicated that the radiation induced signal had a good dose response within a relatively wide dose range. The dose response was linear in the dose range of 0–400 Gy with a correlation coefficient of 0.9999 and had dose accumulative effect in the experimental dose range of 0–100 Gy. In a wider dose range up to 30 kGy, the dose response also presented linear feature in double-logarithmic coordinate system with a correlation coefficient of 0.9970. The dose detection limit was about 0.34Gy with a given probability of 95% confidence level depending upon a rigid calculation algorithm. The signal was extremely stable in the observation time of 360 days with a variation coefficient of 3.8%. The radiation sensitivity of the material showed no remarkable variation against photon energy from 662 KeV to 1.25 MeV and dose rate from 0.86 Gy/min to 12.17 Gy/min. The material showed more sensitive in lower photon energy range below 662 keV, which hint additional calibration may need when using in special photon energy condition. The preliminary results suggested that this newly developed dosimeter was potential to become a practical dosimeter that would expand the application fields of ESR dosimetry.